The influence of shape and sliding distance of femoral head movement loci on the wear of acetabular cups in total hip arthroplasty

被引:17
作者
Bennett, D
Orr, JF
Beverland, DE
Baker, R
机构
[1] Musgrave Pk Hosp, Belfast BT9 7JB, Antrim, North Ireland
[2] Queens Univ Belfast, Sch Mech & Mfg Engn, Belfast, Antrim, North Ireland
[3] Royal Childrens Hosp, Hugh Williamson Gait Lab, Parkville, Vic 3052, Australia
关键词
movement loci; total hip replacement (THR); wear; UHMWPE;
D O I
10.1243/095441102321032184
中图分类号
R318 [生物医学工程];
学科分类号
0831 [生物医学工程];
摘要
Wear of the polyethylene acetabular component is the most serious threat to the long-term success of total hip replacements (THRs). Greatly reduced wear rates have been reported for unidirectional, compared to multidirectional, articulation in vitro. This study considers the multidirectional motions experienced at the hip joint as described by movement loci of points on the femoral head for individual THR patients. A three-dimensional computer program determined the movement loci of selected points on the femoral head for THR patients and normal subjects using kinematic data obtained from gait analysis. The sizes and shapes of these loci were quantified by their sliding distances and aspect ratios with substantial differences exhibited between individual THR patients. The average sliding distances ranged from 10.0 to 18.1 mm and the average aspect ratios of the loci ranged from 2.5 to 9.2 for the THR patients. Positive correlations were found between wear rate and average sliding distance, the inverse of the average aspect ratio of the loci and the product of the average sliding distance and the inverse of the average aspect ratio of the loci. Patients with a normal hip joint range of motion produce multidirectional motion loci and tend to experience more wear than patients with more unidirectional motion loci. Differing patterns of multidirectional motion at the hip joint for individual THR patients may explain widely differing wear rates in vivo.
引用
收藏
页码:393 / 402
页数:10
相关论文
共 30 条
[1]
CONTACT AND RUBBING OF FLAT SURFACES [J].
ARCHARD, JF .
JOURNAL OF APPLIED PHYSICS, 1953, 24 (08) :981-988
[2]
LABORATORY WEAR TESTS AND CLINICAL OBSERVATIONS OF THE PENETRATION OF FEMORAL HEADS INTO ACETABULAR CUPS IN TOTAL REPLACEMENT HIP JOINTS .3. THE MEASUREMENT OF INTERNAL VOLUME CHANGES IN EXPLANTED CHARNLEY SOCKETS AFTER 2-16 YEARS INVIVO AND THE DETERMINATION OF WEAR FACTORS [J].
ATKINSON, JR ;
DOWSON, D ;
ISAAC, JH ;
WROBLEWSKI, BM .
WEAR, 1985, 104 (03) :225-244
[3]
Bankston A B, 1994, J Arthroplasty, V9, P533, DOI 10.1016/0883-5403(94)90101-5
[4]
A hip joint simulator study using simplified loading and motion cycles generating physiological wear paths and rates [J].
Barbour, PSM ;
Stone, MH ;
Fisher, J .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 1999, 213 (H6) :455-467
[5]
Movement loci of selected points on the femoral head for individual total hip arthroplasty patients using three-dimensional computer simulation [J].
Bennett, DB ;
Orr, JF ;
Baker, R .
JOURNAL OF ARTHROPLASTY, 2000, 15 (07) :909-915
[6]
BESONG AA, 1999, P IMECHE C KNEE REPL, P70
[7]
BLAND M, 1997, INTRO MED STAT, P196
[8]
Bragdon C R, 1996, Proc Inst Mech Eng H, V210, P157, DOI 10.1243/PIME_PROC_1996_210_408_02
[9]
The effects of the concentration of high-density polyethylene particles on the bone-implant interface [J].
Brooks, RA ;
Sharpe, JR ;
Wimhurst, JA ;
Myer, BJ ;
Dawes, EN ;
Rushton, N .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2000, 82B (04) :595-600
[10]
BYGOTT J, 1967, INTRO MAPWORK PRACTI, P228