Model evaluation and calibration for prospective respiratory motion correction in coronary MR angiography based on 3-D image registration

被引:38
作者
Manke, D
Rösch, P
Nehrke, K
Börnert, P
Dössel, O
机构
[1] Univ Karlsruhe, Inst Biomed Engn, D-76128 Karlsruhe, Germany
[2] Philips Res Labs, Sector Tech Syst, D-22335 Hamburg, Germany
关键词
coronary magnetic resonance angiography; image registration; prospective motion correction;
D O I
10.1109/TMI.2002.804428
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
Image processing was used as a fundamental tool to derive motion information from magnetic resonance (MR) images, which was fed back into prospective respiratory motion correction during subsequent data acquisition to improve image quality in coronary MR angiography (CMRA) scans. This reduces motion artifacts in the images and, in addition, enables the usage of a broader gating window than commonly used today to increase the scan efficiency. The aim of the study reported in this paper was to find a suitable motion model to be used for respiratory motion correction in cardiac imaging and to develop a calibration procedure to adapt the motion model to the individual patient. At first, the performance of three motion models [one-dimensional translation in feet-head (FH) direction, three-dimensional (3-D) translation, and 3-D affine transformation] was tested in a small volunteer study. An elastic image registration algorithm was applied to 3-D MR images of the coronary vessels obtained at different respiratory levels. A strong intersubject variability was observed. The 3-D translation and affine transformation model were found to be superior over the conventional FH translation model used today. Furthermore, a new approach is presented, which utilizes a fast model-based image registration to extract motion information from time series of low-resolution 3-D MR images, which reflects the respiratory motion of the heart. The registration is based on a selectable global 3-D motion model (translation, rigid, or affine transformation). All 3-D MR images were registered with respect to end expiration. The resulting time series of model parameters were analyzed in combination with additionally acquired motion information from a diaphragmatic MR pencil-beam navigator to calibrate the respiratory motion model. To demonstrate the potential of a calibrated motion model for prospective motion correction in coronary imaging, the approach was tested in CMRA examinations in five volunteers.
引用
收藏
页码:1132 / 1141
页数:10
相关论文
共 28 条
[21]  
Schnabel JA., 2001, LECT NOTES COMPUTER, P573, DOI DOI 10.1007/3-540-45468-3_69
[22]  
Stuber M, 2001, J MAGN RESON IMAGING, V13, P185
[23]   Double-oblique free-breathing high resolution three-dimensional coronary magnetic resonance angiography [J].
Stuber, M ;
Botnar, RM ;
Danias, PG ;
Sodickson, DK ;
Kissinger, KV ;
Van Cauteren, M ;
De Becker, J ;
Manning, WJ .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1999, 34 (02) :524-531
[24]   Submillimeter three-dimensional coronary MR angiography with real-time navigator correction: Comparison of navigator locations [J].
Stuber, M ;
Botnar, RM ;
Danias, PG ;
Kissinger, KV ;
Manning, WJ .
RADIOLOGY, 1999, 212 (02) :579-587
[25]  
Taylor AM, 1999, J MAGN RESON IMAGING, V9, P786, DOI 10.1002/(SICI)1522-2586(199906)9:6<786::AID-JMRI5>3.0.CO
[26]  
2-T
[27]   RESPIRATORY MOTION OF THE HEART - KINEMATICS AND THE IMPLICATIONS FOR THE SPATIAL-RESOLUTION IN CORONARY IMAGING [J].
WANG, Y ;
RIEDERER, SJ ;
EHMAN, RL .
MAGNETIC RESONANCE IN MEDICINE, 1995, 33 (05) :713-719
[28]  
WOLBERG G, 1992, DIGITAL IMAGE WARPIN, P41