A model for calculating dead time and mass discrimination correction factors from inductively coupled plasma mass spectrometry calibration curves

被引:33
作者
Appelblad, PK [1 ]
Baxter, DC [1 ]
机构
[1] Lulea Univ Technol, Div Inorgan Chem, S-97187 Lulea, Sweden
关键词
D O I
10.1039/b001152p
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A model allowing simultaneous determination of the detector dead time and the mass discrimination factor in inductively coupled plasma mass spectrometry (ICP-MS), as well as the corresponding uncertainties, is presented and compared for three representative isotope systems, namely magnesium, indium and thallium. The advantages of using the model presented are firstly that both the detector dead time and the mass discrimination factor can be obtained simultaneously and secondly that the sampling time can be spent entirely on the isotopes of interest.