The incidence of revision of the metal component of total knee arthroplasties in different tibial-insert designs

被引:9
作者
Huang, CH
Liau, JJ
Lung, CY
Lan, CT
Cheng, CK
机构
[1] Natl Yang Ming Univ, Inst Biomed Engn, Orthopaed Biomech Lab, Taipei 112, Taiwan
[2] Mackay Mem Hosp, Dept Orthopaed Surg, Taipei, Taiwan
关键词
polyethylene wear; metallic wear; mobile bearing; fixed bearing; revision;
D O I
10.1016/S0968-0160(02)00054-6
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Seventy-one revisions of knee arthroplasty due to wear of modular tibial polyethylene inserts undertaken from May 1990 to May 1999 were reviewed. In 38 cases (53.5%), severe damage was noted that necessitated replacement of the metal components. This proportion was less than in a previous study (88.7%). All the cases were further divided into two groups based on the design of the tibial component, one group having mobile bearings and the other fixed. In only 17.7% of the mobile group was there a need to replace metal parts at revision, whereas 83.8% of the fixed group required such replacement, which is much more comparable to the previous study. Therefore, the incidence of revision of the metallic component of a total knee arthroplasty depends greatly on the design of the tibial components. (C) 2002 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:331 / 334
页数:4
相关论文
共 25 条
[1]
PERFORMANCE OF THE TIBIAL COMPONENT IN TOTAL KNEE REPLACEMENT - CONVENTIONAL AND REVISION DESIGNS [J].
BARTEL, DL ;
BURSTEIN, AH ;
SANTAVICCA, EA ;
INSALL, JN .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1982, 64 (07) :1026-1033
[2]
The incidence of modular tibial polyethylene insert exchange in total knee arthroplasty when polyethylene failure occurs [J].
Bert, JM ;
Reuben, J ;
Kelly, F ;
Gross, M ;
Elting, J .
JOURNAL OF ARTHROPLASTY, 1998, 13 (06) :609-614
[3]
COLLIER JP, 1991, CLIN ORTHOP RELAT R, P232
[4]
FATIGUE FAILURE OF NONCEMENTED POROUS-COATED IMPLANTS - A RETRIEVAL STUDY [J].
COOK, SD ;
THOMAS, KA .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1991, 73 (01) :20-24
[5]
POLYETHYLENE WEAR OF METAL-BACKED TIBIAL COMPONENTS IN TOTAL AND UNICOMPARTMENTAL KNEE PROSTHESES [J].
ENGH, GA ;
DWYER, KA ;
HANES, CK .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1992, 74 (01) :9-17
[6]
KINEMATIC TOTAL KNEE REPLACEMENT [J].
EWALD, FC ;
JACOBS, MA ;
MIEGEL, RE ;
WALKER, PS ;
POSS, R ;
SLEDGE, CB .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1984, 66A (07) :1032-1040
[7]
EZZET KA, 1995, CLIN ORTHOP RELAT R, P86
[8]
FRIEDMAN RJ, 1994, AAOS INSTR COURS LEC, V43, P233
[9]
Fracture of the femoral component associated with polyethylene wear and osteolysis after total knee arthroplasty [J].
Huang, CH ;
Yang, CY ;
Cheng, CK .
JOURNAL OF ARTHROPLASTY, 1999, 14 (03) :375-379
[10]
Huang CH, 1998, J BIOMED MATER RES, V39, P153, DOI 10.1002/(SICI)1097-4636(199801)39:1<153::AID-JBM17>3.0.CO