Investigation of a high-efficiency nebulizer and a thimble glass frit nebulizer for elemental analysis of biological materials by inductively coupled plasma-atomic emission spectrometry

被引:38
作者
Liu, HY
Clifford, RH
Dolan, SP
Montaser, A
机构
[1] GEORGE WASHINGTON UNIV,DEPT CHEM,WASHINGTON,DC 20052
[2] US FDA,ELEMENTAL RES BRANCH,WASHINGTON,DC 20204
关键词
biological material; core nebulizer; demountable thimble glass frit nebulizer; detection limit; high-efficiency nebulizer; inductively coupled plasma atomic emission spectrometry; thimble frit nebulizer;
D O I
10.1016/0584-8547(95)01423-3
中图分类号
O433 [光谱学];
学科分类号
0703 ; 070302 ;
摘要
A high-efficiency nebulizer (HEN) and a demountable thimble glass frit nebulizer are evaluated for argon inductively coupled plasma-atomic emission spectrometry (Ar ICP-AES). Aerosol characteristics for the HEN and the thimble frit are studied using laser Fraunhofer diffraction. Analyte transport efficiencies, mass transport rates, detection limits, and short- and long-term stabilities are measured. Four standard reference materials (trace elements in water, oyster tissue, bovine liver, and orchard leaves) are analyzed. For comparison, results obtained by a high-solid pneumatic nebulizer (conespray nebulizer) are also presented. Transport efficiencies of up to 50-60% are obtained for the HEN and the thimble frit nebulizer at an uptake rate of 11 mu l min(-1). For the elements tested. short- and long-term precisions and detection limits obtained with the thimble frit are comparable to those of a commonly used pneumatic nebulizer. Good agreement with certified values is obtained for six trace elements in a standard reference material (SRM) water sample nebulized by the thimble frit. However, signal suppression is observed when the biological or plant SRMs are nebulized with the thimble frit nebulizer. In contrast, satisfactory results are obtained for the same matrices when the HEN and the conespray nebulizer are used.
引用
收藏
页码:27 / 40
页数:14
相关论文
共 43 条
[1]  
APEL CT, 1989, ICP INFORMATION NEWS, V9, P503
[2]  
APEL CT, 1988, Patent No. 4743407
[3]  
BROWNER RF, 1987, INDUCTIVELY COUPLE 2
[4]   DRIFT DIAGNOSTICS IN INDUCTIVELY COUPLED PLASMA ATOMIC EMISSION-SPECTROMETRY [J].
CARRE, M ;
POUSSEL, E ;
MERMET, JM .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1992, 7 (06) :791-797
[5]   A SIMPLE, NON-INVASIVE METHOD FOR THE MEASUREMENT OF GAS-FLOW VELOCITIES IN THE INDUCTIVELY COUPLED PLASMA [J].
CICERONE, MT ;
FARNSWORTH, PB .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1989, 44 (09) :897-907
[6]   DUAL-BEAM, LIGHT-SCATTERING INTERFEROMETRY FOR SIMULTANEOUS MEASUREMENTS OF DROPLET-SIZE AND VELOCITY DISTRIBUTIONS OF AEROSOLS FROM COMMONLY USED NEBULIZERS [J].
CLIFFORD, RH ;
ISHII, I ;
MONTASER, A ;
MEYER, GA .
ANALYTICAL CHEMISTRY, 1990, 62 (04) :390-394
[7]   THIMBLE GLASS FRIT NEBULIZER FOR ATOMIC SPECTROMETRY [J].
CLIFFORD, RH ;
MONTASER, A ;
SINEX, SA ;
CAPAR, SG .
ANALYTICAL CHEMISTRY, 1989, 61 (24) :2777-2784
[8]   PARTICLE-SIZE MEASUREMENTS IN THE SUBMICRON RANGE BY THE DIFFERENTIAL ELECTROMOBILITY TECHNIQUE - COMPARISON OF AEROSOLS FROM THERMOSPRAY, ULTRASONIC, PNEUMATIC AND FRIT-TYPE NEBULIZERS [J].
CLIFFORD, RH ;
TAN, HM ;
LIU, HY ;
MONTASER, A ;
ZARRIN, F ;
KEADY, PB .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1993, 48 (10) :1221-1235
[9]   DIAGNOSTIC STUDIES ON DESOLVATED AEROSOLS FROM ULTRASONIC NEBULIZERS [J].
CLIFFORD, RH ;
SOHAL, P ;
LIU, HY ;
MONTASER, A .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1992, 47 (09) :1107-1122
[10]   CONVERSION OF AN ULTRASONIC HUMIDIFIER TO A CONTINUOUS-TYPE ULTRASONIC NEBULIZER FOR ATOMIC SPECTROMETRY [J].
CLIFFORD, RH ;
MONTASER, A ;
DOLAN, SP ;
CAPAR, SG .
ANALYTICAL CHEMISTRY, 1990, 62 (24) :2745-2749