Spatially resolved diffuse reflectance for the determination of tissue optical properties and metabolic state.

被引:5
作者
Forrester, KR [1 ]
Shymkiw, R [1 ]
Yeung, G [1 ]
Irwin, P [1 ]
Tulip, J [1 ]
Bray, R [1 ]
机构
[1] Univ Calgary, Calgary, AB T2N 4N1, Canada
来源
LASER-TISSUE INTERACTION XI: PHOTOCHEMICAL, PHOTOTHERMAL, AND PHOTOMECHANICAL | 2000年 / 3914卷
关键词
diffuse reflectance spectroscopy; Monte Carlo; collagen; light guide; metabolic state;
D O I
10.1117/12.388060
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A multiple wavelength diffuse reflectance instrument was evaluated for the in-situ measurement of tissue optical properties. A diffusion dipole model with a non-linear least-squares fitting procedure was used to generate optical properties from the radial profiles of diffuse reflectance measurements. The diffusion model was compared with Monte Carlo simulations showing acceptable agreement for a range of distances far enough away from the source point. Instrumentation was evaluated by experimental measurement of tissue simulating phantoms where measured optical properties were compared to those obtained using a collimated transmission setup. The results showed agreement to within 5-10%. The optical properties of rabbit ligament and tendon are reported where anisotropic scattering due to collagen alignment was observed. This study represents the first known measurement of rabbit ligament and tendon optical properties and the use of diffuse reflectance to measure the "lightguide" effects of collagen alignment. To evaluate instrument response to changes in tissue oxygenation levels the optical properties of human skin were measured before and during arterial occlusion. During arterial occlusion absorption at 633 nm increased while absorption at 810 nm remained relatively constant. These results are consistent with the deoxygenation of hemoglobin that occurs during occlusion.
引用
收藏
页码:333 / 344
页数:12
相关论文
共 16 条
  • [1] Bevington R., 1969, DATA REDUCTION ERROR
  • [2] Evaluation of laser Doppler imaging to measure blood flow in knee ligaments of adult rabbits
    Bray, R
    Forrester, K
    McDougall, JJ
    Damji, A
    Ferrell, WR
    [J]. MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1996, 34 (03) : 227 - 231
  • [3] A DIFFUSION-THEORY MODEL OF SPATIALLY RESOLVED, STEADY-STATE DIFFUSE REFLECTANCE FOR THE NONINVASIVE DETERMINATION OF TISSUE OPTICAL-PROPERTIES INVIVO
    FARRELL, TJ
    PATTERSON, MS
    WILSON, B
    [J]. MEDICAL PHYSICS, 1992, 19 (04) : 879 - 888
  • [4] FARRELL TJ, P SPIE, V2135, P117
  • [5] OPTICAL-PROPERTIES OF INTRALIPID - A PHANTOM MEDIUM FOR LIGHT-PROPAGATION STUDIES
    FLOCK, ST
    JACQUES, SL
    WILSON, BC
    STAR, WM
    VANGEMERT, MJC
    [J]. LASERS IN SURGERY AND MEDICINE, 1992, 12 (05) : 510 - 519
  • [6] In vivo comparison of scanning technique and wavelength in laser Doppler perfusion imaging: measurement in knee ligaments of adult rabbits
    Forrester, K
    Doschak, M
    Bray, R
    [J]. MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1997, 35 (06) : 581 - 586
  • [7] THE USE OF INDIA INK AS AN OPTICAL ABSORBER IN TISSUE-SIMULATING PHANTOMS
    MADSEN, SJ
    PATTERSON, MS
    WILSON, BC
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1992, 37 (04) : 985 - 993
  • [8] MEASUREMENTS AND CALCULATIONS OF THE ENERGY FLUENCE RATE IN A SCATTERING AND ABSORBING PHANTOM AT 633 NM
    MOES, CJM
    VANGEMERT, MJC
    STAR, WM
    MARIJNISSEN, JPA
    PRAHL, SA
    [J]. APPLIED OPTICS, 1989, 28 (12): : 2292 - 2296
  • [9] DEPTH DISCRIMINATION IN LASER DOPPLER SKIN BLOOD-FLOW MEASUREMENT USING DIFFERENT LASERS
    OBEID, AN
    BOGGETT, DM
    BARNETT, NJ
    DOUGHERTY, G
    ROLFE, P
    [J]. MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1988, 26 (04) : 415 - 419
  • [10] Prahl S., TABULATED MOLAR EXTI