Three-dimensional diffuse optical tomography in the parallel plane transmission geometry: Evaluation of a hybrid frequency domain/continuous wave clinical system for breast imaging

被引:236
作者
Culver, JP
Choe, R
Holboke, MJ
Zubkov, L
Durduran, T
Slemp, A
Ntziachristos, V
Chance, B
Yodh, AG
机构
[1] Univ Penn, Dept Phys & Astron, Philadelphia, PA 19104 USA
[2] Univ Penn, Dept Biochem & Biophys, Philadelphia, PA 19104 USA
关键词
optical tomography; NIR spectroscopy; breast imaging; clinical system;
D O I
10.1118/1.1534109
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Three-dimensional diffuse optical tomography (DOT) of breast requires large data sets for even modest resolution (1 cm). We present a hybrid DOT system that combines a limited number of frequency domain (FD) measurements with a large set of continuous wave (cw) measurements. The FD measurements are used to quantitatively determine tissue averaged absorption and scattering coefficients. The larger cw data sets (105 measurements) collected with a lens coupled CCD, permit 3D DOT reconstructions of a 1-liter tissue volume. To address the computational complexity of large data sets and 3D volumes we employ finite difference based reconstructions computed in parallel. Tissue phantom measurements evaluate imaging performance. The tests include the following: point spread function measures of resolution, characterization of the size and contrast of single objects, field of view measurements and spectral characterization of constituent concentrations. We also report in vivo measurements. Average tissue optical properties of a healthy breast are used to deduce oxy- and deoxy-hemoglobin concentrations. Differential imaging with a tumor simulating target adhered to the surface of a healthy breast evaluates the influence of physiologic fluctuations on image noise. This tomography system provides robust, quantitative, full 3D image reconstructions with the advantages of high data throughput, single detector-tissue coupling path, and large (1L) imaging domains. In addition, we find that point spread function measurements provide a useful and comprehensive representation of system performance. (C) 2003 American Association of Physicists in Medicine.
引用
收藏
页码:235 / 247
页数:13
相关论文
共 50 条
[21]   TIME-RESOLVED OPTICAL IMAGING OF A SOLID TISSUE-EQUIVALENT PHANTOM [J].
HEBDEN, JC ;
HALL, DJ ;
FIRBANK, M ;
DELPY, DT .
APPLIED OPTICS, 1995, 34 (34) :8038-8047
[22]   Three-dimensional time-resolved optical tomography of a conical breast phantom [J].
Hebden, JC ;
Veenstra, H ;
Dehghani, H ;
Hillman, EMC ;
Schweiger, M ;
Arridge, SR ;
Delpy, DT .
APPLIED OPTICS, 2001, 40 (19) :3278-3287
[23]   Time resolved optical tomography of the human forearm [J].
Hillman, EMC ;
Hebden, JC ;
Schweiger, M ;
Dehghani, H ;
Schmidt, FEW ;
Delpy, DT ;
Arridge, SR .
PHYSICS IN MEDICINE AND BIOLOGY, 2001, 46 (04) :1117-1130
[24]   Three-dimensional diffuse optical mammography with ultrasound localization in a human subject [J].
Holboke, MJ ;
Tromberg, BJ ;
Li, X ;
Shah, N ;
Fishkin, J ;
Kidney, D ;
Butler, J ;
Chance, B ;
Yodh, AG .
JOURNAL OF BIOMEDICAL OPTICS, 2000, 5 (02) :237-247
[25]   Three-dimensional optical tomographic Imaging of breast in a human subject [J].
Jiang, H ;
Xu, Y ;
Iftimia, N ;
Eggert, J ;
Klove, K ;
Baron, L ;
Fajardo, L .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 2001, 20 (12) :1334-1340
[26]   Experimental three-dimensional optical image reconstruction of heterogeneous turbid media from continuous-wave data [J].
Jiang, HB ;
Xu, Y ;
Iftimia, N .
OPTICS EXPRESS, 2000, 7 (05) :204-209
[27]   Near-field diffraction tomography with diffuse photon density waves [J].
Li, XD ;
Pattanayak, DN ;
Durduran, T ;
Culver, JP ;
Chance, B ;
Yodh, AG .
PHYSICAL REVIEW E, 2000, 61 (04) :4295-4309
[28]   Inverse problem in optical diffusion tomography. I. Fourier-Laplace inversion formulas [J].
Markel, VA ;
Schotland, JC .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 2001, 18 (06) :1336-1347
[29]   Initial studies of in vivo absorbing and scattering heterogeneity in near-infrared tomographic breast imaging [J].
McBride, TO ;
Pogue, BW ;
Jiang, SD ;
Österberg, UL ;
Paulsen, KD ;
Poplack, SP .
OPTICS LETTERS, 2001, 26 (11) :822-824
[30]   A parallel-detection frequency-domain near-infrared tomography system for hemoglobin imaging of the breast in vivo [J].
McBride, TO ;
Pogue, BW ;
Jiang, S ;
Osterberg, UL ;
Paulsen, KD .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2001, 72 (03) :1817-1824