Measurement of 25-OH-vitamin D in human serum using liquid chromatography tandem-mass spectrometry with comparison to radioimmunoassay and automated immunoassay

被引:121
作者
van den Ouweland, Johannes M. W. [1 ]
Beijers, Antonius M. [1 ]
Demacker, Pierre N. M. [1 ]
van Daal, Henny [1 ]
机构
[1] Canisius Wilhelmina Hosp, Dept Clin Chem, NL-6532 SZ Nijmegen, Netherlands
来源
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES | 2010年 / 878卷 / 15-16期
关键词
25-Hydroxyvitamin D; LC-MS/MS; Radioimmunoassay; Method comparison; VITAMIN-D DEFICIENCY; 25-HYDROXYVITAMIN D-3; QUANTIFICATION; ACCURACY;
D O I
10.1016/j.jchromb.2010.03.035
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
The plasma 25-OH vitamin D concentration is a reliable biomarker for vitamin D status but assay's variability makes adequate monitoring of vitamin D status difficult. We employed isotope-dilution liquid chromatography (LC) tandem-mass spectrometry (MS/MS) for the measurement of both 25-OH vitamin D(3) and 25-OH vitamin D(2) in human serum. Hexadeuterium labelled 25-OH vitamin D(3) internal standard (IS) was added to calibrators (prepared in phosphate-buffered saline with 60 g/L albumin), controls or patient sera and 25-OH vitamin D metabolites were released from vitamin D binding protein by adding sodium hydroxide prior to protein precipitation by acetonitrile/methanol (9:1, v/v). Subsequent off-line solid-phase extraction was followed by chromatographic separation on a C-18 column using a water/methanol/ammonium acetate gradient. Detection was by Atmospheric Pressure Electrospray Ionisation (AP-EI) followed by selected reaction monitoring. We compared the LC-MS/MS assay to the DiaSorin radioimmunoassay (RIA) and a recently re-standardised version of an automated electrochemi-luminescent immunoassay (ECLIA) from Roche Diagnostics. We also analysed external quality control samples from the International Vitamin D External Quality Assessment Scheme (DEQAS) for comparison with other participating laboratories using LC-MS. The method was linear from 5 to at least 550 nmol/L with intra- and interday CV's <= 6% for both 25-OH vitamin D(3) and 25-OH vitamin D(2). Recoveries ranged between 94.9 and 106.9% for 25-OH vitamin D(3) and 82.7 and 100.3% for 25-OH vitamin D(2). Our results for the DEQAS serum pools averaged 7.2% from the overall LC-MS method mean. The DiaSorin RIA agreed well with the LC-MS/MS method (r(2) = 0.90; average bias 1.61 nmol/L), the Roche ECLIA considerably disagreed (r2 = 0.58: bias 10.13 nmol/L). This LC-MS/MS method is reliable and robust for the measurement of both 25-OH vitamin D(3) and 25-OH vitamin D(2) in human serum. (C) 2010 Elsevier B.V. All rights reserved.
引用
收藏
页码:1163 / 1168
页数:6
相关论文
共 13 条
[1]  
[Anonymous], EP10A3 CLIN LAB STAN
[2]   25-Hydroxyvitamin D Assays: The Quest For Accuracy [J].
Carter, Graham D. .
CLINICAL CHEMISTRY, 2009, 55 (07) :1300-1302
[3]   Use of a common standard improves the performance of liquid chromatography-tandem mass spectrometry methods for serum 25-hydroxyvitamin-D [J].
Carter, Graham D. ;
Jones, Julia C. .
ANNALS OF CLINICAL BIOCHEMISTRY, 2009, 46 :79-81
[4]   Measurement of 25-hydroxyvitamin D3 (25OHD3) and 25-hydroxyvitamin D2 (25OHD2) in human serum using liquid chromatography-tandem mass spectrometry and its comparison to a radioimmunoassay method [J].
Chen, Huiping ;
McCoy, Leslie F. ;
Schleicher, Rosemary L. ;
Pfeiffer, Christine M. .
CLINICA CHIMICA ACTA, 2008, 391 (1-2) :6-12
[5]   Vitamin D deficiency [J].
Holick, Michael F. .
NEW ENGLAND JOURNAL OF MEDICINE, 2007, 357 (03) :266-281
[6]   A simple automated solid-phase extraction procedure for measurement of 25-hydroxyvitamin D3 and D2 by liquid chromatography-tandem mass spectrometry [J].
Knox, Susan ;
Harris, John ;
Calton, Lisa ;
Wallace, A. Michael .
ANNALS OF CLINICAL BIOCHEMISTRY, 2009, 46 :226-230
[7]   Automated Measurement of 25-OH Vitamin D3 on the Roche Modular E170 Analyzer [J].
Leino, Aila ;
Turpeinen, Ursula ;
Koskinen, Pertti .
CLINICAL CHEMISTRY, 2008, 54 (12) :2059-2062
[8]   Routine isotope-dilution liquid chromatography-tandem mass spectrometry assay for simultaneous measurement of the 25-hydroxy metabolites of vitamins D2 and D3 [J].
Maunsell, Z ;
Wright, DJ ;
Rainbow, SJ .
CLINICAL CHEMISTRY, 2005, 51 (09) :1683-1690
[9]   Accuracy and clinical implications of seven 25-hydroxyvitamin D methods compared with liquid chromatography-tandem mass spectrometry as a reference [J].
Roth, Heinz Juergen ;
Schmidt-Gayk, Heinrich ;
Weber, Holger ;
Niederau, Christoph .
ANNALS OF CLINICAL BIOCHEMISTRY, 2008, 45 :153-159
[10]   Quantification of serum 25-hydroxyvitamin D2 and D3 using HPLC-tandem mass spectrometry and examination of reference intervals for diagnosis of vitamin D deficiency [J].
Saenger, Amy K. ;
Laha, Thomas J. ;
Bremner, Deborah E. ;
Sadrzadeh, Sayed M. H. .
AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 2006, 125 (06) :914-920