A review of Doppler ultrasound quality assurance protocols and test devices

被引:35
作者
Browne, Jacinta E. [1 ,2 ]
机构
[1] Dublin Inst Technol, Sch Phys, Med Ultrasound Grp, Dublin 8, Ireland
[2] Dublin Inst Technol, IEO, FOCAS Inst, Dublin 8, Ireland
来源
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS | 2014年 / 30卷 / 07期
关键词
Doppler ultrasound quality assurance; Doppler test devices; Flow phantoms; String phantoms and Doppler ultrasound performance; BLOOD-MIMICKING FLUID; REAL VESSEL PHANTOM; SPECTRAL DOPPLER; STRING PHANTOM; FLOW PHANTOMS; TEST OBJECT; MOVING TARGET; STEADY FLOW; PERFORMANCE TESTS; CAROTID-ARTERY;
D O I
10.1016/j.ejmp.2014.08.003
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
100231 [临床病理学]; 100902 [航空航天医学];
摘要
In this paper, an overview of Doppler ultrasound quality assurance (QA) testing will be presented in three sections. The first section will review the different Doppler ultrasound parameters recommended by professional bodies for use in QA protocols. The second section will include an evaluation and critique of the main test devices used to assess Doppler performance, while the final section of this paper will discuss which of the wide range of test devices have been found to be most suitable for inclusion in Doppler QA programmes. Pulsed Wave Spectral Doppler, Colour Doppler Imaging QA test protocols have been recommended over the years by various professional bodies, including the UK's Institute of Physics and Engineering in Medicine (IPEM), the American Institute for Ultrasound in Medicine (AIUM), and the International Electrotechnical Commission (IEC). However, despite the existence of such recommended test protocols, very few commercial or research test devices exist which can measure the full range of both PW Doppler ultrasound and colour Doppler imaging performance parameters, particularly quality control measurements such as: (i) Doppler sensitivity (ii) colour Doppler spatial resolution (iii) colour Doppler temporal resolution (iv) colour Doppler velocity resolution (v) clutter filter performance and (vi) tissue movement artefact suppression. In this review, the merits of the various commercial and research test devices will be considered and a summary of results obtained from published studies which have made use of some of these Doppler test devices, such as the flow, string, rotating and belt phantom, will be presented. (C) 2014 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:742 / 751
页数:10
相关论文
共 71 条
[1]
*AIUM, 2002, PERF CRIT MEAS DOPPL
[2]
[Anonymous], 1994, 70 IPSM
[3]
A DOPPLER SIGNAL SIMULATOR [J].
BASTOS, CAC ;
FISH, PJ .
CLINICAL PHYSICS AND PHYSIOLOGICAL MEASUREMENT, 1991, 12 (02) :177-183
[4]
PERFORMANCE TESTS OF DOPPLER ULTRASOUND EQUIPMENT WITH A TISSUE AND BLOOD-MIMICKING PHANTOM [J].
BOOTE, EJ ;
ZAGZEBSKI, JA .
JOURNAL OF ULTRASOUND IN MEDICINE, 1988, 7 (03) :137-147
[5]
Colour Doppler Spatial Resolution Performance Testing [J].
Brown, Jacinta E. ;
Brown, Ian ;
Hoskins, Peter R. ;
Watson, Amanda J. ;
Elliott, Alex T. .
ULTRASOUND, 2007, 15 (03) :162-166
[6]
Validation of a sensitivity performance index test protocol and evaluation of colour Doppler sensitivity for a range of ultrasound scanners [J].
Browne, JE ;
Watson, AJ ;
Hoskins, PR ;
Elliott, AT .
ULTRASOUND IN MEDICINE AND BIOLOGY, 2004, 30 (11) :1475-1483
[7]
Assessment of the acoustic properties of common tissue-mimicking test phantoms [J].
Browne, JE ;
Ramnarine, KV ;
Watson, AJ ;
Hoskins, PR .
ULTRASOUND IN MEDICINE AND BIOLOGY, 2003, 29 (07) :1053-1060
[8]
Comission IE, 2001, 61685 COM IE
[9]
An audit of a hospital-based Doppler ultrasound quality control protocol using a commercial string Doppler phantom [J].
Cournane, S. ;
Fagan, A. J. ;
Browne, J. E. .
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2014, 30 (03) :380-384
[10]
A real vessel phantom for flow imaging: 3-D Doppler ultrasound of steady flow [J].
Dabrowski, W ;
Dunmore-Buyze, J ;
Cardinal, HN ;
Fenster, A .
ULTRASOUND IN MEDICINE AND BIOLOGY, 2001, 27 (01) :135-141