Unusual cause for an increase of the sensing integrity counter in a patient with inappropriate implantable cardioverter-defibrittator therapy

被引:5
作者
Vollmann, Dirk [1 ]
Luethje, Lars [1 ]
Zabel, Markus [1 ]
机构
[1] Klinikum Univ Gottingen, Abt Kardiol & Pneumol, Herzzentrum, D-37075 Gottingen, Germany
来源
EUROPACE | 2007年 / 9卷 / 05期
关键词
implantable-cardioverter defibrillator; oversensing; inappropriate therapy; sensing integrity counter; DEFIBRILLATOR LEAD FAILURE;
D O I
10.1093/europace/eum028
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
We describe the case of a patient who presented with multiple implantable cardioverter-defibrillator (ICD) shock discharges 12 months after device implantation. Upon device interrogation, intermittent oversensing of electrical noise and potential ICD lead failure were suggested by a significant increase in the sensing integrity counter (SIC), a cumulative count of very short ventricular sensed intervals. Analysis of stored episodes, however, revealed that inappropriate ICD therapy had been caused by intermittent T-wave oversensing (TWO), and that the increase of the SIC resulted from the coincidence of TWO and premature ventricular complexes (PVCs). T-wave oversensing resolved and the SIC did not increase any more during follow-up after adjustment of ventricular sensitivity. The coincidence of TWO and PVCs should therefore be considered as an uncommon cause for short ventricular sensed intervals in ICD patients presenting with a suspect increase in the SIC.
引用
收藏
页码:275 / 277
页数:3
相关论文
共 10 条
[1]   Patient alert in implantable cardioverter defibrillators: Toy or tool? [J].
Becker, R ;
Ruf-Richter, J ;
Senges-Becker, JC ;
Bauer, A ;
Weretka, S ;
Voss, F ;
Katus, HA ;
Schoels, W .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2004, 44 (01) :95-98
[2]   Detection and management of an implantable cardioverter defibrillator lead failure - Incidence and clinical implications [J].
Ellenbogen, KA ;
Wood, MA ;
Shepard, RK ;
Clemo, HF ;
Vaughn, T ;
Holloman, K ;
Dow, M ;
Leffler, J ;
Abeyratne, A ;
Verness, D .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2003, 41 (01) :73-80
[3]  
GUNDERSON B, 2000, PACING CLIN ELECTROP, V23, P591
[4]   An algorithm to predict implantable cardioverter-defibrillator lead failure [J].
Gunderson, BD ;
Patel, AS ;
Bounds, CA ;
Shepard, RK ;
Wood, MA ;
Ellenbogen, KA .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2004, 44 (09) :1898-1902
[5]   Adverse events with transvenous implantable cardioverter-defibrillators - A prospective multicenter study [J].
Rosenqvist, M ;
Beyer, T ;
Block, M ;
den Dulk, K ;
Minten, J ;
Lindemans, F .
CIRCULATION, 1998, 98 (07) :663-670
[6]   Inappropriate arrhythmia detection in implantable defibrillator therapy due to oversensing of diaphragmatic myopotentials [J].
Schulte, B ;
Sperzel, J ;
Carlsson, J ;
Dürsch, M ;
Erdogan, A ;
Pitschner, HF ;
Neuzner, J .
JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 2001, 5 (04) :487-493
[7]   Advanced ICD troubleshooting: Part I [J].
Swerdlow, CD ;
Friedman, PA .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2005, 28 (12) :1322-1346
[8]   Patient Alert™ to detect ICD lead failure:: efficacy, limitations, and implications for future algorithms [J].
Vollmann, D ;
Erdogan, A ;
Himmrich, E ;
Neuzner, J ;
Becker, D ;
Unterberg-Buchwald, C ;
Sperzel, J .
EUROPACE, 2006, 8 (05) :371-376
[9]   Inappropriate therapy and fatal proarrhythmia by an implantable cardioverter-defibrittator [J].
Vollmann, D ;
Lüthje, L ;
Vonhof, S ;
Unterberg, C .
HEART RHYTHM, 2005, 2 (03) :307-309
[10]   Ventricular oversensing: A study of 101 patients implanted with dual chamber defibrillators and two different lead systems [J].
Weretka, S ;
Michaelsen, J ;
Becker, R ;
Karle, F ;
Voss, F ;
Hilbel, T ;
Osswald, BR ;
Bahner, ML ;
Senges, JC ;
Kuebler, W ;
Schoels, W .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2003, 26 (01) :65-70