Miniature high frequency focused ultrasonic transducers for minimally invasive imaging procedures

被引:30
作者
Fleischman, A [1 ]
Modi, R [1 ]
Nair, A [1 ]
Talman, J [1 ]
Lockwood, G [1 ]
Roy, S [1 ]
机构
[1] Cleveland Clin Fdn, Dept Biomed Engn, Cleveland, OH 44195 USA
关键词
ultrasonic transducer; polymer transducer; PVDF; high frequency imaging; MEMS; minimally invasive procedures;
D O I
10.1016/S0924-4247(02)00323-0
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Spherically focused, high frequency (30-50 MHz) ultrasonic imaging transducers are fabricated using poly(vinylidene) difluoride (PVDF) film and a membrane deflection technique that is compatible with CMOS microelectronics fabrication. Micromachined ultrasonic transducers with 0.75-2.00 mm-diameter apertures and f-numbers of 1.30-2.45 are fabricated and characterized. The transducers exhibit focused radiation patterns with 50 mum axial resolution and 51-92 mum lateral resolution and bandwidths of 80-110% around center frequency values. A miniature preamplifier hybrid circuit, which is designed to be mounted within the transducer housing, is fabricated and characterized to reveal a low noise level of 5.3 nV/rootH(z). Tissue imaging capabilities of the micromachined ultrasonic transducers are demonstrated through successful imaging of human cadaveric aorta. (C) 2003 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:76 / 82
页数:7
相关论文
共 15 条
[1]   A P(VDF-TRFE)-BASED INTEGRATED ULTRASONIC TRANSDUCER [J].
FIORILLO, AS ;
VANDERSPIEGEL, J ;
BLOOMFIELD, PE ;
ESMAILZANDI, D .
SENSORS AND ACTUATORS A-PHYSICAL, 1990, 22 (1-3) :719-725
[2]   Focused high frequency ultrasonic transducers for minimally invasive imaging [J].
Fleischman, A ;
Modi, R ;
Nair, A ;
Lock-Wood, G ;
Roy, S .
FIFTEENTH IEEE INTERNATIONAL CONFERENCE ON MICRO ELECTRO MECHANICAL SYSTEMS, TECHNICAL DIGEST, 2002, :300-303
[3]   A history of medical and biological imaging with polyvinylidene fluoride (PVDF) transducers [J].
Foster, FS ;
Harasiewicz, EA ;
Sherar, MD .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2000, 47 (06) :1363-1371
[4]   Fabrication of dome-shaped diaphragm with circular clamped boundary on silicon substrate [J].
Han, CH ;
Kim, ES .
MEMS '99: TWELFTH IEEE INTERNATIONAL CONFERENCE ON MICRO ELECTRO MECHANICAL SYSTEMS, TECHNICAL DIGEST, 1999, :505-510
[5]   THE DESIGN OF PROTECTION CIRCUITRY FOR HIGH-FREQUENCY ULTRASOUND IMAGING-SYSTEMS [J].
LOCKWOOD, GR ;
HUNT, JW ;
FOSTER, FS .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 1991, 38 (01) :48-55
[6]   Beyond 30 MHz - Applications of high-frequency ultrasound imaging [J].
Lockwood, GR ;
Turnbull, DH ;
Christopher, DA ;
Foster, FS .
IEEE ENGINEERING IN MEDICINE AND BIOLOGY MAGAZINE, 1996, 15 (06) :60-71
[7]   Miniature polymer transducers for high frequency medical imaging [J].
Lockwood, GR ;
Hazard, CR .
ULTRASONIC TRANSDUCER ENGINEERING: MEDICAL IMAGING 1998, 1998, 3341 :228-236
[8]   Quantification of intramural calcification in coronary intravascular ultrasound images with automated image analysis [J].
Masseroli, M ;
Cothren, RM ;
Meier, DS ;
Tuzcu, M ;
Vince, G ;
Nissen, SE ;
Thomas, JD ;
Cornhill, JF .
AMERICAN HEART JOURNAL, 1998, 136 (01) :78-86
[9]   MICROMACHINING FOR IMPROVEMENT OF INTEGRATED ULTRASONIC TRANSDUCER SENSITIVITY [J].
MO, JH ;
ROBINSON, AL ;
FITTING, DW ;
TERRY, FL ;
CARSON, PL .
IEEE TRANSACTIONS ON ELECTRON DEVICES, 1990, 37 (01) :134-140
[10]  
MOTCHENBACHER CD, 1973, LOW NOISE ELECT DESI