Axonal stimulation under MRI magnetic field z gradients: A modeling study

被引:13
作者
Carbunaru, R [1 ]
Durand, DM [1 ]
机构
[1] CASE WESTERN RESERVE UNIV,DEPT BIOMED ENGN,CLEVELAND,OH 44106
关键词
MRI; magnetic stimulation; magnetic field gradient; axon stimulation;
D O I
10.1002/mrm.1910380511
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The stimulation of axons under MRI magnetic fields is analyzed solving the cable equation along the axons in the presence of magnetic field z gradients, Axons are represented using a one-dimensional compartmental cable model including the kinetics of mammalian myelinated fibers, Computer simulations of the model were performed for sinusoidal and trapezoidal fields, Several axon and field parameters were tested to determine the threshold values for axonal stimulation of the magnetic field, induced electric field and time derivative of the magnetic field, The results indicate that 1) threshold for stimulation is lowest for the largest diameter axons terminating in the region where the magnetic field is maximum, 2) trapezoidal waveforms can be optimized to allow better sub-threshold resolution than sinusoidal waveforms, and 3) the induced electric field is better than the magnetic field and time derivative of the magnetic field as indicators of stimulation threshold.
引用
收藏
页码:750 / 758
页数:9
相关论文
共 25 条
[1]   FOCAL STIMULATION OF HUMAN CEREBRAL-CORTEX WITH THE MAGNETIC COIL - A COMPARISON WITH ELECTRICAL-STIMULATION [J].
AMASSIAN, VE ;
CRACCO, RQ ;
MACCABEE, PJ .
ELECTROENCEPHALOGRAPHY AND CLINICAL NEUROPHYSIOLOGY, 1989, 74 (06) :401-416
[2]  
BARKER AT, 1987, NEUROSURGERY, V20, P100
[3]   THE ACTIVATING FUNCTION FOR MAGNETIC STIMULATION DERIVED FROM A 3-DIMENSIONAL VOLUME CONDUCTOR MODEL [J].
BASSER, PJ ;
WIJESINGHE, RS ;
ROTH, BJ .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1992, 39 (11) :1207-1210
[4]   PHYSIOLOGICAL-EFFECTS OF FAST OSCILLATING MAGNETIC-FIELD GRADIENTS [J].
BUDINGER, TF ;
FISCHER, H ;
HENTSCHEL, D ;
REINFELDER, HE ;
SCHMITT, F .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1991, 15 (06) :909-914
[5]  
CARBUNARU R, 1996, P 4 ANN M INT SOC MA, P1768
[6]   SENSORY STIMULATION BY TIME-VARYING MAGNETIC-FIELDS [J].
COHEN, MS ;
WEISSKOFF, RM ;
RZEDZIAN, RR ;
KANTOR, HL .
MAGNETIC RESONANCE IN MEDICINE, 1990, 14 (02) :409-414
[7]  
FAIERSTEIN RC, 1996, THESIS CASE W RESERV
[8]   ACTION POTENTIAL IN MYELINATED NERVE FIBRE OF XENOPUS LAEVIS AS COMPUTED ON BASIS OF VOLTAGE CLAMP DATA [J].
FRANKENHAEUSER, B ;
HUXLEY, AF .
JOURNAL OF PHYSIOLOGY-LONDON, 1964, 171 (02) :302-&
[9]   AVOIDING PERIPHERAL-NERVE STIMULATION - GRADIENT WAVE-FORM CRITERIA FOR OPTIMUM RESOLUTION IN ECHO-PLANAR IMAGING [J].
HARVEY, PR ;
MANSFIELD, P .
MAGNETIC RESONANCE IN MEDICINE, 1994, 32 (02) :236-241
[10]  
HINES ML, 1992, NEURON PROGRAM SIMUL