A new high resolution color flow system using an eigendecomposition-based adaptive filter for clutter rejection (vol 49, pg 1384, 2002)

被引:27
作者
Kruse, DE [1 ]
Ferrara, KW [1 ]
机构
[1] Univ Calif Davis, Dept Biomed Engn, Davis, CA 95616 USA
关键词
D O I
10.1109/TUFFC.2002.1159852
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
We present a new signal processing strategy for high frequency color flow mapping in moving tissue environments. A new application of an eigendecomposition-based clutter rejection filter is presented with modifications to deal with high blood-to-clutter ratios (BCR). Additionally, a new method for correcting blood velocity estimates with an estimated tissue motion profile is detailed. The performance of the clutter filter and velocity estimation strategies is quantified using a new swept-scan signal model. In vivo color flow images are presented to illustrate the potential of the system for mapping blood flow in the microcirculation with external tissue motion.
引用
收藏
页码:1739 / 1754
页数:16
相关论文
共 31 条
[1]   Isomorphism between pulsed-wave Doppler ultrasound and direction-of-arrival estimation .1. Experimental results [J].
Allam, ME ;
Kinnick, RR ;
Greenleaf, JF .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 1996, 43 (05) :923-935
[2]   Isomorphism between pulsed-wave Doppler ultrasound and direction-of-arrival estimation .1. Basic principles [J].
Allam, ME ;
Greenleaf, JF .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 1996, 43 (05) :911-922
[3]  
[Anonymous], 1989, Modern Filters
[4]  
BJAERUM S, 1998, P IEEE ULTRASON S, V2, P1571
[5]  
BJAERUM S, 1999, P IEEE ULTR S, V2, P1503
[6]  
BJAERUM S, 1997, P IEEE ULTR S, V2, P1223
[7]  
DEMMEL J, 2000, TEMPLATES SOLUTION A, P39
[8]   Estimation of blood velocity with high frequency ultrasound [J].
Ferrara, KW ;
Zagar, BG ;
SokilMelgar, JB ;
Silverman, RH ;
Aslanidis, IM .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 1996, 43 (01) :149-157
[9]   EFFECT OF THE BEAM-VESSEL ANGLE ON THE RECEIVED ACOUSTIC-SIGNAL FROM BLOOD [J].
FERRARA, KW .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 1995, 42 (03) :416-428
[10]  
FOSTER FS, 1994, P IEEE ULTR S, V3, P1599