In vivo alanine/EPR dosimetry in daily clinical practice:: A feasibility study

被引:41
作者
Ciesielski, B
Schultka, K
Kobierska, A
Nowak, R
Peimel-Stuglik, Z
机构
[1] Med Univ Gdansk, Dept Phys & Biophys, PL-80211 Gdansk, Poland
[2] Med Univ Gdansk, Dept Radiotherapy & Oncol, PL-80211 Gdansk, Poland
[3] Inst Nucl Chem & Technol, Lab Measurments Technol Doses, PL-03195 Warsaw, Poland
来源
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS | 2003年 / 56卷 / 03期
关键词
alanine; EPR; dosimetry; radiotherapy;
D O I
10.1016/S0360-3016(03)00196-2
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: The objective of this study was evaluation of accuracy of in vivo dosimetry using electron paramagnetic resonance (EPR) in alanine. Additionally, we aimed to identify sources of uncertainty in dose determination and quantitative assessment of physical factors that may result in discrepancies between the measured and planned single-fraction doses. Methods and Materials: The measurements were performed using detectors in a form of 1.6 cm X 1.6 cm polyethylene sachets filled with powdered L-alanine. The detectors were taped to the patient's skin and measured the entrance doses for Co-60 and electron beams. Some detectors were covered with buildup material, and some measured the "skin dose." The EPR measurements were performed with a Varian E-4 spectrometer. Results: The calculated uncertainty of EPR measured doses was dependent on measured doses and varied from 6.6% for 0.5 Gy to 3.2% for 2 Gy. The calculated uncertainty was in concordance with experimentally determined reproducibility of EPR signals. However, the deviations between measured and planned doses exceeded the uncertainty range of EPR measurements, which can be attributed to uncertainty in determination of actually delivered doses to the detectors, on the basis of treatment planning data. Conclusion: The accuracy of dose determination by EPR measurements was shown to be achievable within the 5% limit recommended by the ICRU for doses above 0.7 Gy. The accuracy of in vivo verification of radiotherapy doses by in vivo EPR dosimetry can be improved by meticulous selection of measurement conditions, i.e., radiation fields and detector positions, ensuring accurate calculation of doses delivered to the dosimeters. (C) 2003 Elsevier Inc.
引用
收藏
页码:899 / 905
页数:7
相关论文
共 38 条
[11]   THEORETICAL AND EXPERIMENTAL RADIATION EFFECTIVENESS OF THE FREE-RADICAL DOSIMETER ALANINE TO IRRADIATION WITH HEAVY CHARGED-PARTICLES [J].
HANSEN, JW ;
OLSEN, KJ .
RADIATION RESEARCH, 1985, 104 (01) :15-27
[12]   A high sensitivity EPR technique for alanine dosimetry [J].
Haskell, EH ;
Hayes, RB ;
Kenner, GH .
RADIATION PROTECTION DOSIMETRY, 1998, 77 (1-2) :43-49
[13]   Metal ion-organic compound for high sensitive ESR dosimetry [J].
Hassan, GM ;
Ikeya, M .
APPLIED RADIATION AND ISOTOPES, 2000, 52 (05) :1247-1254
[14]   PHOTON MASS ATTENUATION AND ENERGY-ABSORPTION COEFFICIENTS FROM 1 KEV TO 20 MEV [J].
HUBBELL, JH .
INTERNATIONAL JOURNAL OF APPLIED RADIATION AND ISOTOPES, 1982, 33 (11) :1269-1290
[15]   Strategy for finding new materials for ESR dosimeters [J].
Ikeya, M ;
Hassan, GM ;
Sasaoka, H ;
Kinoshita, Y ;
Takaki, S ;
Yamanaka, C .
APPLIED RADIATION AND ISOTOPES, 2000, 52 (05) :1209-1215
[16]  
International Commission in Radiation Units and Measurements, 1984, 37 ICRU
[17]  
International Commission on Radiation Units and Measurements - ICRU, 1976, 24 ICRU
[18]   COMPREHENSIVE QA FOR RADIATION ONCOLOGY - REPORT OF AAPM RADIATION-THERAPY COMMITTEE TASK-GROUP-40 [J].
KUTCHER, GJ ;
COIA, L ;
GILLIN, M ;
HANSON, WF ;
LEIBEL, S ;
MORTON, RJ ;
PALTA, JR ;
PURDY, JA ;
REINSTEIN, LE ;
SVENSSON, GK ;
WELLER, M ;
WINGFIELD, L .
MEDICAL PHYSICS, 1994, 21 (04) :581-618
[19]   ASSESSMENT OF DOSE INHOMOGENEITY AT TARGET LEVEL BY INVIVO DOSIMETRY - CAN THE RECOMMENDED 5-PERCENT ACCURACY IN THE DOSE DELIVERED TO THE TARGET VOLUME BY FULFILLED IN DAILY PRACTICE [J].
LEUNENS, G ;
VERSTRAETE, J ;
DUTREIX, A ;
VANDERSCHUEREN, E .
RADIOTHERAPY AND ONCOLOGY, 1992, 25 (04) :242-250
[20]  
Lobodziec W, 1998, REP PRACT ONCOL RADI, V3, P25