摘要
Background Coronary flow velocity reserve (CFVR), defined as the ratio of maximal hyperaemic to baseline flow velocity, has been validated as a marker of physiological significance of a coronary lesion. Clinically, this parameter is measured invasively during X-ray angiography using the Doppler guide wire. With magnetic resonance (MR) imaging it is possible to quantify CFVR non-invasively. Design The purpose of the study was to compare CFVR, acquired with MR imaging and the Doppler guide wire in patients with coronary artery disease. Methods Twenty-two patients suffering from one- or two-vessel coronary artery disease as derived from diagnostic X-ray coronary angiography were included. Coronary flow velocity reserve was measured at baseline and during maximal hyperaemia, obtained by intravenous administration of adenosine using MR phase contrast velocity quantification. Within 2 weeks CFVR was measured invasively with a Doppler guide wire. Results In 26 coronary arteries CFVR was acquired with both techniques. Mean CFVR in the stenosed and healthy reference arteries was 1.5 +/- 0.7 and 2.7 +/- 1.0 (P < 0.01) respectively for MR measurements and 1.9 +/- 0.7 and 3.1 +/- 0.6 (P < 0.01) respectively for Doppler measurements. Bland-Altman analysis revealed a non-significant mean difference between the two techniques of 0.4 +/- 1.2. Conclusion In a selected group of stable patients with coronary artery disease MR flow velocity quantification provides non-invasive data equivalent to the invasive Doppler guide wire data. Variability in both the MR and Doppler ultrasound measurement resulted in a significant scatter of data without systematic difference.
引用
收藏
页码:365 / 372
页数:8
相关论文
共 38 条
  • [1] Accuracy of segmented MR velocity mapping to measure small vessel pulsatile flow in a phantom simulating cardiac motion
    Arheden, H
    Saeed, M
    Törnqvist, E
    Lund, G
    Wendland, MF
    Higgins, CB
    Ståhlberg, F
    [J]. JOURNAL OF MAGNETIC RESONANCE IMAGING, 2001, 13 (05) : 722 - 728
  • [2] QUANTITATIVE CORONARY ANGIOGRAPHY IN THE ESTIMATION OF THE FUNCTIONAL-SIGNIFICANCE OF CORONARY STENOSIS - CORRELATIONS WITH DOBUTAMINE-ATROPINE STRESS TEST
    BAPTISTA, J
    ARNESE, M
    ROELANDT, JRTC
    FIORETTI, P
    KEANE, D
    ESCANED, J
    BOERSMA, E
    DIMARIO, C
    SERRUYS, PW
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1994, 23 (06) : 1434 - 1439
  • [3] Velocity-encoded, phase-difference cine MRI measurements of coronary artery flow: Dependence of flow accuracy on the number of cine frames
    Clarke, GD
    Hundley, WG
    McColl, RW
    Eckels, R
    Smith, D
    Chaney, C
    Li, HF
    Peshock, RM
    [J]. JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1996, 6 (05): : 733 - 742
  • [4] MEASUREMENT OF ABSOLUTE EPICARDIAL CORONARY-ARTERY FLOW AND FLOW RESERVE WITH BREATH-HOLD CINE PHASE-CONTRAST MAGNETIC-RESONANCE-IMAGING
    CLARKE, GD
    ECKELS, R
    CHANEY, C
    SMITH, D
    DITTRICH, J
    HUNDLEY, WG
    NESSAIVER, M
    LI, HF
    PARKEY, RW
    PESHOCK, RM
    [J]. CIRCULATION, 1995, 91 (10) : 2627 - 2634
  • [5] Coronary flow and coronary flow reserve measurements in humans with breath-held magnetic resonance phase contrast velocity mapping
    Davis, CP
    Liu, PF
    Hauser, M
    Gohde, SC
    vonSchulthess, GK
    Debatin, JF
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1997, 37 (04) : 537 - 544
  • [6] VALIDATION OF A DOPPLER GUIDE WIRE FOR INTRAVASCULAR MEASUREMENT OF CORONARY-ARTERY FLOW VELOCITY
    DOUCETTE, JW
    CORL, PD
    PAYNE, HM
    FLYNN, AE
    GOTO, M
    NASSI, M
    SEGAL, J
    [J]. CIRCULATION, 1992, 85 (05) : 1899 - 1911
  • [7] FLOW VELOCITY QUANTIFICATION IN HUMAN CORONARY-ARTERIES WITH FAST, BREATH-HOLD MR-ANGIOGRAPHY
    EDELMAN, RR
    MANNING, WJ
    GERVINO, E
    LI, W
    [J]. JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1993, 3 (05): : 699 - 703
  • [8] CORONARY-ARTERIES - BREATH-HOLD MR ANGIOGRAPHY
    EDELMAN, RR
    MANNING, WJ
    BURSTEIN, D
    PAULIN, S
    [J]. RADIOLOGY, 1991, 181 (03) : 641 - 643
  • [9] RELATION BETWEEN GEOMETRIC DIMENSIONS OF CORONARY-ARTERY STENOSES AND MYOCARDIAL PERFUSION RESERVE IN MAN
    GOLDSTEIN, RA
    KIRKEEIDE, RL
    DEMER, LL
    MERHIGE, M
    NISHIKAWA, A
    SMALLING, RW
    MULLANI, NA
    GOULD, KL
    [J]. JOURNAL OF CLINICAL INVESTIGATION, 1987, 79 (05) : 1473 - 1478
  • [10] PHYSIOLOGIC BASIS FOR ASSESSING CRITICAL CORONARY STENOSIS - INSTANTANEOUS FLOW RESPONSE AND REGIONAL DISTRIBUTION DURING CORONARY HYPEREMIA AS MEASURES OF CORONARY FLOW RESERVE
    GOULD, KL
    LIPSCOMB, K
    HAMILTON, GW
    [J]. AMERICAN JOURNAL OF CARDIOLOGY, 1974, 33 (01) : 87 - 94