Detection of right-to-left shunt with ultrasound contrast agent and transcranial Doppler sonography

被引:325
作者
Jauss, M [1 ]
Zanette, E [1 ]
机构
[1] Univ Giessen, Dept Neurol, D-35385 Giessen, Germany
关键词
right-to-left shunt; ultrasound contrast agent; transcranial Doppler sonography; patent foramen ovale; valsalva maneuver;
D O I
10.1159/000016119
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
An international Consensus Meeting to determine a standard in the examination technique for the detection of right-to-left shunt (RLS) using contrast transcranial Doppler sonography (TCD) led to the following recommendations to standardize the examination procedure: The patient should be prepared with an 18-gauge needle inserted into the cubital vein and should be in the supine position. Insonation of at least one middle cerebral artery (MCA) using TCD is performed. The contrast agent is prepared using 9 mi isotonic saline solution and 1 mi air mixed with a three-way stopcock by exchange of saline! air mixture between the syringes and injected as a bolus. In case of little or no detection of microbubbles (MB) in the MCA under basal conditions, the examination will be repeated using the Valsalva maneuver (VM), Contrast agent will be injected 5 s before the start of the VM; the overall VM duration should be 10 s. The patient should start the VM on examiner's command. The strength of the VM can be controlled by peak flow velocity of the Doppler curve. The time when the first MB appears at the MCA level wilt be noted. A four-level categorization according to the MB count should be applied: (1) 0 MB (negative result); (2) 1-10 MB; (3)>10 MB and no curtain, and (4) curtain. ('Curtain' refers to a shower of MB, where a single bubble cannot be identified.) The results should be documented for basal condition and VM testing separately. The clinical significance of the diagnosis of a RLS in a particular patient is not fully evaluated and requires further studies. A minimum amount of MB suggestive of a clinical relevant RLS is not yet established. It probably depends on interindividual differences in hemodynamics that are currently not fully understood. Transesophageal echocardiography remains the gold standard for detection of a patent foramen ovate or an atrial septum defect. However, TCD with a contrast agent has been turned out as a potential method to diagnose a RLS in several studies which have been published during the last years, and a RLS other than at the atrial level may be detected only by this method. Furthermore, the VM can be applied more comfortably and more reliably during Doppler examination than during transesophageal echocardiography. Copyright (C) 2000 S. Karger AG,Basel.
引用
收藏
页码:490 / 496
页数:7
相关论文
共 29 条
  • [1] VALIDATION OF TRANSCRANIAL DOPPLER SONOGRAPHY IN THE ASSESSMENT OF PATENT FORAMEN OVALE
    ANZOLA, GP
    RENALDINI, E
    MAGONI, M
    COSTA, A
    COBELLI, M
    GUINDANI, M
    [J]. CEREBROVASCULAR DISEASES, 1995, 5 (03) : 194 - 198
  • [2] TRANSCRANIAL DOPPLER ULTRASOUND IDENTIFIES PATIENTS WITH RIGHT-TO-LEFT CARDIAC OR PULMONARY SHUNTS
    CHIMOWITZ, MI
    NEMEC, JJ
    MARWICK, TH
    LORIG, RJ
    FURLAN, AJ
    SALCEDO, EE
    [J]. NEUROLOGY, 1991, 41 (12) : 1902 - 1904
  • [3] Complementarity of contrast transcranial Doppler and contrast transesophageal echocardiography for the detection of patent foramen ovale in stroke patients
    Devuyst, G
    Despland, PA
    Bogousslavsky, J
    Jeanrenaud, X
    [J]. EUROPEAN NEUROLOGY, 1997, 38 (01) : 21 - 25
  • [4] COMPARISON OF DIAGNOSTIC-TECHNIQUES FOR THE DETECTION OF A PATENT FORAMEN OVALE IN STROKE PATIENTS
    DITULLIO, M
    SACCO, RL
    VENKETASUBRAMANIAN, N
    SHERMAN, D
    MOHR, JP
    HOMMA, S
    [J]. STROKE, 1993, 24 (07) : 1020 - 1024
  • [5] Contrast transcranial Doppler ultrasound in the detection of right-to-left shunts - Comparison of different procedures and different contrast agents
    Droste, DW
    Kriete, JU
    Stypmann, J
    Castrucci, M
    Wichter, T
    Tietje, R
    Weltermann, B
    Young, P
    Ringelstein, EB
    [J]. STROKE, 1999, 30 (09) : 1827 - 1832
  • [6] Contrast transcranial Doppler ultrasound in the detection of right-to-left shunts -: Reproducibility, comparison of 2 agents, and distribution of microemboli
    Droste, DW
    Reisener, M
    Kemény, V
    Dittrich, R
    Schulte-Altedorneburg, G
    Stypmann, J
    Wichter, T
    Ringelstein, EB
    [J]. STROKE, 1999, 30 (05) : 1014 - 1018
  • [7] PSYCHOSOCIAL CONCOMITANTS TO REHABILITATION IN CHRONIC OBSTRUCTIVE PULMONARY-DISEASE .1. PSYCHOSOCIAL AND PSYCHOLOGICAL CONSIDERATIONS
    DUDLEY, DL
    GLASER, EM
    JORGENSON, BN
    LOGAN, DL
    [J]. CHEST, 1980, 77 (03) : 413 - 420
  • [8] CONSENSUS METHODS - CHARACTERISTICS AND GUIDELINES FOR USE
    FINK, A
    KOSECOFF, J
    CHASSIN, M
    BROOK, RH
    [J]. AMERICAN JOURNAL OF PUBLIC HEALTH, 1984, 74 (09) : 979 - 983
  • [10] Femoral injection of echo contrast medium may increase the sensitivity of testing for a patent foramen ovale
    Hamann, GF
    Schätzer-Klotz, D
    Fröhlig, G
    Strittmatter, M
    Jost, V
    Berg, G
    Stopp, M
    Schimrigk, K
    Schieffer, H
    [J]. NEUROLOGY, 1998, 50 (05) : 1423 - 1428