Relationship of blood pressure and pump flow in an implantable centrifugal blood pump during hypertension

被引:14
作者
Akimoto, T
Yamazaki, K
Litwak, P
Litwak, KN
Tagusari, O
Mori, T
Antaki, JF
Kameneva, MV
Watach, MJ
Umezu, M
Tomioka, J
Kormos, RL
Koyanagi, H
Griffith, BP
机构
[1] Univ Pittsburgh, Dept Surg, McGowan Ctr, Pittsburgh, PA 15219 USA
[2] Tokyo Womens Med Coll, Dept Cardiovasc Surg, Tokyo 162, Japan
[3] SunMed Technol Res Corp, Nagano, Japan
[4] Waseda Univ, Dept Mech Engn, Tokyo, Japan
关键词
D O I
10.1097/00002480-200009000-00018
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The purpose of this study was to evaluate the real time relationship between pump flow and pump differential pressure (D-P) during experimentally induced hypertension (HT). Two calves (80 and 68 kg) were implanted with the EVA-HEART centrifugal blood pump (SunMedical Technology Research Corp., Nagano, Japan) under general anesthesia. Blood pressure (BP) in diastole was increased to 100 mm Hg by norepinephrine to simulate HT. Pump flow, D-P, EGG, and BP were measured at pump speeds of 1,800, 2,100, and 2,300 rpm. All data were separated into systole and diastole, and pump flow during HT was compared with normotensive (NT) conditions at respective pump speeds. Diastolic BP was increased to 99.3 +/- 4.1 mm Hg from 66.5 +/- 4.4mm Hg (p < 0.01). D-P in systole was under 40 mm Hg (range of change was 10 to 40 mm Hg) even during HT. During NT, the average systolic pump flow volume was 60% of the total pump flow. However, during HT, the average systolic pump flow was 100% of total pump flow volume, although the pump flow volume in systole during HT decreased (33.1 +/- 5.7 vs 25.9 +/- 4.0 ml/systole, p < 0.01). In diastole, the average flow volume through the pump was 19.6 +/- 6.9 ml/diastole during NT and -2.2 +/- 11.1 ml/diastole during HT (p < 0.01). The change in pump flow volume due to HT, in diastole, was greater than the change in pump flow in systole at each pump speed (p < 0.001). This study suggests that the decrease of mean pump flow during HT is mainly due to the decrease of the diastolic pump flow and, to a much lesser degree, systolic pump flow.
引用
收藏
页码:596 / 599
页数:4
相关论文
共 11 条
[1]   Rotary blood pump flow spontaneously increases during exercise under constant pump speed: Results of a chronic study [J].
Akimoto, T ;
Yamazaki, K ;
Litwak, P ;
Litwak, KN ;
Tagusari, O ;
Mori, T ;
Antaki, JF ;
Kameneva, MV ;
Watach, MJ ;
Umezu, M ;
Tomioka, J ;
Kormos, RL ;
Koyanagi, H ;
Griffith, BP .
ARTIFICIAL ORGANS, 1999, 23 (08) :797-801
[2]   Development of the Nimbus/Pittsburgh axial flow left ventricular assist system [J].
Butler, K ;
Thomas, D ;
Antaki, J ;
Borovetz, H ;
Griffith, B ;
Kameneva, M ;
Kormos, R ;
Litwak, P .
ARTIFICIAL ORGANS, 1997, 21 (07) :602-610
[3]   IMPROVED MORTALITY AND REHABILITATION OF TRANSPLANT CANDIDATES TREATED WITH A LONG-TERM IMPLANTABLE LEFT-VENTRICULAR ASSIST SYSTEM [J].
FRAZIER, OH ;
ROSE, EA ;
MCCARTHY, P ;
BURTON, NA ;
TECTOR, A ;
LEVIN, H ;
KAYNE, HL ;
POIRIER, VL ;
DASSE, KA .
ANNALS OF SURGERY, 1995, 222 (03) :327-338
[4]   Effects of exercise during long-term support with a left ventricular assist device - Results of the experience with left ventricular assist device with exercise (EVADE) pilot trial [J].
Jaski, BE ;
Kim, J ;
Maly, RS ;
Branch, KR ;
Adamson, R ;
Favrot, LK ;
Smith, SC ;
Dembitsky, WP .
CIRCULATION, 1997, 95 (10) :2401-2406
[5]  
KONISHI H, 1994, T AM SOC ART INT ORG, V40, pM299
[6]   STRUCTURAL AND LEFT-VENTRICULAR HISTOLOGIC-CHANGES AFTER IMPLANTABLE LVAD INSERTION [J].
MCCARTHY, PM ;
NAKATANI, S ;
VARGO, R ;
KOTTKEMARCHANT, K ;
HARASAKI, H ;
JAMES, KB ;
SAVAGE, RM ;
THOMAS, JD .
ANNALS OF THORACIC SURGERY, 1995, 59 (03) :609-613
[7]  
NAKATANI T, 1995, ASAIO J, V41, pM526, DOI 10.1097/00002480-199507000-00066
[8]  
Nishimura K, 1997, ASAIO J, V43, pM553
[9]   Effect of pressure-flow relationship of centrifugal pump on in vivo hemodynamics: A consideration for design [J].
Tagusari, O ;
Yamazaki, K ;
Litwak, P ;
Antaki, JF ;
Watach, M ;
Gordon, LM ;
Kono, K ;
Mori, T ;
Koyanagi, H ;
Griffith, BP ;
Kormos, RL .
ARTIFICIAL ORGANS, 1998, 22 (05) :399-404
[10]   Development of a flow estimation and control system of an implantable centrifugal blood pump for circulatory assist [J].
Wakisaka, Y ;
Okuzono, Y ;
Taenaka, Y ;
Chikanari, K ;
Endo, S ;
Masuzawa, T ;
Takano, H .
ARTIFICIAL ORGANS, 1998, 22 (06) :488-492