Monolithic low-transition-temperature superconducting magnetometers for high resolution imaging magnetic fields of room temperature samples

被引:41
作者
Baudenbacher, F
Fong, LE
Holzer, JR
Radparvar, M
机构
[1] Vanderbilt Univ, Dept Phys & Astron, Nashville, TN 37235 USA
[2] Hypres Inc, Elmsford, NY 10523 USA
关键词
D O I
10.1063/1.1572968
中图分类号
O59 [应用物理学];
学科分类号
摘要
We have developed a monolithic low-temperature superconducting quantum interference device (SQUID) magnetometer and incorporated the device in a scanning microscope for imaging magnetic fields of room temperature samples. The instrument has a similar to100 mum spatial resolution and a 1.4 pT/Hz(1/2) field sensitivity above a few hertz. We discuss design constraints on and potential applications of the SQUID microscope. (C) 2003 American Institute of Physics.
引用
收藏
页码:3487 / 3489
页数:3
相关论文
共 14 条
[1]   High resolution imaging of biomagnetic fields generated by action currents in cardiac tissue using a LTS-SQUID microscope [J].
Baudenbacher, F ;
Peters, NT ;
Baudenbacher, P ;
Wikswo, JP .
PHYSICA C-SUPERCONDUCTIVITY AND ITS APPLICATIONS, 2002, 368 (1-4) :24-31
[2]   MAGNETIC MICROSCOPY USING A LIQUID-NITROGEN COOLED YBA2CU3O7 SUPERCONDUCTING QUANTUM INTERFERENCE DEVICE [J].
BLACK, RC ;
MATHAI, A ;
WELLSTOOD, FC ;
DANTSKER, E ;
MIKLICH, AH ;
NEMETH, DT ;
KINGSTON, JJ ;
CLARKE, J .
APPLIED PHYSICS LETTERS, 1993, 62 (17) :2128-2130
[3]   Magnetic microscopies: the new additions [J].
Dahlberg, ED ;
Proksch, R .
JOURNAL OF MAGNETISM AND MAGNETIC MATERIALS, 1999, 200 (1-3) :720-728
[4]   Low excess flux noise in YBa2Cu3O7-x dc SQUIDs cooled in static magnetic fields [J].
Dantsker, E ;
Tanaka, S ;
Nilsson, PA ;
Kleiner, R ;
Clarke, J .
IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 1997, 7 (02) :2772-2775
[5]   A 5-MHZ BANDWIDTH SQUID MAGNETOMETER WITH ADDITIONAL POSITIVE FEEDBACK [J].
DRUNG, D ;
MATZ, H ;
KOCH, H .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1995, 66 (04) :3008-3015
[6]   PLANAR COUPLING SCHEME FOR ULTRA LOW-NOISE DC SQUIDS [J].
JAYCOX, JM ;
KETCHEN, MB .
IEEE TRANSACTIONS ON MAGNETICS, 1981, 17 (01) :400-403
[7]   MINIATURE SQUID SUSCEPTOMETER [J].
KETCHEN, MB ;
KOPLEY, T ;
LING, H .
APPLIED PHYSICS LETTERS, 1984, 44 (10) :1008-1010
[8]   High-T-c SQUID microscope for room temperature samples [J].
Lee, TS ;
Chemla, YR ;
Dantsker, E ;
Clarke, J .
IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 1997, 7 (02) :3147-3150
[9]   Scanning electron microscope with polarization analysis: Micromagnetic structures in ultrathin films [J].
Oepen, HP ;
Steierl, G ;
Kirschner, J .
JOURNAL OF VACUUM SCIENCE & TECHNOLOGY B, 2002, 20 (06) :2535-2538
[10]   USING A MAGNETOMETER TO IMAGE A TWO-DIMENSIONAL CURRENT DISTRIBUTION [J].
ROTH, BJ ;
SEPULVEDA, NG ;
WIKSWO, JP .
JOURNAL OF APPLIED PHYSICS, 1989, 65 (01) :361-372