Utility of an automatic instrument for blood pressure measurement in children - The Bogalusa Heart Study

被引:28
作者
Wattigney, WA [1 ]
Webber, LS [1 ]
Lawrence, MD [1 ]
Berenson, GS [1 ]
机构
[1] TULANE UNIV, SCH PUBL HLTH & TROP MED, DEPT APPL HLTH SCI, NEW ORLEANS, LA 70112 USA
关键词
ausculatory blood pressure; Dinamap automatic blood pressure monitor; oscillometric blood pressure; automatic blood pressure monitoring;
D O I
10.1016/0895-7061(95)00334-7
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
The Dinamap 845XT automatic blood pressure monitor, Dinamap 8100 (an update model), and two mercury sphygmomanometers were compared in 417 school-aged children examined in the spring of 1992 as part of the Bogalusa Heart Study. This study was conducted in the nearby community of Franklinton, Louisiana, to verify data obtained as part of a cross-sectional survey (1987 to 1988) of school-aged children in Bogalusa. Systolic blood pressure levels were on the average 3 mm Hg higher on the Dinamap instruments than on the sphygmomanometers. Mean levels of diastolic blood pressure using either Dinamap instrument were slightly higher until eight years of age and then were considerably lower than mercury sphygmomanometer fourth phase readings. Diastolic blood pressure levels on the Dinamap 8100 were 4 mm Hg lower than on the Dinamap 845XT. Height was identified as the predominant predictor variable of differences in diastolic blood pressure between the mercury sphygmomanometer and either Dinamap instrument. A 10% random sample of children was reexamined each screening day in the cross-sectional survey to estimate measurement errors. The diastolic readings of the Dinamap 845XT had a lower intraclass correlation (0.68) compared to the mercury sphygmomanometers (0.83 fourth phase and 0.76 fifth phase). The Dinamap offers the ease of measuring systolic blood pressure although the diastolic blood pressure appears to be biased and especially low, particularly on the new 8100 model.
引用
收藏
页码:256 / 262
页数:7
相关论文
共 22 条
[1]  
[Anonymous], 1987, Pediatrics, V79, P1
[2]   THE IMPACT OF AUTOMATED BLOOD-PRESSURE DEVICES ON THE EFFICIENCY OF CLINICAL-TRIALS [J].
APPEL, LJ ;
MARWAHA, S ;
WHELTON, PK ;
PATEL, M .
CONTROLLED CLINICAL TRIALS, 1992, 13 (03) :240-247
[3]   COMPARISON OF 3 DEVICES FOR MEASURING BLOOD-PRESSURE [J].
BASSEIN, L ;
BORGHI, C ;
COSTA, FV ;
STROCCHI, E ;
MUSSI, A ;
AMBROSIONI, E .
STATISTICS IN MEDICINE, 1985, 4 (03) :361-368
[4]   COMPARISON OF 2 AUTOMATIC DEVICES AND THE STANDARD MERCURY SPHYGMOMANOMETER IN HYPERTENSIVE PATIENTS [J].
BASSEIN, L ;
BORGHI, C ;
COSTA, FV ;
STROCCHI, E ;
MUSSI, A ;
AMBROSIONI, E .
CLINICAL AND EXPERIMENTAL HYPERTENSION PART A-THEORY AND PRACTICE, 1985, 7 (2-3) :387-390
[5]  
BERENSON GS, 1980, CARDIOVASCULAR RISK, P72
[6]  
BRUCE NG, 1990, J HYPERTENS, V8, pS11
[7]   ANTHROPOMETRIC AND MATURATION MEASUREMENTS OF CHILDREN, AGES 5 TO 14 YEARS, IN A BIRACIAL COMMUNITY - BOGALUSA HEART STUDY [J].
FOSTER, TA ;
VOORS, AW ;
WEBBER, LS ;
FRERICHS, RR ;
BERENSON, GS .
AMERICAN JOURNAL OF CLINICAL NUTRITION, 1977, 30 (04) :582-591
[8]   COMPARISON OF AUTOMATIC OSCILLOMETRIC ARTERIAL-PRESSURE MEASUREMENT WITH CONVENTIONAL AUSCULTATORY MEASUREMENT IN THE LABOR WARD [J].
HASAN, MA ;
THOMAS, TA ;
PRYSROBERTS, C .
BRITISH JOURNAL OF ANAESTHESIA, 1993, 70 (02) :141-144
[9]   OBSERVER ERROR IN SYSTOLIC BLOOD-PRESSURE MEASUREMENT IN THE ELDERLY - A CASE FOR AUTOMATIC RECORDERS [J].
HLA, KM ;
VOKATY, KA ;
FEUSSNER, JR .
ARCHIVES OF INTERNAL MEDICINE, 1986, 146 (12) :2373-2376
[10]  
HUNTER SMD, 1980, PREV MED, V9, P701