Analysis of the accuracy of linear measurements obtained by cone beam computed tomography (CBCT-NewTom)

被引:352
作者
Lascala, C
Panella, J
Marques, MM
机构
[1] Univ Sao Paulo, Fac Odontol, Dept Estomatol, BR-05508900 Sao Paulo, Brazil
[2] Univ Sao Paulo, Sch Dent, Dept Radiol, Sao Paulo, Brazil
[3] Univ Sao Paulo, Sch Dent, Dept Operat Dent, Sao Paulo, Brazil
关键词
computed tomography; cone beam; x-ray computed; skull;
D O I
10.1259/dmfr/25500850
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Objectives: The conical beam computed tomography (CBCT) technique presents an innovation of tomographic imaging systems and subsequent volumetric image reconstruction for dentistry. When compared with other methods of tomographic imaging CBCT is characterized by rapid volumetric image acquisition from a single low radiation dose scan of the patient. The NewTom (NewTom 9000; Quantitative Radiology, Verona, Italy) is an example of such a CBCT machine dedicated to dental and maxillofacial imaging, particularly for surgical and/or prosthetics implant planning in the field of dentistry. The aim of this study was to evaluate the accuracy of the linear measurements obtained in CBCT images using a NewTom. Methods: Thirteen measurements were obtained in dry skulls (n = 8) between internal and external anatomical sites using a caliper. These were considered as real measurements. Then the dry skulls were submitted to CBCT imaging examinations. Radiographic distance measurements of the same dry skull anatomical sites were made using the NewTom QR-DVT 9000 software of the 2 mm-CTs axial section images and sagittal or coronal reconstructions. The data were compared by paired Student's t-test. Results: The results showed that the real measurements were always larger than those for the CBCT images, but these differences were only significant for measurements of the internal structures of the skull base. Conclusions: The conclusion of this study is that, although the CBCT image underestimates the real distances between skull sites, differences are only significant for the skull base and therefore it is reliable for linear evaluation measurements of other structures more closely associated with dentomaxillofacial imaging.
引用
收藏
页码:291 / 294
页数:4
相关论文
共 14 条
[1]  
BIANCHI S, 2003, DENTOMAXILLOFAC RAD, V30, pS28
[2]   Quantitative analysis of spiral computed tomography for craniofacial clinical applications [J].
Cavalcanti, MGP ;
Vannier, MW .
DENTOMAXILLOFACIAL RADIOLOGY, 1998, 27 (06) :344-350
[3]   CONE-BEAM CT FOR RADIOTHERAPY APPLICATIONS [J].
CHO, PS ;
JOHNSON, RH ;
GRIFFIN, TW .
PHYSICS IN MEDICINE AND BIOLOGY, 1995, 40 (11) :1863-1883
[4]  
CURRY TS, 1984, CHISTENSENS INTRO PH, P234
[5]   A comparison of a new limited cone beam computed tomography machine for dental use with a multidetector row helical CT machine [J].
Hashimoto, K ;
Arai, Y ;
Iwai, K ;
Araki, M ;
Kawashima, S ;
Terakado, M .
ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY AND ENDODONTOLOGY, 2003, 95 (03) :371-377
[6]   Quantitative analysis of three-dimensional rendered imaging of the human skull acquired from multi-detector row computed tomography [J].
Jung, H ;
Kim, HJ ;
Kim, DO ;
Hong, SI ;
Jeong, HK ;
Kim, KD ;
Kim, YO ;
Yoo, SK ;
Yoo, HS .
JOURNAL OF DIGITAL IMAGING, 2002, 15 (04) :232-239
[7]   Dosimetry of two extraoral direct digital imaging devices: NewTom cone beam CT and Orthophos Plus DS panoramic unit [J].
Ludlow, JB ;
Davies-Ludlow, LE ;
Brooks, SL .
DENTOMAXILLOFACIAL RADIOLOGY, 2003, 32 (04) :229-234
[8]   Radiation absorbed in maxillofacial imaging with a new dental computed tomography device [J].
Mah, JK ;
Danforth, RA ;
Bumann, A ;
Hatcher, D .
ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY AND ENDODONTICS, 2003, 96 (04) :508-513
[9]  
Misch C E, 1990, Int J Oral Implantol, V6, P23
[10]   A new volumetric CT machine for dental imaging based on the cone-beam technique: preliminary results [J].
Mozzo, P ;
Procacci, C ;
Tacconi, A ;
Martini, PT ;
Andreis, IAB .
EUROPEAN RADIOLOGY, 1998, 8 (09) :1558-1564