The importance of AIF ROI selection in DCE-MRI renography: Reproducibility and variability of renal perfusion and filtration

被引:44
作者
Cutajar, M. [1 ,2 ]
Mendichovszky, I. A. [3 ]
Tofts, P. S. [2 ,4 ]
Gordon, I. [1 ]
机构
[1] UCL Inst Child Hlth, Radiol & Phys Unit, London WC1N 1EH, England
[2] Univ Sussex, Brighton & Sussex Med Sch, Brighton BN1 9PX, E Sussex, England
[3] Univ Manchester, Manchester M13 9PT, Lancs, England
[4] UCL Inst Neurol, London WC1N 3BG, England
关键词
Renal dynamic contrast-enhanced magnetic resonance imaging (DCE-MRI); Aortic input function analysis; Reproducibility study; Gadolinium; Three-compartment tracer kinetics model; Perfusion; GFR; CONTRAST-ENHANCED MRI; PATLAK PLOT TECHNIQUE; GLOMERULAR-FILTRATION; COMPARTMENT MODEL; KIDNEY; DTPA; IMAGES;
D O I
10.1016/j.ejrad.2009.05.041
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
100231 [临床病理学]; 100902 [航空航天医学];
摘要
Purpose: The aim of this study was to investigate (a) the effect the choice of the region of interest (ROI) defining the aortic input function (AIF) has on the estimation of renal perfusion and filtration in dynamic contrast-enhanced (DCE) magnetic resonance imaging (MRI) renography, and (b) the reproducibility of these parameters. Using renal DCE-MRI and a three-compartment model analysis, this work evaluated the effect two different AIFs, derived from variable sized ROIs in the aorta, has on calculating DCE-MRI renal perfusion and filtration values in a group of healthy adult volunteers who underwent two consecutive renal DCE-MRI studies. Methods: Fifteen healthy volunteers underwent two DCE-MRI studies under similar physiological conditions. Oblique-coronal DCE-MRI data volumes were acquired on a 1.5 T Siemens Avanto scanner with a 3D-FLASH pulse-sequence (TE/TR = 0.53/1.63 ms, flip angle = 17 degrees, acquisition matrix = 128 x 104 voxels, strong fat saturation, PAT factor = 2 (GRAPPA) and 400 mm x 325 mm FOV). Each dynamic dataset consisted of 18 slices of 7.5 mm thickness (no gap) and an in-plane resolution of 3.1 mm x 3.1 mm, acquired every 2.5 s for not less than 5 minutes. During the MR scan a dose of 0.05 mmol (0.1 mL) kg(-1) body weight of dimeglumine gadopentetate (Magnevist) was injected intravenously (2 mL s(-1) injection rate), followed by a 15 mL saline flush at the same rate, using a MR-compatible automated injector (Spectris). Two AIFs were defined for each volunteer by drawing two ROIs in the aorta for each study. Renal perfusion and glomerular filtration rate (GFR) values were then calculated for each of the AIFs using a modified Tofts Renal Model (TRM). Both renal perfusion and GFR were expressed in mL min(-1) 100 mL(-1) of tissue. Results and conclusion: Inter-individual reproducibility tests for renal perfusion and glomerular filtration rate showed that the size of AIF ROIs significantly affects calculated values of perfusion and GFR (p-values <0.02). No significant differences were observed when comparing perfusion and GFR values in the same volunteer between scans performed on different days (p-values >0.22). From our study we conclude that while DCE-MRI derived indices of renal function are reproducible in the same individual when imaged on different days, the size of the aortic ROI and hence the AIF has a significant influence on calculated renal perfusion and GFR values. Currently there is no accepted standard for drawing the aortic ROI and no standardized approach for the AIF definition in renal DCE-MRI studies. (C) 2009 Elsevier Ireland Ltd. All rights reserved.
引用
收藏
页码:E155 / E161
页数:7
相关论文
共 22 条
[1]
Glomerular filtration rate: Assessment with dynamic contrast-enhanced MRI and a cortical-compartment model in the rabbit kidney [J].
Annet, L ;
Hermoye, L ;
Peeters, F ;
Jamar, F ;
Dehoux, JP ;
Van Beers, BE .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2004, 20 (05) :843-849
[2]
Reproducibilities and responses to food intake of GFR measured with chromium-51-EDTA and iohexol simultaneously and independently in normal subjects [J].
Bird, Nicholas J. ;
Peters, Christina ;
Michell, A. Robert ;
Peters, A. Michael .
NEPHROLOGY DIALYSIS TRANSPLANTATION, 2008, 23 (06) :1902-1909
[3]
Measurement of single kidney function using dynamic contrast-enhanced MRI: Comparison of two models in human subjects [J].
Buckley, David L. ;
Shurrab, Ala'a E. ;
Cheung, Ching M. ;
Jones, Andrew P. ;
Mamtora, Hari ;
Kalra, Philip A. .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2006, 24 (05) :1117-1123
[4]
Improvement of MRI-functional measurement with automatic movement correction in native and transplanted kidneys [J].
de Senneville, Baudouin Denis ;
Mendichovszky, Iosif Alexandru ;
Roujol, Sebastien ;
Gordon, Isky ;
Moonen, Chrit ;
Grenier, Nicolas .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2008, 28 (04) :970-978
[5]
Measurement of glomerular filtration rate with magnetic resonance imaging: Principles, limitations, and expectations [J].
Grenier, Nicolas ;
Mendichovszky, Iosif ;
De Senneville, Baudouin Denis ;
Rouiol, Sbastien ;
Desbarats, Pascal ;
Pedersen, Michael ;
Wells, Kevin ;
Frokiaer, Jorgen ;
Gordon, Isky .
SEMINARS IN NUCLEAR MEDICINE, 2008, 38 (01) :47-55
[6]
Glomerular filtration rate measured using the Patlak plot technique and contrast-enhanced dynamic MRI with different amounts of gadolinium-DTPA [J].
Hackstein, N ;
Kooijman, H ;
Tomaselli, S ;
Rau, WS .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2005, 22 (03) :406-414
[7]
Measurement of single-kidney glomerular filtration rate using a contrast-enhanced dynamic gradient-echo sequence and the Rutland-Patlak plot technique [J].
Hackstein, N ;
Heckrodt, J ;
Rau, WS .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2003, 18 (06) :714-725
[8]
Renal function measurements from MR renography and a simplified multicompartmental model [J].
Lee, Vivian S. ;
Rusinek, Henry ;
Bokacheva, Louisa ;
Huang, Ambrose J. ;
Oesingmann, Niels ;
Chen, Qun ;
Kaur, Manmeen ;
Prince, Keyma ;
Song, Ting ;
Kramer, Elissa L. ;
Leonard, Edward F. .
AMERICAN JOURNAL OF PHYSIOLOGY-RENAL PHYSIOLOGY, 2007, 292 (05) :F1548-F1559
[9]
Reproducibility of the aortic input function (AIF) derived from dynamic contrast-enhanced magnetic resonance imaging (DCE-MRI) of the kidneys in a volunteer study [J].
Mendichovszky, I. A. ;
Cutajar, M. ;
Gordon, I. .
EUROPEAN JOURNAL OF RADIOLOGY, 2009, 71 (03) :576-581
[10]
How accurate is dynamic contrast-enhanced MRI in the assessment of renal glomerular filtration rate?: A critical appraisal [J].
Mendichovszky, Iosif ;
Pederson, Michael ;
Frokiaer, Jorgen ;
Dissing, Thomas ;
Grenier, Nicolas ;
Anderson, Peter ;
McHugh, Kieran ;
Yang, Qing ;
Gordon, Isky .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2008, 27 (04) :925-931