Microfluidic library screening for mapping antibody epitopes

被引:47
作者
Bessette, Paul H.
Hu, Xiaoyuan
Soh, Hyongsok T. [1 ]
Daugherty, Patrick S.
机构
[1] Univ Calif Santa Barbara, Dept Chem Engn, Santa Barbara, CA 93106 USA
[2] Univ Calif Santa Barbara, Dept Mech Engn, Santa Barbara, CA 93106 USA
[3] Univ Calif Santa Barbara, Inst Collaborat Biotechnol, Santa Barbara, CA 93106 USA
关键词
D O I
10.1021/ac0616916
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The capability to screen molecular libraries using disposable microfluidic devices provides the potential to simplify and automate reagent generation and to develop integrated bioanalytical systems for clinical diagnostics. Here, antibody epitopes were mapped using a disposable microfluidic device to screen a combinatorial peptide library composed of 5 x 10(8) members displayed on bacterial cells. On-chip library screening was achieved in a two-stage, continuous-flow microfluidic sorter that separates antibody-binding target cells captured on microspheres through dielectrophoretic funneling. The antibody fingerprints identified were comparable to those obtained using state-of-the-art commercial cell sorting instrumentation.
引用
收藏
页码:2174 / 2178
页数:5
相关论文
共 20 条
[1]   BioMEMS: state-of-the-art in detection, opportunities and prospects [J].
Bashir, R .
ADVANCED DRUG DELIVERY REVIEWS, 2004, 56 (11) :1565-1586
[2]   Rapid isolation of high-affinity protein binding peptides using bacterial display [J].
Bessette, PH ;
Rice, JJ ;
Daugherty, PS .
PROTEIN ENGINEERING DESIGN & SELECTION, 2004, 17 (10) :731-739
[3]   Flow cytometric screening of cDNA expression libraries for fluorescent proteins [J].
Bessette, PH ;
Daugherty, PS .
BIOTECHNOLOGY PROGRESS, 2004, 20 (03) :963-967
[4]  
BRIZZARD BL, 1994, BIOTECHNIQUES, V16, P730
[5]   Epitope mapping and affinity purification of monospecific antibodies by Escherichia coli cell surface display of gene-derived random peptide libraries [J].
Christmann, A ;
Wentzel, A ;
Meyer, C ;
Meyers, G ;
Kolmar, H .
JOURNAL OF IMMUNOLOGICAL METHODS, 2001, 257 (1-2) :163-173
[6]   Go with the microflow [J].
Clayton, J .
NATURE METHODS, 2005, 2 (08) :621-627
[7]   Cells on chips [J].
El-Ali, Jamil ;
Sorger, Peter K. ;
Jensen, Klavs F. .
NATURE, 2006, 442 (7101) :403-411
[8]   Marker-specific sorting of rare cells using dielectrophoresis [J].
Hu, XY ;
Bessette, PH ;
Qian, JR ;
Meinhart, CD ;
Daugherty, PS ;
Soh, HT .
PROCEEDINGS OF THE NATIONAL ACADEMY OF SCIENCES OF THE UNITED STATES OF AMERICA, 2005, 102 (44) :15757-15761
[9]   Continuous dielectrophoretic size-based particle sorting [J].
Kralj, Jason G. ;
Lis, Michael T. W. ;
Schmidt, Martin A. ;
Jensen, Klavs F. .
ANALYTICAL CHEMISTRY, 2006, 78 (14) :5019-5025
[10]   Fully integrated PCR-capillary electrophoresis microsystem for DNA analysis [J].
Lagally, ET ;
Emrich, CA ;
Mathies, RA .
LAB ON A CHIP, 2001, 1 (02) :102-107