Pitfalls and challenges when assessing the depth of hypnosis during general anaesthesia by clinical signs and electronic indices

被引:44
作者
Jensen, EW
Litvan, H
Struys, M
Vazquez, PM
机构
[1] Hosp Santa Creu & Sant Pau, Dept Cardiac Anaesthesia, Barcelona, Spain
[2] Danmeter AS, Res Sect, Odense, Denmark
[3] Hosp Santa Creu & Sant Pau, Postoperat Intens Care Unit, Barcelona, Spain
[4] Ghent Univ Hosp, Dept Anaesthesia, B-9000 Ghent, Belgium
关键词
auditory evoked potentials; Bispectral index; depth of hypnosis; general anaesthesia; monitoring;
D O I
10.1111/j.1399-6576.2004.00521.x
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
The objective of this article was to review the present methods used for validating the depth of hypnosis. We introduce three concepts, the real depth of hypnosis (DHreal), the observed depth of hypnosis (DHobs), and the electronic indices of depth of hypnosis (DHel-ind). The DHreal is the real state of hypnosis that the patient has in a given moment during the general anaesthesia. The DHobs is the subjective assessment of the anaesthesiologist based on clinical signs. The DHel-ind is any estimation of the depth of hypnosis given by an electronic device. The three entities DHreal, DHobs and DHel-ind should in the ideal situation be identical. However, this is rarely the case. The correlation between the DHobs and the DHel-ind can be affected by a number of factors such as the stimuli used for the assessment of the level of consciousness or the administration of analgesic agents or neuro muscular blocking agents. Opioids, for example, can block the response to tactile and noxious stimuli, and even the response to verbal command could vanish, hence deeming the patient in a lower depth of hypnosis than the real patient state. The DHel-ind can be disturbed by the presence of facial muscular activity. In conclusion, although several monitors and clinical scoring scales are available to assess the depth of hypnosis during general anaesthesia, care should be taken when interpreting their results.
引用
收藏
页码:1260 / 1267
页数:8
相关论文
共 38 条
[1]   CORNEAL AND BLINK REFLEXES IN PARKINSONS-DISEASE WITH ON-OFF FLUCTUATIONS [J].
AGOSTINO, R ;
BERARDELLI, A ;
CRUCCU, G ;
STOCCHI, F ;
MANFREDI, M .
MOVEMENT DISORDERS, 1987, 2 (04) :227-235
[2]   Effects on the bispectral index during medium-high dose fentanyl induction with or without propofol supplement [J].
Barr, G ;
Anderson, RE ;
Öwall, A ;
Jakobsson, JG .
ACTA ANAESTHESIOLOGICA SCANDINAVICA, 2000, 44 (07) :807-811
[3]   Nitrous oxide does not alter bispectral index:: study with nitrous oxide as sole agent and as an adjunct to i.v. anaesthesia [J].
Barr, G ;
Jakobsson, JG ;
Öwall, A ;
Anderson, RE .
BRITISH JOURNAL OF ANAESTHESIA, 1999, 82 (06) :827-830
[4]   The effects of nitrous oxide on the auditory evoked potential index during sevoflurane anaesthesia [J].
Barr, G ;
Anderson, R ;
Jakobsson, J .
ANAESTHESIA, 2002, 57 (08) :736-739
[5]   Correlation of approximate entropy, bispectral index, and spectral edge frequency 95 (SEF95) with clinical signs of "anesthetic depth" during coadministration of propofol and remifentanil [J].
Bruhn, J ;
Bouillon, TW ;
Radulescu, L ;
Hoeft, A ;
Bertaccini, E ;
Shafer, SL .
ANESTHESIOLOGY, 2003, 98 (03) :621-627
[6]  
CHERNIK DA, 1990, J CLIN PSYCHOPHARM, V10, P244
[7]  
DECASTRO J, 1976, ANN ANESTHESIOL FR, V17, P1071
[8]   Patient state index - Titration of delivery and recovery from propofol, alfentanil, and nitrous oxide anesthesia [J].
Drover, DR ;
Lemmens, HJ ;
Pierce, ET ;
Plourde, G ;
Loyd, G ;
Ornstein, E ;
Prichep, LS ;
Chabot, RJ ;
Gugino, L .
ANESTHESIOLOGY, 2002, 97 (01) :82-89
[9]   A neurophysiological approach to brainstem reflexes. Blink reflex [J].
Esteban, A .
NEUROPHYSIOLOGIE CLINIQUE-CLINICAL NEUROPHYSIOLOGY, 1999, 29 (01) :7-38
[10]   Unintentional intraoperative awareness during sufentanil anaesthesia for cardiac surgery [J].
Gilron, I ;
Solomon, P ;
Plourde, G .
CANADIAN JOURNAL OF ANAESTHESIA-JOURNAL CANADIEN D ANESTHESIE, 1996, 43 (03) :295-298