Direct characterization of intact polypeptides by matrix-assisted laser desorption electrospray ionization quadrupole Fourier transform ion cyclotron resonance mass spectrometry

被引:46
作者
Sampson, Jason S. [1 ]
Hawkridge, Adam M. [1 ]
Muddiman, David C. [1 ]
机构
[1] N Carolina State Univ, Dept Chem, WM Keck FT ICR Mass Spectrometry Lab, Raleigh, NC 27695 USA
关键词
D O I
10.1002/rcm.2947
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
We report the characterization of a recently introduced hybrid ionization source, matrix-assisted laser desorption electrospray ionization (MALDESI), coupled to a quadrupole Fourier transform ion cyclotron resonance mass spectrometry (QFT-ICR-MS) system. We first demonstrate the ability of MALDESI-QFT-ICR MS to directly analyze and provide high mass measurement accuracy (similar to 1 part-per-million) of a polypeptide using internal calibration. Second, we show the potential of MALDESI-QFT-ICR MS for the top-down characterization of multiply charged polypeptide cations. Finally, we demonstrate sub-femtomole detection limits in MALDESI-QFT-ICR MS using a combination of naturally occurring peptides and their respective stable isotope labeled forms. The results presented herein demonstrate the feasibility of several potential applications for MALDESI-QFT-CR MS for the direct analysis of intact biological molecules. Copyright (c) 2007 John Wiley & Sons, Ltd.
引用
收藏
页码:1150 / 1154
页数:5
相关论文
共 28 条
[1]   Versatile new ion source for the analysis of materials in open air under ambient conditions [J].
Cody, RB ;
Laramée, JA ;
Durst, HD .
ANALYTICAL CHEMISTRY, 2005, 77 (08) :2297-2302
[2]   Atmospheric pressure laser desorption/chemical ionization mass spectrometry: a new ionization method based on existing themes [J].
Coon, JJ ;
McHale, KJ ;
Harrison, WW .
RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2002, 16 (07) :681-685
[3]   ELECTROSPRAY IONIZATION FOR MASS-SPECTROMETRY OF LARGE BIOMOLECULES [J].
FENN, JB ;
MANN, M ;
MENG, CK ;
WONG, SF ;
WHITEHOUSE, CM .
SCIENCE, 1989, 246 (4926) :64-71
[4]   Accessible proteomics space and its implications for peak capacity for zero-, one- and two-dimensional separations coupled with FT-ICR and TOF mass spectrometry [J].
Frahm, JL ;
Howard, BE ;
Heber, S ;
Muddiman, DC .
JOURNAL OF MASS SPECTROMETRY, 2006, 41 (03) :281-288
[5]   A dual electrospray ionization source combined with hexapole accumulation to achieve high mass accuracy of biopolymers in fourier transform ion cyclotron resonance mass spectrometry [J].
Hannis, JC ;
Muddiman, DC .
JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 2000, 11 (10) :876-883
[6]   Analytical performance of a venturi device integrated into an electrospray ionization Fourier transform ion cyclotron resonance mass spectrometer for analysis of nucleic acids [J].
Hawkridge, AM ;
Zhou, L ;
Lee, ML ;
Muddiman, DC .
ANALYTICAL CHEMISTRY, 2004, 76 (14) :4118-4122
[7]   FOURIER-TRANSFORM MASS-SPECTROMETRY OF LARGE MOLECULES BY ELECTROSPRAY IONIZATION [J].
HENRY, KD ;
WILLIAMS, ER ;
WANG, BH ;
MCLAFFERTY, FW ;
SHABANOWITZ, J ;
HUNT, DF .
PROCEEDINGS OF THE NATIONAL ACADEMY OF SCIENCES OF THE UNITED STATES OF AMERICA, 1989, 86 (23) :9075-9078
[8]   Direct protein detection from biological media through electrospray-assisted laser desorption ionization/mass spectrometry [J].
Huang, MZ ;
Hsu, HJ ;
Lee, LY ;
Jeng, JY ;
Shiea, JT .
JOURNAL OF PROTEOME RESEARCH, 2006, 5 (05) :1107-1116
[9]   LASER DESORPTION IONIZATION OF PROTEINS WITH MOLECULAR MASSES EXCEEDING 10000 DALTONS [J].
KARAS, M ;
HILLENKAMP, F .
ANALYTICAL CHEMISTRY, 1988, 60 (20) :2299-2301
[10]   Top-down proteomics [J].
Kelleher, NL .
ANALYTICAL CHEMISTRY, 2004, 76 (11) :196A-203A