Fabrication of oxygen electrode arrays and their incorporation into sensors for measuring biochemical oxygen demand

被引:55
作者
Yang, Z
Sasaki, S
Karube, I
Suzuki, H
机构
[1] Univ Tokyo, Adv Sci & Technol Res Ctr, Meguro Ku, Tokyo 153, Japan
[2] Univ Tsukuba, Inst Mat Sci, Tsukuba, Ibaraki 305, Japan
关键词
oxygen electrode arrays; sensors; biochemical oxygen demand;
D O I
10.1016/S0003-2670(97)00560-6
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Miniature Clark-type oxygen electrode arrays were fabricated using thin film technology. Each oxygen electrode comprised a silver cathode and a Ag/AgCl anode. The use of a dry electrolyte containing polyvinylpyrrolidone (PVP) permitted the use of solid state fabrication processes. The electrolyte was left dry until it was used, when the water essential for oxygen reduction was introduced through a gas-permeable membrane and absorbed by the PVP. The miniature oxygen electrode arrays showed good characteristics for monitoring dissolved oxygen concentrations and could be mass produced with assured quality, The yeast Trichosporon cutaneum was immobilized onto the cathode of an oxygen electrode using a photo-crosslinkable resin and the response of the sensor was the difference between the output of a yeast-immobilized electrode and that of a bare oxygen electrode. Measurements using the dual-type sensor for biochemical oxygen demand (BOD) were markedly simplified as air-saturated samples were not required, so air pumps and stirrers can be omitted from the sensing system. The BOD sensor was used for the analysis of real samples and the results agreed well with those determined using the conventional 5-day BOD determination method. This is a significant step toward the development of a portable BOD measuring system incorporating disposable electrodes and containing no mechanical parts. (C) 1997 Elsevier Science B.V.
引用
收藏
页码:41 / 49
页数:9
相关论文
共 25 条
[11]   2-DIMENSIONAL IMAGING OF O-2, H2O2, AND GLUCOSE DISTRIBUTIONS BY AN ARRAY OF 400 INDIVIDUALLY ADDRESSABLE MICROELECTRODES [J].
MEYER, H ;
DREWER, H ;
GRUNDIG, B ;
CAMMANN, K ;
KAKEROW, R ;
MANOLI, Y ;
MOKWA, W ;
ROSPERT, M .
ANALYTICAL CHEMISTRY, 1995, 67 (07) :1164-1170
[12]   LONG-LIFE PLANAR OXYGEN SENSOR [J].
MIYAHARA, Y ;
TSUKADA, K ;
WATANABE, Y .
SENSORS AND ACTUATORS B-CHEMICAL, 1994, 20 (2-3) :89-94
[13]   INTEGRATION OF ENZYME-IMMOBILIZED COLUMN WITH ELECTROCHEMICAL FLOW CELL USING MICROMACHINING TECHNIQUES FOR A GLUCOSE DETECTION SYSTEM [J].
MURAKAMI, Y ;
TAKEUCHI, T ;
YOKOYAMA, K ;
TAMIYA, E ;
KARUBE, I ;
SUDA, M .
ANALYTICAL CHEMISTRY, 1993, 65 (20) :2731-2735
[14]   BIOREACTORS AND BIOMEMBRANES FOR BIOCHEMICAL OXYGEN-DEMAND ESTIMATION [J].
PRAET, E ;
REUTER, V ;
GAILLARD, T ;
VASEL, JL .
TRAC-TRENDS IN ANALYTICAL CHEMISTRY, 1995, 14 (07) :371-378
[15]   DESIGN AND VALIDATION OF THE TRANSPARENT OXYGEN SENSOR ARRAY [J].
SARGENT, BJ ;
GOUGH, DA .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1991, 38 (05) :476-482
[16]   DISPOSABLE OXYGEN ELECTRODES FABRICATED BY SEMICONDUCTOR TECHNIQUES AND THEIR APPLICATION TO BIOSENSORS [J].
SUZUKI, H ;
KOJIMA, N ;
SUGAMA, A ;
TAKEI, F ;
IKEGAMI, K .
SENSORS AND ACTUATORS B-CHEMICAL, 1990, 1 (1-6) :528-532
[17]   A MINIATURE CLARK-TYPE OXYGEN-ELECTRODE USING A POLYELECTROLYTE AND ITS APPLICATION AS A GLUCOSE SENSOR [J].
SUZUKI, H ;
SUGAMA, A ;
KOJIMA, N ;
TAKEI, F ;
IKEGAMI, K .
BIOSENSORS & BIOELECTRONICS, 1991, 6 (05) :395-400
[18]   MICROMACHINED CLARK OXYGEN-ELECTRODE [J].
SUZUKI, H ;
SUGAMA, A ;
KOJIMA, N .
SENSORS AND ACTUATORS B-CHEMICAL, 1993, 10 (02) :91-98
[19]   FABRICATION OF AN OXYGEN-ELECTRODE USING SEMICONDUCTOR TECHNOLOGY [J].
SUZUKI, H ;
TAMIYA, E ;
KARUBE, I .
ANALYTICAL CHEMISTRY, 1988, 60 (10) :1078-1080
[20]   DEVELOPMENT OF A MINIATURE CLARK-TYPE OXYGEN-ELECTRODE USING SEMICONDUCTOR TECHNIQUES AND ITS IMPROVEMENT FOR PRACTICAL APPLICATIONS [J].
SUZUKI, H ;
KOJIMA, N ;
SUGAMA, A ;
TAKEI, F .
SENSORS AND ACTUATORS B-CHEMICAL, 1990, 2 (03) :185-191