Evaluation of a new body composition phantom for quality control and cross-calibration of DXA devices

被引:33
作者
Diessel, E
Fuerst, T
Njeh, CF
Tylavsky, F
Cauley, J
Dockrell, M
Genant, HK
机构
[1] Univ Calif San Francisco, Osteoporosis & Arthrit Res Grp, San Francisco, CA 94143 USA
[2] Synarc Inc, San Francisco, CA 94105 USA
[3] Univ Tennessee, Memphis, TN 38103 USA
[4] Univ Pittsburgh, Pittsburgh, PA 15238 USA
关键词
dual-energy X-ray absorptiometry; in vivo; percentage body fat; precision;
D O I
10.1152/jappl.2000.89.2.599
中图分类号
Q4 [生理学];
学科分类号
071003 ;
摘要
This study evaluated a new body composition phantom and its use for quality control and cross-calibration of dual-energy X-ray absorptiometry (DXA) instruments for measurements of body composition. We imaged the variable composition phantom (Lunar, Madison, WI) on eight different DXA devices. Deviations of up to 7% fat were observed when we compared the percent fat values measured by the different devices with the nominal values provided by the manufacturer. Absolute precision error of percent fat measurements for the phantom ranged from 0.6 to 0.8%. The phantom's percent fat values were also compared with whole body composition measurements from 130 female and male volunteers. The phantom detected differences in percent fat values that were similar to those found by comparing in vivo measurements with values from different DXA scanner models from the same manufacturer. When comparing different models of scanners from different manufacturers, such as the Hologic QDR-4500 and the Lunar DPX-IQ, the phantom showed a different relationship than was seen for patients. Therefore, corrections or comparisons based on the phantom data alone would be incorrect. In conclusion, the Lunar variable composition phantom is capable of accurately measuring the fat calibration of DXA devices and may be suitable for cross-sectional cross-calibration between scanners from the same manufacturer; however, for comparison of DXA scanners from different manufacturers, in vivo cross-calibration is still the only accurate method. The phantom may be used in longitudinal quality control to verify an instrument's temporal stability.
引用
收藏
页码:599 / 605
页数:7
相关论文
共 29 条
[1]   CROSS CALIBRATION OF QDR-2000 AND QDR-1000 DUAL-ENERGY X-RAY DENSITOMETERS FOR BONE-MINERAL AND SOFT-TISSUE MEASUREMENTS [J].
ABRAHAMSEN, B ;
GRAM, J ;
HANSEN, TB ;
BECKNIELSEN, H .
BONE, 1995, 16 (03) :385-390
[2]   Replacing DXA scanners: Cross-calibration with phantoms may be misleading [J].
Blake, GM .
CALCIFIED TISSUE INTERNATIONAL, 1996, 59 (01) :1-5
[3]   Bone mineral and body composition measurements: Cross-calibration of pencil-beam and fan-beam dual-energy X-ray absorptiometers [J].
Ellis, KJ ;
Shypailo, RJ .
JOURNAL OF BONE AND MINERAL RESEARCH, 1998, 13 (10) :1613-1618
[4]   QUALITY-CONTROL OF DXA INSTRUMENTS IN MULTICENTER TRIALS [J].
FAULKNER, KG ;
MCCLUNG, MR .
OSTEOPOROSIS INTERNATIONAL, 1995, 5 (04) :218-227
[5]   ASSESSMENT OF THE COMPOSITION OF MAJOR BODY REGIONS BY DUAL-ENERGY X-RAY ABSORPTIOMETRY (DEXA), WITH SPECIAL REFERENCE TO LIMB MUSCLE MASS [J].
FULLER, NJ ;
LASKEY, MA ;
ELIA, M .
CLINICAL PHYSIOLOGY, 1992, 12 (03) :253-266
[6]   DXA longitudinal quality control: A comparison of inbuilt quality assurance, visual inspection, multi-rule Shewhart charts and Cusum analysis [J].
Garland, SW ;
Lees, B ;
Stevenson, JC .
OSTEOPOROSIS INTERNATIONAL, 1997, 7 (03) :231-237
[7]  
GENANT HK, 1994, J BONE MINER RES, V9, P1503
[8]   QUALITY ASSURANCE FOR BONE DENSITOMETRY RESEARCH STUDIES - CONCEPT AND IMPACT [J].
GLUER, CC ;
FAULKNER, KG ;
ESTILO, MJ ;
ENGELKE, K ;
ROSIN, J ;
GENANT, HK .
OSTEOPOROSIS INTERNATIONAL, 1993, 3 (05) :227-235
[9]   ACCURATE ASSESSMENT OF PRECISION ERRORS - HOW TO MEASURE THE REPRODUCIBILITY OF BONE DENSITOMETRY TECHNIQUES [J].
GLUER, CC ;
BLAKE, G ;
LU, Y ;
BLUNT, BA ;
JERGAS, M ;
GENANT, HK .
OSTEOPOROSIS INTERNATIONAL, 1995, 5 (04) :262-270
[10]   EVALUATION OF A NEW SET OF CALIBRATION STANDARDS FOR THE MEASUREMENT OF FAT-CONTENT VIA DPA AND DXA [J].
GOODSITT, MM .
MEDICAL PHYSICS, 1992, 19 (01) :35-44