Sixteen-detector row CT of abdomen and pelvis: Study for optimization of Z-axis modulation technique performed in 153 patients

被引:124
作者
Kalra, MK
Maher, MM
Kamath, RS
Horiuchi, T
Toth, TL
Halpern, EF
Saini, S
机构
[1] Massachusetts Gen Hosp, Dept Radiol, Div Abdominal Imaging & Intervent, Boston, MA 02114 USA
[2] Harvard Univ, Sch Med, Boston, MA 02114 USA
[3] GE Med Syst, Waukesha, WI USA
[4] GE Yokogawa Med Syst, Tokyo, Japan
关键词
abdomen; CT; computed tomography (CT); image quality computed tomography (CT); radiation exposure pelvis;
D O I
10.1148/radiol.2331031505
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PURPOSE: To retrospectively determine the optimal noise indexes required to obtain diagnostically acceptable computed tomographic (CT) images of the abdomen and pelvis with z-axis modulation. MATERIALS AND METHODS: Ninety-five patients underwent 16-section multi-detector row CT of the abdomen and pelvis with z-axis modulation at noise indexes of 120.5, 11.0, and 12.0 HU with 10-380 mA. Subsequently, 58 patients were scanned at noise indexes of 12.5 and 15.0 HU with 75-380 mA. The weights of all subjects were recorded, and transverse and anteroposterior diameters were measured. The CTimages were evaluated for abnormalities and graded for image quality in terms of noise and diagnostic acceptability by using a five-point scale. Objective noise in the liver parenchyma was measured, and the tube current was recorded at each section in all 153 patients. Statistical analyses were performend to determine the appropriate noise index and to assess the effect of the patient weight and abdominal diameters on image noise and diagnostic acceptability at different noise indexes. Tube current-time products (in milliampere seconds) at various noise indexes were compared with those at CT previously performed without z-axis modulation. RESULTS: No significant difference insubjective image noise or diagnostic acceptaptability was found at noise indexes of 10.5-15.0 HU (p = .14), and objective noise was significantly inferior only at a noise index of 15.0 HU (p = .009). Compared with CT scanning at a 10.5-HU noise index, CT scanning at 12.5- and 15-HU noise indexes yielded, respectively, 10.0% and 41.3% reductions in radiation exposure. Patient weight and abdominal diameters affected subjective image quality. CONCLUSION: Use of a 15.0-HU noise index at 75-380 mA results in acceptable subjective image noise and diagnostic acceptability but acceptability but significantly greater objective image noise at routine abdominal-pelvic CT. For greater image quality demands, a noise index of 12.5 HU results in acceptable image quality demands, a noise index of 12.5 HU results in acceptable image quality and a 19.6% reduction in radiation exposure.
引用
收藏
页码:241 / 249
页数:9
相关论文
共 14 条
  • [1] Multi-slice CT-colonography in low-dose technique -: Preliminary results
    Cohnen, M
    Vogt, C
    Aurich, V
    Beck, A
    Häussinger, D
    Mödder, U
    [J]. ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2002, 174 (07): : 835 - 838
  • [2] The Smart-Scan-Technique in spiral computed tomography. A new method for dose reduction.
    Giacomuzzi, SM
    Erckert, B
    Schopf, T
    Freund, MC
    Springer, P
    Dessl, A
    Jaschke, W
    [J]. ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 1996, 165 (01): : 10 - 16
  • [3] Dose reduction in CT examination of children by an attenuation-based on-line modulation of tube current (CARE Dose)
    Greess, H
    Nömayr, A
    Wolf, H
    Baum, U
    Lell, M
    Böwing, B
    Kalender, W
    Bautz, WA
    [J]. EUROPEAN RADIOLOGY, 2002, 12 (06) : 1571 - 1576
  • [4] Dose reduction in computed tomography by attenuation-based on-line modulation of tube current: evaluation of six anatomical regions
    Greess, H
    Wolf, H
    Baum, U
    Lell, M
    Pirkl, M
    Kalender, W
    Bautz, WA
    [J]. EUROPEAN RADIOLOGY, 2000, 10 (02) : 391 - 394
  • [5] Greess H, 1999, ROFO-FORTSCHR RONTG, V170, P246
  • [6] Generalized multi-dimensional adaptive filtering for conventional and spiral single-slice, multi-slice, and cone-beam CT
    Kachelriess, M
    Watzke, O
    Kalender, WA
    [J]. MEDICAL PHYSICS, 2001, 28 (04) : 475 - 490
  • [7] Dose reduction in CT by on-line tube current control: principles and validation on phantoms and cadavers
    Kalender, WA
    Wolf, H
    Suess, C
    Gies, M
    Greess, H
    Bautz, WA
    [J]. EUROPEAN RADIOLOGY, 1999, 9 (02) : 323 - 328
  • [8] Strategies for CT radiation dose optimization
    Kalra, MK
    Maher, MM
    Toth, TL
    Hamberg, LM
    Blake, MA
    Shephard, JA
    Saini, S
    [J]. RADIOLOGY, 2004, 230 (03) : 619 - 628
  • [9] Can noise reduction filters improve low-radiation-dose chest CT images? Pilot study
    Kalra, MK
    Wittram, C
    Maher, MM
    Sharma, A
    Avinash, GB
    Karau, K
    Toth, TL
    Halpern, E
    Saini, S
    Shepard, JA
    [J]. RADIOLOGY, 2003, 228 (01) : 257 - 264
  • [10] Low-dose CT of the abdomen: Evaluation of image improvement with use of noise reduction filters - Pilot study
    Kalra, MK
    Maher, MM
    Sahani, DV
    Blake, MA
    Hahn, PF
    Avinash, GB
    Toth, TL
    Halpern, E
    Saini, S
    [J]. RADIOLOGY, 2003, 228 (01) : 251 - 256