Influence of calcium on glucose biosensor response and on hydrogen peroxide detection

被引:11
作者
Labat-Allietta, N [1 ]
Thevenot, DR [1 ]
机构
[1] Univ Paris 12, Fac Sci & Technol, LABAM, F-94010 Creteil, France
关键词
glucose sensor; calcium; hydrogen peroxide; Koutecky-Levich plots; platinum;
D O I
10.1016/S0956-5663(97)00084-5
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Of small species capable of reaching a platinum working electrode from biological samples, calcium cations have been found to inhibit significantly glucose biosensor responses. The sensitivities to glucose of sensors immersed in carbonate buffer saline solutions decreased when 0.5 mM calcium chloride was added. The degree of inhibition was proportional to the glucose response in the absence of calcium (0-17% of the normalized current). Likewise, sensor sensitivities to hydrogen peroxide decreased, in the 5-90% range, in the presence of 0.5 mM calcium. Bare Pt-Ir wires show a reversible inhibition of hydrogen peroxide sensitivity. This reversible inhibition is directly related to the decrease of hydrogen peroxide oxidation rate at the platinum anode: this has been evidenced, using rotating disk electrodes, by plotting Koutecky-Levich plots. Such inhibition has been found both for free and chelated calcium cations at levels below 1 mM. Several hypotheses for possible reactions between platinum, hydrogen peroxide and calcium are discussed. (C) 1997 Elsevier Science Limited. All rights reserved.
引用
收藏
页码:19 / 29
页数:11
相关论文
共 25 条
[1]  
[Anonymous], 1969, REACTIONS CHIMIQUES
[2]  
BARD AJ, 1980, ELECTROCHEMICAL METH
[3]   DESIGN AND INVITRO STUDIES OF A NEEDLE-TYPE GLUCOSE SENSOR FOR SUBCUTANEOUS MONITORING [J].
BINDRA, DS ;
ZHANG, YN ;
WILSON, GS ;
STERNBERG, R ;
THEVENOT, DR ;
MOATTI, D ;
REACH, G .
ANALYTICAL CHEMISTRY, 1991, 63 (17) :1692-1696
[4]   STIMULATION OF HEXOSE MONOPHOSPHATE SHUNT IN HUMAN NEUTROPHILS BY ASCORBIC-ACID - MECHANISM OF ACTION [J].
DECHATEL.LR ;
MCCALL, CE ;
COOPER, MR .
ANTIMICROBIAL AGENTS AND CHEMOTHERAPY, 1972, 1 (01) :12-&
[5]   ASSESSMENT OF SUBCUTANEOUS GLUCOSE-CONCENTRATION - VALIDATION OF THE WICK TECHNIQUE AS A REFERENCE FOR IMPLANTED ELECTROCHEMICAL SENSORS IN NORMAL AND DIABETIC DOGS [J].
FISCHER, U ;
ERTLE, R ;
ABEL, P ;
REBRIN, K ;
BRUNSTEIN, E ;
VONDORSCHE, HH ;
FREYSE, EJ .
DIABETOLOGIA, 1987, 30 (12) :940-945
[6]   Oxidation and reduction of vitamin C [J].
Hand, DB ;
Greisen, EC .
JOURNAL OF THE AMERICAN CHEMICAL SOCIETY, 1942, 64 :358-361
[7]  
HANNING I, 1986, ANAL LETT, V19, P461
[8]  
KERNER W, 1993, BIOM CONC ACT CHEM S
[9]  
Kotrly S., 1985, HDB CHEM EQUILIBRIA
[10]   INVIVO BEHAVIOR OF HYPODERMICALLY IMPLANTED MICROFABRICATED GLUCOSE SENSORS [J].
KOUDELKA, M ;
ROHNERJEANRENAUD, F ;
TERRETTAZ, J ;
BOBBIONIHARSCH, E ;
DEROOIJ, NF ;
JEANRENAUD, B .
BIOSENSORS & BIOELECTRONICS, 1991, 6 (01) :31-36