A simple reversed phase high-performance liquid chromatography method for polysorbate 80 quantitation in monoclonal antibody drug products

被引:35
作者
Adamo, Michael [3 ]
Dick, Lawrence W., Jr. [1 ]
Qiu, Difei [1 ]
Lee, An-Horng [1 ]
Devincentis, John [2 ]
Cheng, Kuang-Chuan [1 ]
机构
[1] Medarex Inc, Analyt Dev, Bloomsbury, NJ 08804 USA
[2] Bristol Myers Squibb Co, Syracuse, NY 13057 USA
[3] Bristol Myers Squibb Co, Pennington, NJ 08534 USA
来源
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES | 2010年 / 878卷 / 21期
关键词
Polysorbate; 80; Tween-80; RP-H PLC; Oleic acid; Surfactant; LIGHT-SCATTERING DETECTION; MASS-SPECTROMETRY; DOCETAXEL; VEHICLE;
D O I
10.1016/j.jchromb.2010.04.039
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
In this paper, we discuss an improved high-performance liquid chromatography (HPLC) method for the quantitation of polysorbate 80 (polyoxyethylenesorbitan monooleate), a commonly used stabilizing excipient in therapeutic drug solutions. This method is performed by quantitation of oleic acid, a hydrolysis product of polysorbate 80. Using base hydrolysis, polysorbate 80 releases the oleic acid at a 1:1 molar ratio. The oleic acid can then be separated from other polysorbate 80 hydrolysis products and matrices using reversed phase HPLC. The oleic acid is monitored without derivatization using the absorbance at 195 nm. The method was validated and also shown to be accurate for the quantitation of polysorbate 80 in a high protein concentration monoclonal antibody drug product. For the measured polysorbate 80 concentrations, the repeatability was less than 6.2% relative standard deviation of the mean (% RSD) with the day-to-day intermediate precision being less than 8.2% RSD. The accuracy of the oleic acid quantitation averaged 94-109% in different IgG(1) and IgG(4) drug solutions with variable polysorbate 80 concentrations. In this study, polyoxyethylene, a by-product of the polysorbate 80 hydrolysis was also identified. This peak was not identified by previous methods and also increased proportionally to the polysorbate 80 concentration. We have developed and qualified a method which uses common equipment found in most laboratories and is usable over a range of monoclonal antibody subclasses and protein concentrations. (C) 2010 Elsevier B.V. All rights reserved.
引用
收藏
页码:1865 / 1870
页数:6
相关论文
共 10 条
[1]   Simultaneous analysis of docetaxel and the formulation vehicle polysorbate 80 in human plasma by liquid chromatography/tandem mass spectrometry [J].
Baker, SD ;
Zhao, M ;
He, P ;
Carducci, MA ;
Verweij, J ;
Sparreboom, A .
ANALYTICAL BIOCHEMISTRY, 2004, 324 (02) :276-284
[2]   Quantitation of polysorbate 20 in protein solutions using mixed-mode chromatography and evaporative light scattering detection [J].
Hewitt, Daniel ;
Zhang, Taylor ;
Kao, Yung-Hsiang .
JOURNAL OF CHROMATOGRAPHY A, 2008, 1215 (1-2) :156-160
[3]   High-performance liquid chromatographic determination of polysorbate 80 in pharmaceutical suspensions [J].
Hu, M ;
Niculescu, M ;
Zhang, XM ;
Hui, A .
JOURNAL OF CHROMATOGRAPHY A, 2003, 984 (02) :233-236
[4]  
JI S, 2006, WCBP 2006 10 S INT R
[5]   Analysis methods of polysorbate 20: A new method to assess the stability of polysorbate 20 and established methods that may overlook degraded polysorbate 20 [J].
Khossravi, M ;
Kao, YH ;
Mrsny, RJ ;
Sweeney, TD .
PHARMACEUTICAL RESEARCH, 2002, 19 (05) :634-639
[6]   Determination of polysorbate 80 in parenteral formulations by high-performance liquid chromatography and evaporative light scattering detection [J].
Nair, LM ;
Stephens, NV ;
Vincent, S ;
Raghavan, N ;
Sand, PJ .
JOURNAL OF CHROMATOGRAPHY A, 2003, 1012 (01) :81-86
[7]   Determination of the docetaxel vehicle, polysorbate 80, in patient samples by liquid chromatography-tandem mass spectrometry [J].
Sparreboom, A ;
Zhao, M ;
Brahmer, JR ;
Verweij, J ;
Baker, SD .
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2002, 773 (02) :183-190
[8]   Single step method for the accurate concentration determination of polysorbate 80 [J].
Tani, TH ;
Moore, JM ;
Patapoff, TW .
JOURNAL OF CHROMATOGRAPHY A, 1997, 786 (01) :99-106
[9]  
TRACY S, 2008, AAPS J, V10, P647
[10]  
ZSOLT O, 1998, J PHARM BIOMED ANAL, V18, P715