Optimum detection times for bacteria and yeast species with the BACTEC 9120 aerobic blood culture system:: Evaluation for a 5-year period in a Turkish university hospital

被引:43
作者
Durmaz, G [1 ]
Us, T [1 ]
Aydinli, A [1 ]
Kiremitci, A [1 ]
Kiraz, N [1 ]
Akgün, Y [1 ]
机构
[1] Osmangazi Univ, Med Fac Hosp, Dept Clin Microbiol, TR-26480 Eskisehir, Turkey
关键词
D O I
10.1128/JCM.41.2.819-821.2003
中图分类号
Q93 [微生物学];
学科分类号
071005 ; 100705 ;
摘要
We tracked and documented the time of positivity of blood cultures by using the BACTEC 9120 (Becton Dickinson Diagnostic Instrument Systems) blood culture system over a 5-year study period. A 7-day protocol of the incubation period was selected, and a total of 11,156 blood cultures were evaluated. The clinically significant microorganisms (32.95%) were isolated in 3,676 specimens. Gram-positive and -negative bacterial isolation rates were found to be 41.07 and 44.88%, respectively. Yeasts were found in 14.03% of all pathogens. Both the false-positivity and -negativity rates were very low (0.1 and 0.3%, respectively). The mean detection times for all of the pathogens were determined to be 19.45 h. Yeasts, nonfermentative gram-negative bacteria, and Brucella melitensis strains were isolated within 5 days. By taking these data into account, we decided to establish a 5-day-incubation protocol in our laboratory instead of the 7 days that are commonly used.
引用
收藏
页码:819 / 821
页数:3
相关论文
共 17 条
[1]   Rapid diagnosis of Brucella bacteremia by using the BACTEC 9240 system [J].
Bannatyne, RM ;
Jackson, MC ;
Memish, Z .
JOURNAL OF CLINICAL MICROBIOLOGY, 1997, 35 (10) :2673-2674
[2]  
Bayram A, 1997, J TURK MICROBIOL SOC, V27, P64
[3]   Clinical comparison of BACTEC 9240 plus aerobic/F resin bottles and the isolator aerobic culture system for detection of bloodstream infections [J].
Cockerill, FR ;
Reed, GS ;
Hughes, JG ;
Torgerson, CA ;
Vetter, EA ;
Harmsen, WS ;
Dale, JS ;
Roberts, GD ;
Ilstrup, DM ;
Henry, NK .
JOURNAL OF CLINICAL MICROBIOLOGY, 1997, 35 (06) :1469-1472
[4]   Detection of Brucella melitensis by BACTEC NR 730 and BACTEC 9120 systems [J].
Gedikoglu, S ;
Helvaci, S ;
Ozakin, C ;
Gokirmak, F ;
Kihcturgay, K .
EUROPEAN JOURNAL OF EPIDEMIOLOGY, 1996, 12 (06) :649-650
[5]   Experience of changing between signal and Bactec 9240 blood culture systems in a children's hospital [J].
Gray, J ;
Brockwell, M ;
Das, I .
JOURNAL OF CLINICAL PATHOLOGY, 1998, 51 (04) :302-305
[6]  
Huang AH, 1998, EUR J CLIN MICROBIOL, V17, P637, DOI 10.1007/s100960050147
[7]   Multicenter clinical comparison of Resin-containing bottles with standard aerobic and anaerobic bottles for culture of microorganisms from blood [J].
Lelievre, H ;
Gimenez, M ;
Vandenesch, F ;
Reinhardt, A ;
Lenhardt, D ;
Just, HM ;
Pau, M ;
Ausina, V ;
Etienne, J .
EUROPEAN JOURNAL OF CLINICAL MICROBIOLOGY & INFECTIOUS DISEASES, 1997, 16 (09) :669-674
[8]   Laboratory diagnosis of bacteremia and fungemia [J].
Magadia, RR ;
Weinstein, MP .
INFECTIOUS DISEASE CLINICS OF NORTH AMERICA, 2001, 15 (04) :1009-+
[9]   Outpatient pediatric blood cultures: Time to positivity [J].
McGowan, KL ;
Foster, JA ;
Coffin, SE .
PEDIATRICS, 2000, 106 (02) :251-255
[10]  
RUITZ J, 1997, J CLIN MICROBIOL, V35, P2417