Invasive human magnetic resonance imaging: Feasibility during revascularization in a combined XMR suite

被引:38
作者
Dick, AJ
Raman, VK
Raval, AN
Guttman, MA
Thompson, RB
Ozturk, C
Peters, DC
Stine, AM
Wright, VJ
Schenke, WH
Lederman, RJ
机构
[1] NHLBI, Div Intramural Res, Cardiovasc Branch, NIH, Bethesda, MD 20892 USA
[2] NHLBI, Cardiac Energet Lab, Bethesda, MD 20892 USA
关键词
interventional magnetic resonance imaging; angioplasty; invasive imaging; catheterization; peripheral artery disease;
D O I
10.1002/ccd.20302
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
We tested the feasibility and safety of invasive magnetic resonance imaging (MRI) during peripheral angioplasty. Real-time MRI can image soft tissue and may potentially guide therapeutic procedures without ionizing radiation or nephrotoxic contrast. MRI-guided diagnostic catheterization has been described recently, but safe and conspicuous catheter devices are not widely available. An active guidewire, which serves as an MRI receiver antenna, might be useful to guide catheterization or even to image atheroma. We describe a combined interventional suite offering both X-ray fluoroscopy and real-time MRI. We used a 0.030" active guidewire receiver coil for invasive MRI after X-ray lesion traversal in patients undergoing percutaneous iliofemoral artery revascularization. Intravascular MRI was compared with noninvasive MRI, X-ray angiography, and intravascular ultrasound (IVUS). Seven eligible patients consented to participate, but three were excluded because of lengthy revascularization procedures. Four remaining patients safely underwent combined X-ray fluoroscopy and real-time magnetic resonance imaging (XMR) transport, continuous monitoring, and all imaging modalities. There was no device dislodgment, contamination or evidence of heating. The intravascular MRI coil was well visualized except at the tip, but did not provide superior mural imaging compared with IVUS. Therefore, because an adequate safety and workflow experience was obtained, enrollment was terminated after only four subjects. Invasive MRI is feasible and apparently safe during peripheral angioplasty. Patients can safely be transported and monitored in an XMR interventional suite. An active quarter-wavelength guidewire coil does not provide superior imaging compared with IVUS, but provides satisfactory guidewire visualization. These tools may prove useful for advanced therapeutic procedures in the future.
引用
收藏
页码:265 / 274
页数:10
相关论文
共 25 条
[1]   Endovascular interventional magnetic resonance imaging [J].
Bartels, LW ;
Bakker, CJG .
PHYSICS IN MEDICINE AND BIOLOGY, 2003, 48 (14) :R37-R64
[2]   Initial experiences with in vivo intravascular coronary vessel wall imaging [J].
Botnar, RM ;
Bücker, A ;
Kim, WY ;
Viohl, I ;
Günther, RW ;
Spuentrup, E .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2003, 17 (05) :615-619
[3]   Magnetic resonance-guided placement of atrial septal closure device in animal model of patent foramen ovale [J].
Buecker, A ;
Spuentrup, E ;
Grabitz, R ;
Freudenthal, F ;
Muehler, EG ;
Schaeffter, T ;
van Vaals, JJ ;
Günther, RW .
CIRCULATION, 2002, 106 (04) :511-515
[4]   Magnetic resonance fluoroscopy allows targeted delivery of mesenchymal stem cells to infarct borders in swine [J].
Dick, AJ ;
Guttman, MA ;
Raman, VK ;
Peters, DC ;
Pessanha, BSS ;
Hill, JM ;
Smith, S ;
Scott, G ;
McVeigh, ER ;
Lederman, RJ .
CIRCULATION, 2003, 108 (23) :2899-2904
[5]   Real-time volume rendered MRI for interventional guidance [J].
Guttman, MA ;
Lederman, RJ ;
Sorger, JM ;
McVeigh, ER .
JOURNAL OF CARDIOVASCULAR MAGNETIC RESONANCE, 2002, 4 (04) :431-U1
[6]   Active device tracking and high-resolution intravascular MRI using a novel catheter-based, opposed-solenoid phased array coil [J].
Hillenbrand, CM ;
Elgort, DR ;
Wong, EY ;
Reykowski, A ;
Wacker, FK ;
Lewin, JS ;
Duerk, JL .
MAGNETIC RESONANCE IN MEDICINE, 2004, 51 (04) :668-675
[7]   In vivo intravascular MR imaging: Transvenous technique for arterial wall Imaging [J].
Hofmann, LV ;
Lidell, RP ;
Arepally, A ;
Montague, B ;
Yang, XM ;
Bluemke, DA .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2003, 14 (10) :1317-1327
[8]   Active MR visualization of a vascular guidewire in vivo [J].
Ladd, ME ;
Zimmermann, GG ;
Quick, HH ;
Debatin, JF ;
Boesiger, P ;
von Schulthess, GK ;
McKinnon, GC .
JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1998, 8 (01) :220-225
[9]   Catheter-based endomyocardial injection with real-time magnetic resonance imaging [J].
Lederman, RJ ;
Guttman, MA ;
Peters, DC ;
Thompson, RB ;
Sorger, JM ;
Dick, AJ ;
Raman, VK ;
McVeigh, ER .
CIRCULATION, 2002, 105 (11) :1282-1284
[10]   An expandable intravenous RF coil for arterial wall imaging [J].
Martin, AJ ;
McLoughlin, RF ;
Chu, KC ;
Barberi, EA ;
Rutt, BK .
JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1998, 8 (01) :226-234