Active device tracking and high-resolution intravascular MRI using a novel catheter-based, opposed-solenoid phased array coil

被引:71
作者
Hillenbrand, CM
Elgort, DR
Wong, EY
Reykowski, A
Wacker, FK
Lewin, JS
Duerk, JL
机构
[1] Univ Hosp Cleveland, Dept Radiol, MRI, Cleveland, OH 44106 USA
[2] Case Western Reserve Univ, Dept Biomed Engn, Cleveland, OH 44106 USA
关键词
intravascular MRI; catheter RF coils; vessel wall imaging; interventional device tracking; phased array;
D O I
10.1002/mrm.20050
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A novel two-element, catheter-based phased array coil was designed and built for both active MR device tracking and high-resolution vessel wall imaging. The device consists of two independent solenoid coils that are wound in opposite directions, connected to separate receive channels, and mounted collinearly on an angiographic catheter. The elements were used independently or together for tracking or imaging applications, respectively. The array's dual functionality was tested on a clinical 1.5 T MRI scanner in vitro, in vivo, and in situ. During real-time catheter tracking, each element gave rise to a high-amplitude peak in the respective projection data, which enabled reliable and robust device tracking as well as automated slice positioning. In vivo microimaging with 240 mum inplane resolution was achieved in 9 s using the device and TrueFISP imaging. Therefore, a single device was successfully implemented that met the combined requirements of intravascular device tracking and imaging. (C) 2004 Wiley-Liss, Inc.
引用
收藏
页码:668 / 675
页数:8
相关论文
共 34 条
[1]   High resolution intravascular MRI and MRS by using a catheter receiver coil [J].
Atalar, E ;
Bottomley, PA ;
Ocali, O ;
Correia, LCL ;
Kelemen, MD ;
Lima, JAC ;
Zerhouni, EA .
MAGNETIC RESONANCE IN MEDICINE, 1996, 36 (04) :596-605
[2]   Catheter-tracking FOV MR fluoroscopy [J].
Atalar, E ;
Kraitchman, DL ;
Carkhuff, B ;
Lesho, J ;
Ocali, O ;
Solaiyappan, M ;
Guttman, MA ;
Charles, HK .
MAGNETIC RESONANCE IN MEDICINE, 1998, 40 (06) :865-872
[3]   MR-guided endovascular interventions: Susceptibility-based catheter and near-real-time imaging technique [J].
Bakker, CJ ;
Hoogeveen, RM ;
Hurtak, WF ;
vanVaals, JJ ;
Viergever, MA ;
Mali, WPTM .
RADIOLOGY, 1997, 202 (01) :273-276
[4]   REMOTE (INSIDE-OUT) NMR .1. REMOTE PRODUCTION OF A REGION OF HOMOGENEOUS MAGNETIC-FIELD [J].
COOPER, RK ;
JACKSON, JA .
JOURNAL OF MAGNETIC RESONANCE, 1980, 41 (03) :400-405
[5]   Remember true FISP? A high SNR, near 1-second imaging method for T2-like contrast in interventional MRI at .2 T [J].
Duerk, JL ;
Lewin, JS ;
Wendt, M ;
Petersilge, C .
JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1998, 8 (01) :203-208
[6]   REAL-TIME POSITION MONITORING OF INVASIVE DEVICES USING MAGNETIC-RESONANCE [J].
DUMOULIN, CL ;
SOUZA, SP ;
DARROW, RD .
MAGNETIC RESONANCE IN MEDICINE, 1993, 29 (03) :411-415
[7]   Real-time catheter tracking and adaptive imaging [J].
Elgort, DR ;
Wong, EY ;
Hillenbrand, CM ;
Wacker, FK ;
Lewin, JS ;
Duerk, JL .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2003, 18 (05) :621-626
[8]   A method for fast 3D tracking using tuned fiducial markers and a limited projection reconstruction FISP (LPR-FISP) sequence [J].
Flask, C ;
Elgort, D ;
Wong, E ;
Shankaranarayanan, A ;
Lewin, J ;
Wendt, M ;
Duerk, JL .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2001, 14 (05) :617-627
[9]  
GINZTON EL, 1966, MICROWAVE MEASUREMEN, P254
[10]   Catheter visualization using locally induced, actively controlled field inhomogeneities [J].
Glowinski, A ;
Adam, G ;
Bucker, A ;
Neuerburg, J ;
vanVaals, JJ ;
Gunther, RW .
MAGNETIC RESONANCE IN MEDICINE, 1997, 38 (02) :253-258