Rhythm discrimination by rate branch and QRS morphology in dual chamber implantable cardioverter defibrillators

被引:19
作者
Boriani, G
Biffi, M
Dall'Acqua, A
Martignani, C
Frabetti, L
Zannoli, R
Branzi, A
机构
[1] Univ Bologna, Inst Cardiol, Azienda Osped S Orsola Malpighi, I-40138 Bologna, Italy
[2] St Jude Med Italia, Milan, Italy
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 2003年 / 26卷 / 01期
关键词
atrial fibrillation; cardioverter defibrillators; arrhythmia detection; supraventricular tachycardia; ventricular tachyarrhythmia;
D O I
10.1046/j.1460-9592.2003.00073.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Morphology Discrimination is a discriminator based on QRS morphology analysis that has been recently implemented in dual chamber implantable cardioverter defibrillators (ICDs). Detected events are initially classified according to median atrial and ventricular rates (Rate Branch). Then, a series of discriminators (Morphology Discrimination, Stability, Sudden Onset) analyze the rhythm according to specific criteria and the number of discriminators required for VT diagnosis (i.e., requiring "any" or "all" of the specific discriminators to indicate VT). The discriminating accuracy of the algorithm was evaluated in 645 detections recorded during the follow-up of 25 patients. The overall specificity for 397 supraventricular arrhythmias was 73.5% (292/397) with the tachycardia diagnosis criteria set to "any" and 90.9% (361/397) with the tachycardia diagnosis criteria set to "all." Sensitivity for VT was 100% and 98.7% (231/234) with the tachycardia diagnosis criteria set to "any" and "all," respectively. With the tachycardia diagnosis criteria set to "any," specificity for atrial fibrillation was 88.6%, for atrial flutter 40.3%, for atrial tachycardia 0%, and for sinus tachycardia 97.0%. With the tachycardia diagnosis criteria set to "all, " specificity for atrial fibrillation was 92.40%, for atrial flutter 93.5%, for atrial tachycardia 54.7%, and for sinus tachycardia 99.0%. The contribution of Morphology Discrimination was crucial to improve the specificity of the Rate Branch algorithm. The implementation of Morphology Discrimination in a dual chamber ICD with Rate Branch rhythm classification allows the attainment of high specificity and high sensitivity for ventricular tachyarrhythmias.
引用
收藏
页码:466 / 470
页数:5
相关论文
共 14 条
[1]   Combined use of morphology discrimination, sudden onset, and stability as discriminating algorithms in single chamber cardioverter defibrillators [J].
Boriani, G ;
Occhetta, E ;
Pistis, G ;
Menozzi, C ;
Jorfida, M ;
Sermasi, S ;
Pagani, M ;
Gasparini, G ;
Musso, G ;
Dall'Acqua, A ;
Biffi, M ;
Branzi, A .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2002, 25 (09) :1357-1366
[2]   Clinical evaluation of morphology discrimination: An algorithm for rhythm discrimination in cardioverter defibrillators [J].
Boriani, G ;
Biffi, M ;
Frabetti, L ;
Lattuca, JJ ;
Branzi, A .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2001, 24 (06) :994-1001
[3]   Do current dual chamber cardioverter defibrillators have advantages over conventional single chamber cardioverter defibrillators in reducing inappropriate therapies? A randomized, prospective study [J].
Deisenhofer, I ;
Kolb, C ;
Ndrepepa, G ;
Schreieck, J ;
Karch, M ;
Schmieder, S ;
Zrenner, B ;
Schmitt, C .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2001, 12 (02) :134-142
[4]  
Grönefeld GC, 2001, PACE, V24, P1519, DOI 10.1046/j.1460-9592.2001.01519.x
[5]   Stability: An ICD detection criterion for discriminating atrial fibrillation from ventricular tachycardia [J].
Higgins, SL ;
Lee, RS ;
Kramer, RL .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 1995, 6 (12) :1081-1088
[6]   Inappropriate detection of supraventricular arrhythmias by implantable dual chamber defibrillators: A comparison of four different algorithms [J].
Hintringer, F ;
Schwarzacher, S ;
Eibl, G ;
Pachinger, O .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2001, 24 (05) :835-841
[7]   Clinical experience with the new detection algorithms for atrial fibrillation of a defibrillator with dual chamber sensing and pacing [J].
Kühlkamp, V ;
Dörnberger, V ;
Mewis, C ;
Suchalla, R ;
Bosch, RF ;
Seipel, L .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 1999, 10 (07) :905-915
[8]   Inappropriate defibrillator therapies: Are dual chamber devices providing a remedy? [J].
Lee, KL ;
Lau, CP .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2001, 12 (02) :143-144
[9]  
NEUTZNER J, 1995, PACE, V18, P539
[10]   Enhanced detection criteria in implantable cardioverter-defibrillators to avoid inappropriate therapy [J].
Schaumann, A ;
vonzurMuhlen, F ;
Gonska, BD ;
Kreuzer, H .
AMERICAN JOURNAL OF CARDIOLOGY, 1996, 78 :42-50