Do current dual chamber cardioverter defibrillators have advantages over conventional single chamber cardioverter defibrillators in reducing inappropriate therapies? A randomized, prospective study

被引:126
作者
Deisenhofer, I
Kolb, C
Ndrepepa, G
Schreieck, J
Karch, M
Schmieder, S
Zrenner, B
Schmitt, C
机构
[1] Deutsch Herzzentrum Munchen, D-80636 Munich, Germany
[2] Tech Univ Munich, Klinikum Rechts Isar, Med Klin 1, D-8000 Munich, Germany
关键词
dual chamber implantable cardioverter defibrillator; detection algorithm; inappropriate therapy; atrial fibrillation;
D O I
10.1046/j.1540-8167.2001.00134.x
中图分类号
R5 [内科学];
学科分类号
1002 [临床医学]; 100201 [内科学];
摘要
Dual Chamber vs Single Chamber Cardioverter Defibrillators. Introduction: Supraventricular tachyarrhythmias are the main cause of inappropriate therapies in patients with conventional single chamber implantable cardioverter defibrillators (VVI-ICD). It was anticipated that dual chamber cardioverter defibrillators (DDD-ICD), with their capacity to analyze atrial and ventricular rhythm, could substantially reduce inappropriate therapies. Methods and Results: Our prospective study included 92 patients (87 men; mean age 61 +/- 12.7 years) who were randomly assigned to a VVI-ICD (45 patients) or a DDD-ICD (47 patients). Both groups were followed for 7.5 +/- 3.5 and 7.6 +/- 4.1 months, respectively, During the follow-up period, overall 725 ventricular tachycardia (VT)/ventricular fibrillation (VF) episodes were recorded in 45 (49%) of 92 patients. Of these episodes, 404 (56%) occurred in the VVI-ICD group and 321 (44%) episodes occurred in the DDD-ICD group. Twenty-three (51%) patients in the VVI-ICD group and 22 (47%) patients in the DDD-ICD group (P = 0.8) developed VT/VF, Overall, 73 (10%) of 725 treated episodes were inappropriate in 6 (13%) patients in the VVI group and in 10 (21%) patients in the DDD-ICD group (P = 0.2), There were 22 (31%) inappropriately treated episodes in the VVI-ICD group and 51 (69%) in the DDD-ICD group. Thirty-two of the 51 inappropriate episodes in the DDD-ICD patients resulted from intermittent atrial sensing problems that led to failure of the respective dual chamber algorithms. Nonfatal complications occurred in 6 (13%) patients in the VVI-ICD group and in 3 (6%) patients in the DDD-ICD group (P = 07). Conclusions: We conclude that the implanted DDD-ICD and conventional VVI-ICD are equally safe and effective for therapy of life-threatening ventricular tachyarrhythmias, Although DDD-ICDs allow better rhythm classification, the applied detection algorithms do not offer benefits in avoiding inappropriate therapies during supraventricular tachyarrhythmias.
引用
收藏
页码:134 / 142
页数:9
相关论文
共 27 条
[1]
PROSPECTIVE RANDOMIZED TRIAL OF ATRIAL VERSUS VENTRICULAR PACING IN SICK-SINUS SYNDROME [J].
ANDERSEN, HR ;
THUESEN, L ;
BAGGER, JP ;
VESTERLUND, T ;
THOMSEN, PEB .
LANCET, 1994, 344 (8936) :1523-1528
[2]
Inappropriate shocks delivered by implantable cardiac defibrillators during oversensing of activity of diaphagmatic muscle [J].
Babuty, D ;
Fauchier, L ;
Cosnay, P .
HEART, 1999, 81 (01) :94-96
[3]
The potential usage of dual chamber pacing in patients with implantable cardioverter defibrillators [J].
Best, PJM ;
Hayes, DL ;
Stanton, MS .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1999, 22 (01) :79-85
[4]
HOLTER DOCUMENTED SUDDEN-DEATH IN A PATIENT WITH AN IMPLANTED DEFIBRILLATOR [J].
BIRGERSDOTTERGREEN, U ;
ROSENQVIST, M ;
LINDEMANS, FW ;
RYDEN, L ;
RADEGRAN, K .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1992, 15 (07) :1008-1014
[5]
BROUWER J, 1997, PACE, V20, P16
[6]
Myopotential sensing by a dual chamber implantable cardioverter defibrillator: Two case reports [J].
Deshmukh, P ;
Anderson, K .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 1998, 9 (07) :767-772
[7]
Variability of atrial sensing during exercise in a patient with a dual chamber ICD caused inappropriate ICD discharges [J].
Glotzer, TV ;
Radoslovich, GA ;
Neglia, J ;
Zimmerman, JM .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1999, 22 (12) :1838-1841
[8]
ELECTROCARDIOGRAPHICALLY DOCUMENTED UNNECESSARY, SPONTANEOUS SHOCKS IN 241 PATIENTS WITH IMPLANTABLE CARDIOVERTER DEFIBRILLATORS [J].
GRIMM, W ;
FLORES, BF ;
MARCHLINSKI, FE .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1992, 15 (11) :1667-1673
[9]
DELETERIOUS EFFECTS OF LONG-TERM SINGLE-CHAMBER VENTRICULAR PACING IN PATIENTS WITH SICK SINUS SYNDROME - THE HIDDEN BENEFITS OF DUAL-CHAMBER PACING [J].
HESSELSON, AB ;
PARSONNET, V ;
BERNSTEIN, AD ;
BONAVITA, GJ .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1992, 19 (07) :1542-1549
[10]
Discrimination of sinus rhythm, atrial flutter, and atrial fibrillation using bipolar endocardial signals [J].
Jung, J ;
Hohenberg, G ;
Heisel, A ;
Strauss, D ;
Schieffer, H ;
Fries, R .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 1998, 9 (07) :689-695