Inappropriate shocks delivered by implantable cardiac defibrillators during oversensing of activity of diaphagmatic muscle

被引:20
作者
Babuty, D [1 ]
Fauchier, L [1 ]
Cosnay, P [1 ]
机构
[1] Hosp Trousseau, Dept Cardiol B, F-37044 Tours, France
关键词
implantable cardiac defibrillator; inappropriate shocks; myopotentials;
D O I
10.1136/hrt.81.1.94
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Two cases are reported (both men, one 72 and one 54 years old) of inappropriate shocks delivered by an implantable cardiac defibrillator (ICD) device, which oversensed the myopotentials induced by deep breathing and Valsahra manoeuvre. No damage to leads was associated with the oversensing of myopotentials. The mechanism of the inappropriate shocks was determined using real time electrograms. Modification of the duration of ventricular detection and decrease in sensitivity made it possible to avoid the oversensing of myopotentials and to deliver ICD treatment. potentials and to deliver ICD treatment.
引用
收藏
页码:94 / 96
页数:3
相关论文
共 8 条
[1]   Inhibition of bradycardia pacing caused by far-field atrial sensing in a third-generation cardioverter defibrillator with an automatic gain feature [J].
Curwin, JH ;
Roelke, M ;
Ruskin, JN .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1996, 19 (01) :124-126
[2]   CARDIOVERTER DISCHARGES FOLLOWING SENSING OF ELECTRICAL ARTIFACT DUE TO FLUID PENETRATION IN THE CONNECTOR PORT [J].
HIEF, C ;
PODCZECK, A ;
FROHNER, K ;
NURNBERG, M ;
KALTENBRUNNER, W ;
VEIT, F ;
STEINBACH, K .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1995, 18 (08) :1589-1591
[3]   INTERFERENCE FROM A HAND-HELD RADIOFREQUENCY REMOTE-CONTROL CAUSING DISCHARGE OF AN IMPLANTABLE DEFIBRILLATOR [J].
MAN, KC ;
DAVIDSON, T ;
LANGBERG, JJ ;
MORADY, F ;
KALBFLEISCH, SJ .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1993, 16 (08) :1756-1758
[4]   Interaction between electronic article surveillance systems and implantable defibrillators: Insights from a fourth generation ICD [J].
Mathew, P ;
Lewis, C ;
Neglia, J ;
Krol, RB ;
Saksena, S .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1997, 20 (11) :2857-2859
[5]   LIMITATIONS AND LATE COMPLICATIONS OF 3RD-GENERATION AUTOMATIC CARDIOVERTER-DEFIBRILLATORS [J].
NUNAIN, SO ;
ROELKE, M ;
TROUTON, T ;
OSSWALD, S ;
KIM, YH ;
SOSASUAREZ, G ;
BROOKS, DR ;
MCGOVERN, B ;
GUY, M ;
TORCHIANA, DF ;
VLAHAKES, GJ ;
GARAN, H ;
RUSKIN, JN .
CIRCULATION, 1995, 91 (08) :2204-2213
[6]   THE PROARRHYTHMIC POTENTIAL OF IMPLANTABLE CARDIOVERTER-DEFIBRILLATORS [J].
PINSKI, SL ;
FAHY, GJ .
CIRCULATION, 1995, 92 (06) :1651-1664
[7]   INAPPROPRIATE DISCHARGE BY AN IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR - RECOGNITION OF MYOPOTENTIAL SENSING USING TELEMETERED INTRACARDIAC ELECTROGRAMS [J].
SANDLER, MT ;
KUTALEK, SP .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1994, 17 (04) :665-671
[8]   ERRONEOUS DISCHARGE OF AN IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR CAUSED BY AN ELECTRIC RAZOR [J].
SEIFERT, T ;
BLOCK, M ;
BORGGREFE, M ;
BREITHARDT, G .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1995, 18 (08) :1592-1594