Pulse oximetry is a poor predictor of hypoxemia in stable children with sickle cell disease

被引:51
作者
Blaisdell, CJ
Goodman, S
Clark, K
Casella, JF
Loughlin, GM
机构
[1] Johns Hopkins Med Inst, Dept Pediat, Eudowood Div Resp Sci, Baltimore, MD 21205 USA
[2] Johns Hopkins Med Inst, Dept Oncol, Baltimore, MD 21205 USA
[3] Johns Hopkins Med Inst, Div Pediat Hematol, Baltimore, MD 21205 USA
来源
ARCHIVES OF PEDIATRICS & ADOLESCENT MEDICINE | 2000年 / 154卷 / 09期
关键词
D O I
10.1001/archpedi.154.9.900
中图分类号
R72 [儿科学];
学科分类号
100202 ;
摘要
Objective: To evaluate the accuracy of the pulse oximeter to detect hypoxemia in patients with sickle cell disease in an ambulatory care setting. Study Design: Simultaneous measurements of Pao(2), arterial oxygen saturation by co-oximetry (criterion standard), and pulse oximetry were performed in 21 children with sickle cell disease during 22 outpatient visits. The bias and precision of the pulse oximeter compared with measured arterial oxygen saturation by co-oximetry were determined. The sensitivity, specificity, and positive and negative predictive values of the pulse oximeter to detect hypoxemia (PaO2 <70 mm Hg) were also calculated. Results: The mean difference between pulse oximetry and measured oxygen saturation (bias) was 5.0% and the SD (precision) was 5.3. Twenty-one patients had a PaO2 greater than 70 mm Hg; 7 of these (33%) were predicted to be hypoxic by pulse oximetry with values less than 93%, for a specificity to detect normoxia of 67%. Conclusion: Making treatment decisions based on pulse oximetry data alone in patients with sickle cell disease who are not acutely ill may be inappropriate.
引用
收藏
页码:900 / 903
页数:4
相关论文
共 17 条
[1]   OXYGEN DISSOCIATION CURVES IN SICKLE CELL ANEMIA AND IN SUBJECTS WITH THE SICKLE CELL TRAIT [J].
BECKLAKE, MR ;
GRIFFITHS, SB ;
MCGREGOR, M ;
GOLDMAN, HI ;
SCHREVE, JP .
JOURNAL OF CLINICAL INVESTIGATION, 1955, 34 (05) :751-755
[2]   Does sleep-disordered breathing contribute to the clinical severity of sickle cell anemia? [J].
Brooks, LJ ;
Koziol, SM ;
Chiarucci, KM ;
Berman, BW .
JOURNAL OF PEDIATRIC HEMATOLOGY ONCOLOGY, 1996, 18 (02) :135-139
[3]  
CALVIN JE, 1992, CAN MED ASSOC J, V146, P703
[4]   Evaluation of pulse oximetry in sickle cell anemia patients presenting to the emergency department in acute vasoocclusive crisis [J].
Comber, JT ;
Lopez, BL .
AMERICAN JOURNAL OF EMERGENCY MEDICINE, 1996, 14 (01) :16-18
[5]   COMPARISON OF OXYGENATION MEASUREMENTS IN PEDIATRIC-PATIENTS DURING SICKLE-CELL CRISES [J].
CRAFT, JA ;
ALESSANDRINI, E ;
KENNEY, LB ;
KLEIN, B ;
BRAY, G ;
LUBAN, NLC ;
MEEK, R ;
NADKARNI, VM .
JOURNAL OF PEDIATRICS, 1994, 124 (01) :93-95
[6]  
GOEPP J, 1991, PEDIATR EMERG CARE, V7, P387
[7]   PULSE OXIMETRY IN NEONATAL MEDICINE [J].
HAY, WW ;
THILO, E ;
CURLANDER, JB .
CLINICS IN PERINATOLOGY, 1991, 18 (03) :441-472
[8]   Accuracy of pulse oximetry in sickle cell disease [J].
Ortiz, FO ;
Aldrich, TK ;
Nagel, RL ;
Benjamin, LJ .
AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 1999, 159 (02) :447-451
[9]   PULSE OXIMETRY IN SICKLE-CELL DISEASE [J].
PIANOSI, P ;
CHARGE, TD ;
ESSELTINE, DW ;
COATES, AL .
ARCHIVES OF DISEASE IN CHILDHOOD, 1993, 68 (06) :735-738
[10]  
POETS CF, 1994, PEDIATRICS, V93, P737