COMPARISON OF OXYGENATION MEASUREMENTS IN PEDIATRIC-PATIENTS DURING SICKLE-CELL CRISES

被引:29
作者
CRAFT, JA
ALESSANDRINI, E
KENNEY, LB
KLEIN, B
BRAY, G
LUBAN, NLC
MEEK, R
NADKARNI, VM
机构
[1] MED CTR DELAWARE, DEPT PEDIAT, NEWARK, DE 19718 USA
[2] GEORGE WASHINGTON UNIV, MED CTR, CHILDRENS NATL MED CTR, WASHINGTON, DC 20037 USA
关键词
D O I
10.1016/S0022-3476(94)70260-8
中图分类号
R72 [儿科学];
学科分类号
100202 ;
摘要
Measurements of the saturation of arterial blood with oxygen (Sao(2)) were compared in 24 children during sickle cell crises. Simultaneous pulse oximetry (Nellcor N-100 pulse oximeter) and arterial blood analysis showed that Sao(2) measured by pulse oximetry overestimated cooximeter-measured Sao(2) (mean bias, 6.9%; p < 0.001). The blood gas machine-calculated Sao(2) also overestimated cooximeter-measured Sao(2)(p < 0.001). The bias increased with increasing age (p = 0.002) and carboxyhemoglobin level (p = 0.005) but was not related to methemoglobin, total hemoglobin, percentage of hemoglobin S, or percentage of hemoglobin F.
引用
收藏
页码:93 / 95
页数:3
相关论文
共 15 条
[1]   THE EFFECT OF CARBON-MONOXIDE INHALATION ON PULSE OXIMETRY AND TRANS-CUTANEOUS PO2 [J].
BARKER, SJ ;
TREMPER, KK .
ANESTHESIOLOGY, 1987, 66 (05) :677-679
[2]   AGREEMENT BETWEEN NONINVASIVE OXIMETRIC VALUES FOR OXYGEN-SATURATION [J].
CAHAN, C ;
DECKER, MJ ;
HOEKJE, PL ;
STROHL, KP .
CHEST, 1990, 97 (04) :814-819
[3]  
CHARACHE S, 1992, NIH922117 US DEP HLT
[4]   EFFECTS OF OXYGEN INHALATION ON ENDOGENOUS ERYTHROPOIETIN KINETICS, ERYTHROPOIESIS, AND PROPERTIES OF BLOOD-CELLS IN SICKLE-CELL-ANEMIA [J].
EMBURY, SH ;
GARCIA, JF ;
MOHANDAS, N ;
PENNATHURDAS, R ;
CLARK, MR .
NEW ENGLAND JOURNAL OF MEDICINE, 1984, 311 (05) :291-295
[5]  
ENGLE RR, 1971, PEDIATRICS, V47, P723
[6]   OXIMETRY IN CHILDREN RECOVERING FROM DEEP HYPOTHERMIA FOR CARDIAC-SURGERY [J].
MACNAB, AJ ;
BAKERBROWN, G ;
ANDERSON, EE .
CRITICAL CARE MEDICINE, 1990, 18 (10) :1066-1069
[7]   BIAS AND PRECISION OF PULSE OXIMETERS AND ARTERIAL OXIMETERS [J].
NICKERSON, BG ;
SARKISIAN, C ;
TREMPER, K .
CHEST, 1988, 93 (03) :515-517
[8]   PULSE OXIMETRY IN SICKLE-CELL DISEASE [J].
PIANOSI, P ;
CHARGE, TD ;
ESSELTINE, DW ;
COATES, AL .
ARCHIVES OF DISEASE IN CHILDHOOD, 1993, 68 (06) :735-738
[9]  
RACKOFF WR, 1993, BLOOD, V81, P3422