Reduction in hemoglobin A1c with real-time continuous glucose monitoring: Results from a 12-week observational study

被引:113
作者
Bailey, Timothy S.
Zisser, Howard C.
Garg, Satish K.
机构
[1] Univ Colorado Denver, Barbara Davis Ctr Diabet, Aurora, CO 80010 USA
[2] N Cty Endocrine Med Grp, Escondilo, CA USA
[3] Sansum Diabet Res Inst, Santa Barbara, CA USA
关键词
D O I
10.1089/dia.2007.0205
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: Real-time continuous glucose monitoring (CGM) was studied in 140 adults with diabetes over a 12-week period of home use. Hemoglobin A(1c) (HbA1c) was measured on day I (baseline) and at weeks 6 and 12. Methods: On day 1, participants received the CGM device (STS (R) System, DexCom, Inc., San Diego, CA) and underwent training on proper use. Insertion of the first sensor was performed under staff supervision. Subjects inserted subsequent sensors on their own. After calibration, the device (a 3-day sensor, receiver, and transmitter) provided users with real-time glucose values updated at 5-min intervals, glucose trend graphs, configurable high/low alerts, and a hypoglycemia alarm (<= 55 mg/dL). Study participants were given supplies sufficient for 3 weeks of device use. Follow-up visits were performed at 3-week intervals for resupply and to download CGM data, with a final visit at the end of week 12. Results: Overall, a reduction in HbA1c of 0.4 +/- 0.05% (least squares mean +/- SE) was observed, P < 0.0001. Significant HbA1c reductions were observed across subgroups of subjects with both type I and 2 diabetes, and those delivering insulin by multiple daily injections and pumps. The largest HbA1c reduction (1.4 +/- 0.4%) was observed in subjects with baseline HbA1c > 9.0%. Increased CGM use was associated with greater reductions in HbAlc. Conclusions: This observational study showed that home use of real-time GCM was safe and well tolerated and associated with a clinically and statistically significant reduction in HbAlc. Large-scale randomized, controlled outcome studies of CGM are indicated.
引用
收藏
页码:203 / 210
页数:8
相关论文
共 26 条
[1]  
[Anonymous], 1995, DIABETES, V44, P968
[2]   Continuous glucose monitoring used to adjust diabetes therapy improves glycosylated hemoglobin: a pilot study [J].
Bode, BW ;
Gross, TM ;
Thornton, KR ;
Mastrototaro, JJ .
DIABETES RESEARCH AND CLINICAL PRACTICE, 1999, 46 (03) :183-190
[3]   Glycemic variability:: A hemoglobin A1c-independent risk factor for diabetic complications [J].
Brownlee, M ;
Hirsch, IB .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 2006, 295 (14) :1707-1708
[4]   Use of the GlucoWatch biographer in children with type 1 diabetes [J].
Chase, HP ;
Roberts, MD ;
Wightman, C ;
Klingensmith, G ;
Garg, SK ;
Van Wyhe, M ;
Desai, S ;
Harper, W ;
Lopatin, M ;
Bartkowiak, M ;
Tamada, J ;
Eastman, RC .
PEDIATRICS, 2003, 111 (04) :790-794
[5]   Continuous subcutaneous glucose monitoring in children with type 1 diabetes [J].
Chase, HP ;
Kim, LM ;
Owen, SL ;
MacKenzie, TA ;
Klingensmith, GJ ;
Murtfeldt, R ;
Garg, SK .
PEDIATRICS, 2001, 107 (02) :222-226
[6]   The continuous glucose monitoring system is useful for detecting unrecognized hypoglycemias in patients with type 1 and type 2 diabetes but is not better than frequent capillary glucose measurements for improving metabolic control [J].
Chico, A ;
Vidal-Ríos, P ;
Subirà, M ;
Novials, A .
DIABETES CARE, 2003, 26 (04) :1153-1157
[7]   Hypoglycemia in diabetes [J].
Cryer, PE ;
Davis, SN ;
Shamoon, H .
DIABETES CARE, 2003, 26 (06) :1902-1912
[8]   Improved glycemic control in poorly controlled patients with type 1 diabetes using real-time continuous glucose monitoring [J].
Deis, Dorothee ;
Bolinder, Jan ;
Riveline, Jean-Pierre ;
Battelino, Tadej ;
Bosi, Emanuele ;
Tubiana-Rufi, Nadia ;
Kerr, David ;
Phillip, Moshe .
DIABETES CARE, 2006, 29 (12) :2730-2732
[9]  
*DIAB RES CHILDR N, 2005, DIABETES CARE, V28, P1101
[10]   Improvement in glycemic excursions with a transcutaneous, real-time continuous glucose sensor [J].
Garg, S ;
Zisser, H ;
Schwartz, S ;
Bailey, T ;
Kaplan, R ;
Ellis, S ;
Jovanovic, L .
DIABETES CARE, 2006, 29 (01) :44-50