Analysis of the Look-Locker T1 mapping sequence in dynamic contrast uptake studies:: simulation and in vivo validation

被引:21
作者
Karlsson, M [1 ]
Nordell, B
机构
[1] Karolinska Inst, MR Res Ctr, Stockholm, Sweden
[2] Karolinska Inst, Dept Med Radiat Phys, Stockholm, Sweden
关键词
dynamic MRI; LL-EPI; T-1; mapping;
D O I
10.1016/S0730-725X(00)00193-4
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 [临床医学]; 100207 [影像医学与核医学]; 1009 [特种医学];
摘要
An alternative to the pulse sequences at present used in dynamic contrast uptake MRI is the dynamic LL-EPI T-1 mapping method. This method generates T-1 estimates in a few seconds, thereby allowing dynamic studies. A particular advantage of the LL-EPI technique is that it provides the opportunity to generate spatial and temporal information about the paramagnetic contrast agent concentration independently of the inflow rate. This paper illustrates, by computer simulations, the accuracy of the estimated 1/T-1 value when using the LL-EPI technique in situations that are not supported by the model. The simulated situations not supported by the model are those in which the longitudinal and transversal relaxation rates change during the T-1 mapping. The most critical moment occurs during a bolus passage of contrast agent when the concentration gradient is large. The computer simulations of the LL-EPI T-1 mapping method in non-supported situations show that in normal perfused capillary tissue the error in the estimated 1/T-1 value is within the absolute error of 0.1 s(-1) in most simulated situations, although in a typical vessel the simulations do indicate that the stated absolute error tolerance of 0.5 s(-1) is exceeded relatively easily. However, this transgression can be rectified by a non-bolus injection of the contrast agent media. (C) 2000 Elsevier Science Inc. All rights reserved.
引用
收藏
页码:947 / 954
页数:8
相关论文
共 18 条
[1]
HYDRATION OF COMPLEXONE COMPLEXES OF LANTHANIDE CATIONS [J].
ALSAADI, BM ;
ROSSOTTI, FJC ;
WILLIAMS, RJP .
JOURNAL OF THE CHEMICAL SOCIETY-DALTON TRANSACTIONS, 1980, (11) :2151-2154
[2]
Water diffusion and exchange as they influence contrast enhancement [J].
Donahue, KM ;
Weisskoff, RM ;
Burstein, D .
JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1997, 7 (01) :102-110
[3]
Optimization of the ultrafast Look-Locker echo-planar imaging T1 mapping sequence [J].
Freeman, AJ ;
Gowland, PA ;
Mansfield, P .
MAGNETIC RESONANCE IMAGING, 1998, 16 (07) :765-772
[4]
ACCURATE MEASUREMENT OF T(1) IN-VIVO IN LESS-THAN 3 SECONDS USING ECHO-PLANAR IMAGING [J].
GOWLAND, P ;
MANSFIELD, P .
MAGNETIC RESONANCE IN MEDICINE, 1993, 30 (03) :351-354
[5]
Assessment of tumor microcirculation: A new role of dynamic contrast MR imaging [J].
Griebel, J ;
Mayr, NA ;
deVries, A ;
Knopp, MV ;
Gneiting, T ;
Kremser, C ;
Essig, M ;
Hawighorst, H ;
Lukas, PH ;
Yuh, WTC .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 1997, 7 (01) :111-119
[6]
Phantom and in vivo study of the Look-Locher T1 mapping method [J].
Karlsson, M ;
Nordell, B .
MAGNETIC RESONANCE IMAGING, 1999, 17 (10) :1481-1488
[7]
Myocardial perfusion modeling using MRI [J].
Larsson, HBW ;
FritzHansen, T ;
Rostrup, E ;
Sondergaard, L ;
Ring, P ;
Henriksen, O .
MAGNETIC RESONANCE IN MEDICINE, 1996, 35 (05) :716-726
[8]
Look D. C., 1970, Review of Scientific Instruments, V41, P250, DOI 10.1063/1.1684482
[9]
MULTI-PLANAR IMAGE-FORMATION USING NMR SPIN ECHOES [J].
MANSFIELD, P .
JOURNAL OF PHYSICS C-SOLID STATE PHYSICS, 1977, 10 (03) :L55-L58
[10]
AN ALGORITHM FOR LEAST-SQUARES ESTIMATION OF NONLINEAR PARAMETERS [J].
MARQUARDT, DW .
JOURNAL OF THE SOCIETY FOR INDUSTRIAL AND APPLIED MATHEMATICS, 1963, 11 (02) :431-441