Practical aspects for the evaluation of skin doses in interventional cardiology using a new slow film

被引:26
作者
Guibelalde, E [1 ]
Vano, E
Gonzalez, L
Prieto, C
Fernandez, JM
Ten, JI
机构
[1] Univ Complutense Madrid, Dept Radiol, Med Phys Grp, E-28040 Madrid, Spain
[2] San Carlos Hosp, Med Phys Serv, Madrid 28040, Spain
[3] San Carlos Hosp, Diagnost Radiol Serv, Madrid 28040, Spain
关键词
D O I
10.1259/bjr/47455149
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 [临床医学]; 100207 [影像医学与核医学]; 1009 [特种医学];
摘要
Mapping skin doses in complex fluoroscopy interventions is useful to determine the probability of a possible injury, to detect areas of overlapping irradiation fields and to obtain a permanent register of the most exposed patient skin areas. To fulfil this task, large films with slow X-ray response can be used. Recently, Kodak has introduced a new radiotherapy verification film, named EDR2 (Extended Dose Range). The aim of this paper is to analyse the possibilities of using this new film for estimating skin dose distributions in interventions with potentially higher doses, such as complex percutaneous transluminal coronary angioplasty (PTCA), intravascular brachytherapy procedures (IVB) or cardiac ablations. The EDR2 film by Kodak is an improved option to be used in interventional cardiology to obtain maps of patient skin doses and to estimate maximum skin doses up to 1400 mGy. Film kVp dependence is negligible and the processor conditions can be standardized to obtain skin dose estimations. The linear range for accurate dose measurements is from 50 mGy to 500 mGy.
引用
收藏
页码:332 / 336
页数:5
相关论文
共 7 条
[1]
American National Standards Institute, 1982, PH243 ANSI
[2]
A SURVEY OF FILMS FOR USE AS DOSIMETERS IN INTERVENTIONAL RADIOLOGY [J].
FAJARDO, LC ;
GEISE, RA ;
RITENOUR, ER .
HEALTH PHYSICS, 1995, 68 (04) :595-599
[3]
Comparison of four techniques to estimate radiation dose to skin during angiographic and interventional radiology procedures [J].
Fletcher, DW ;
Miller, DL ;
Balter, S ;
Taylor, MA .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2002, 13 (04) :391-397
[4]
Measuring skin dose with radiochromic dosimetry film in the cardiac catheterization laboratory [J].
Giles, ER ;
Murphy, PH .
HEALTH PHYSICS, 2002, 82 (06) :875-880
[5]
Patient dosimetry in interventional radiology using slow films [J].
Vano, E ;
Guibelalde, E ;
Fernandez, JM ;
Gonzalez, L ;
Ten, JI .
BRITISH JOURNAL OF RADIOLOGY, 1997, 70 :195-200
[6]
Skin radiation injuries in patients following repeated coronary angioplasty procedures [J].
Vano, E ;
Goicolea, J ;
Galvan, C ;
Gonzalez, L ;
Meiggs, L ;
Ten, JI ;
Macaya, C .
BRITISH JOURNAL OF RADIOLOGY, 2001, 74 (887) :1023-1031
[7]
WAGNER L, 1998, 84 SCI ASS ANN M RAD, P249