MR-compatible ventilator for small animals: computer-controlled ventilation for proton and noble gas imaging

被引:53
作者
Hedlund, LW [1 ]
Cofer, GP
Owen, SJ
Johnson, GA
机构
[1] Duke Univ, Med Ctr, Dept Radiol, Ctr In Vivo Microscopy, Durham, NC 27710 USA
[2] Duke Univ, Dept Biomed Engn, Durham, NC 27710 USA
关键词
mechanical ventilation; respirator; rodents; hyperpolarized gas; helium; xenon; MR microscopy;
D O I
10.1016/S0730-725X(00)00154-5
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
We describe an MR-compatible ventilator that is computer controlled to generate a variety of breathing patterns, to minimize image degrading effects of breathing motion, and to support delivery of gas anesthesia and experimental inhalational gases. A key feature of this ventilator is the breathing valve that attaches directly to the endotracheal tube to reduce dead volume and allows independent control of inspiratory and expiratory phases of ventilation This ventilator has been used in a wide variety of MR and x-ray microscopy studies of small animals, especially for MR imaging the lungs with hyperpolarized gases (He-3 & Xe-129). (C) 2000 Elsevier Science Inc. All rights reserved.
引用
收藏
页码:753 / 759
页数:7
相关论文
共 29 条
  • [1] Nuclear magnetic resonance imaging of airways in humans with use of hyperpolarized He-3
    Bachert, P
    Schad, LR
    Bock, M
    Knopp, MV
    Ebert, M
    Grossmann, T
    Heil, W
    Hofmann, D
    Surkau, R
    Otten, EW
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1996, 36 (02) : 192 - 196
  • [2] PHYSIOLOGICAL SUPPORT AND MONITORING OF CRITICALLY ILL PATIENTS DURING MAGNETIC-RESONANCE IMAGING
    BARNETT, GH
    ROPPER, AH
    JOHNSON, KA
    [J]. JOURNAL OF NEUROSURGERY, 1988, 68 (02) : 246 - 250
  • [3] In vivo He-3 MR images of guinea pig lungs
    Black, RD
    Middleton, HL
    Cates, GD
    Cofer, GP
    Driehuys, B
    Happer, W
    Hedlund, LW
    Johnson, GA
    Shattuck, MD
    Swartz, JC
    [J]. RADIOLOGY, 1996, 199 (03) : 867 - 870
  • [4] MR microscopy of lung airways with hyperpolarized 3He
    Chen, XJ
    Chawla, MS
    Hedlund, LW
    Möller, HE
    MacFall, JR
    Johnson, GA
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1998, 39 (01) : 79 - 84
  • [5] DUNN V, 1985, Magnetic Resonance Imaging, V3, P169, DOI 10.1016/0730-725X(85)90254-1
  • [6] MR MICROSCOPY OF THE RAT LUNG USING PROJECTION RECONSTRUCTION
    GEWALT, SL
    GLOVER, GH
    HEDLUND, LW
    COFER, GP
    MACFALL, JR
    JOHNSON, GA
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1993, 29 (01) : 99 - 106
  • [7] MODIFIED VENTILATOR WITH LOGIC-CONTROLLER FOR CARDIORESPIRATORY SYNCHRONIZATION OF MAGNETIC-RESONANCE-IMAGING IN SMALL ANIMALS
    GROVE, JR
    BEHBIN, A
    KOST, J
    [J]. MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1995, 33 (01) : 104 - 107
  • [8] HEDLUND L, 1995, MR MICR C WURZB GERM, V36
  • [9] A VENTILATOR FOR MAGNETIC-RESONANCE-IMAGING
    HEDLUND, LW
    DEITZ, J
    NASSAR, R
    HERFKENS, R
    VOCK, P
    DAHLKE, J
    KUBEK, R
    EFFMANN, EL
    PUTMAN, CE
    [J]. INVESTIGATIVE RADIOLOGY, 1986, 21 (01) : 18 - 23
  • [10] MR MICROSCOPY OF THE RAT THORAX
    HEDLUND, LW
    JOHNSON, GA
    KARIS, JP
    EFFMANN, EL
    [J]. JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1986, 10 (06) : 948 - 952