Miniaturized capillary electrophoresis system with a carbon nanotube microelectrode for rapid separation and detection of thiols

被引:188
作者
Chen, G
Zhang, LY
Wang, J
机构
[1] Fudan Univ, Dept Chem, Shanghai 200032, Peoples R China
[2] New Mexico State Univ, Dept Chem & Biochem, Las Cruces, NM 88003 USA
基金
中国国家自然科学基金;
关键词
miniaturization; capillary electrophoresis; amperometric detection; carbon nanotube; homocysteine; cysteine; glutathione; N-acetylcysteine;
D O I
10.1016/j.talanta.2004.04.022
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Multi-walled carbon nanotube (CNT) was mixed with epoxy to fabricate microdisc electrode used as a detector for a specially designed miniaturized capillary electrophoresis (CE)-amperometric detection system for the separation and detection of several bioactive thiols. The end-channel CNT amperometric detector offers favourable signal-to-noise characteristics at a relatively low potential (0.8 V) for detecting thiol compounds. Factors influencing the separation and detection processes were examined and optimized. Four thiols (homocysteine, cysteine, glutathione, and N-acetylcysteine) have been separated within 130 s at a separation voltage of 2000 V using a 20 mM phosphate running buffer (pH 7.8). Highly linear response is obtained for homocysteme, cysteine, glutathione, and N-acetylcysteine over the range of 5-50 muM with detection limits of 0.75, 0.8, 2.9, and 3.3 muM, respectively. Good stability and reproducibility (R.S.D. < 5%) are obtained reflecting the minimal adsorption of thiols at the CNT electrode surface. The new microchip protocol should find a wide range of bioanalytical applications involving assays of thiol compounds. (C) 2004 Elsevier B.V. All rights reserved.
引用
收藏
页码:1018 / 1023
页数:6
相关论文
共 32 条
[1]   A specific HPLC-UV method for the determination of cysteine and related aminothiols in biological samples [J].
Amarnath, K ;
Amarnath, V ;
Amarnath, K ;
Valentine, HL ;
Valentine, WM .
TALANTA, 2003, 60 (06) :1229-1238
[2]   Carbon nanotubes - the route toward applications [J].
Baughman, RH ;
Zakhidov, AA ;
de Heer, WA .
SCIENCE, 2002, 297 (5582) :787-792
[3]   In vivo monitoring of the thiols in rat striatum by liquid chromatography with amperometric detection at a functionalized multi-wall carbon nanotubes modified electrode [J].
Cao, XN ;
Lin, L ;
Xian, YZ ;
Zhang, W ;
Xie, YF ;
Jin, LT .
ELECTROANALYSIS, 2003, 15 (10) :892-897
[4]  
Caussé E, 2000, ELECTROPHORESIS, V21, P2074, DOI 10.1002/1522-2683(20000601)21:10<2074::AID-ELPS2074>3.0.CO
[5]  
2-L
[6]   Determination of purine and pyrimidine bases in DNA by micellar electrokinetic capillary chromatography with electrochemical detection [J].
Chen, G ;
Han, XH ;
Zhang, LY ;
Ye, JN .
JOURNAL OF CHROMATOGRAPHY A, 2002, 954 (1-2) :267-276
[7]   HELICAL MICROTUBULES OF GRAPHITIC CARBON [J].
IIJIMA, S .
NATURE, 1991, 354 (6348) :56-58
[8]   Determination of thiols by capillary electrophoresis with amperometric detection at a coenzyme pyrroloquinoline quinone modified electrode [J].
Inoue, T ;
Kirchhoff, JR .
ANALYTICAL CHEMISTRY, 2002, 74 (06) :1349-1354
[9]   Pre-column derivatization high-performance liquid chromatographic method for determination of cysteine, cysteinyl-glycine, homocysteine and glutathione in plasma and cell extracts [J].
Katrusiak, AE ;
Paterson, PG ;
Kamencic, H ;
Shoker, A ;
Lyon, AW .
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2001, 758 (02) :207-212
[10]  
Krijt J, 2001, CLIN CHEM, V47, P1821