Development of an immunosensor for the determination of rabbit IgG using streptavidin modified screen-printed carbon electrodes

被引:50
作者
Díaz-González, M [1 ]
Hernández-Santos, D [1 ]
González-García, MB [1 ]
Costa-García, A [1 ]
机构
[1] Univ Oviedo, Fac Quim, Dept Quim Fis & Analit, E-33006 Oviedo, Spain
关键词
alkaline phosphatase : 3-indoxyl phosphate; screen-printed carbon electrode; streptavidin : biotin reaction; rabbit IgG; immunosensor;
D O I
10.1016/j.talanta.2004.07.022
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Voltammetric enzyme immunosensors based on the employment of streptavidin modified screen-printed carbon electrodes (SPCEs) for the detection of rabbit IgG, as a model analyte, were described. Alkaline phosphatase (AP) and 3-indoxyl phosphate (3-1P) were used as the enzymatic label and substrate, respectively. The adsorption of streptavidin was performed by deposition of a drop of a streptavidin solution overnight at 4degreesC on the pre-oxidized surface of the SPCEs. The analytical characteristics of these sensors were evaluated using biotin conjugated to AP. The immunosensor devices were based on a specific reaction of rabbit l.-G with its biotinylated antibodies, which were immobilised on the modified screen-printed carbon electrodes through the streptavidin:biotin reaction. The immunosensors were used for a direct determination of AP labelled rabbit IgG, and for free rabbit IgG detection using sequential competitive immunoassay. A calibration curve in the range of 9 5 x 10(-11) to 1 x 10(-9) M of rabbit IgG was obtained with a estimated detection limit of 5 x 10(-11) M (7.0 ng/ml). These immunosensors were stable for 5 months if they were stored at 4degreesC. (C) 2004 Elsevier B.V. All rights reserved.
引用
收藏
页码:565 / 573
页数:9
相关论文
共 34 条
[1]   AVIDIN-BIOTIN BASED ELECTROCHEMICAL IMMUNOASSAY FOR THYROTROPIN [J].
ATHEY, D ;
BALL, M ;
MCNEIL, CJ .
ANNALS OF CLINICAL BIOCHEMISTRY, 1993, 30 :570-577
[2]   Disposable amperometric immunosensor system for rabbit IgG using a conducting polymer modified screen-printed electrode [J].
Darain, F ;
Park, SU ;
Shim, YB .
BIOSENSORS & BIOELECTRONICS, 2003, 18 (5-6) :773-780
[3]  
Díaz-González M, 2002, ELECTROANAL, V14, P665, DOI 10.1002/1521-4109(200205)14:10<665::AID-ELAN665>3.0.CO
[4]  
2-C
[5]   Indirect determination of alkaline phosphatase based on the amperometric detection of indigo carmine at a screen-printed electrode in a flow system [J].
Díaz-González, M ;
Fernández-Sánchez, C ;
Costa-García, A .
ANALYTICAL SCIENCES, 2002, 18 (11) :1209-1213
[6]   Competitive enzyme immunosensor developed on a renewable carbon paste electrode support [J].
Fernández-Sánchez, C ;
Costa-García, A .
ANALYTICA CHIMICA ACTA, 1999, 402 (1-2) :119-127
[7]   AC voltammetric carbon paste-based enzyme immunosensors [J].
Fernández-Sánchez, C ;
González-García, MB ;
Costa-García, A .
BIOSENSORS & BIOELECTRONICS, 2000, 14 (12) :917-924
[8]  
FERNANDEZSANCHE.C, 1998, ELECTROANAL, V10, P403
[9]   Immunosensors: Electrochemical sensing and other engineering approaches [J].
Ghindilis, AL ;
Atanasov, P ;
Wilkins, M ;
Wilkins, E .
BIOSENSORS & BIOELECTRONICS, 1998, 13 (01) :113-131
[10]  
GREEN MJ, 1987, BIOSENSORS FUNDAMENT, P57