Precise analysis of primary amino acids in urine by an automated high-performance liquid chromatography method: comparison with ion-exchange chromatography

被引:29
作者
Fekkes, D [1 ]
Voskuilen-Kooyman, A
Jankie, R
Huijmans, J
机构
[1] Dept Psychiat, Sect Pathophysiol Behav, Rotterdam, Netherlands
[2] Erasmus Univ, Dept Clin Genet, NL-3000 DR Rotterdam, Netherlands
来源
JOURNAL OF CHROMATOGRAPHY B | 2000年 / 744卷 / 01期
关键词
amino acids;
D O I
10.1016/S0378-4347(00)00234-6
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A precise, simple and rapid method for the quantitative determination of primary amino acids in urine based on high-performance liquid chromatography and o-phthaldialdehyde pre-column derivatization is described. All primary urinary amino acids could be determined within 49 min (injection to injection). Amino acid concentrations in 40 urinary samples were measured by this method and the results were compared with those measured by ion-exchange chromatography. The correlation coefficient for the common amino acids was greater than 0.90. This is the first study in which such a detailed comparison has been made on urine samples. It appeared that the method described is an excellent alternative to the classical ion-exchange method for the quantitation of urinary amino acids. (C) 2000 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:183 / 188
页数:6
相关论文
共 14 条
[1]  
BLOM W, 1992, Amino Acids (Vienna), V2, P25, DOI 10.1007/BF00806075
[2]   Automated method for the measurement of amino acids in urine by high-performance liquid chromatography [J].
Carducci, C ;
Birarelli, M ;
Leuzzi, V ;
Santagata, G ;
Serafini, P ;
Antonozzi, I .
JOURNAL OF CHROMATOGRAPHY A, 1996, 729 (1-2) :173-180
[3]   AMINO-ACID-ANALYSIS OF PHYSIOLOGICAL FLUIDS BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY WITH PHENYLISOTHIOCYANATE DERIVATIZATION AND COMPARISON WITH ION-EXCHANGE CHROMATOGRAPHY [J].
DAVEY, JF ;
ERSSER, RS .
JOURNAL OF CHROMATOGRAPHY-BIOMEDICAL APPLICATIONS, 1990, 528 (01) :9-23
[4]   VALIDATION OF THE DETERMINATION OF AMINO-ACIDS IN PLASMA BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY USING AUTOMATED PRECOLUMN DERIVATIZATION WITH O-PHTHALDIALDEHYDE [J].
FEKKES, D ;
VANDALEN, A ;
EDELMAN, M ;
VOSKUILEN, A .
JOURNAL OF CHROMATOGRAPHY B-BIOMEDICAL APPLICATIONS, 1995, 669 (02) :177-186
[5]   State-of-the-art of high-performance liquid chromatographic analysis of amino acids in physiological samples [J].
Fekkes, D .
JOURNAL OF CHROMATOGRAPHY B-BIOMEDICAL APPLICATIONS, 1996, 682 (01) :3-22
[6]   REVERSED-PHASE CHROMATOGRAPHY OF PHENYLTHIOCARBAMYL AMINO-ACID DERIVATIVES OF PHYSIOLOGICAL AMINO-ACIDS - AN EVALUATION AND A COMPARISON WITH ANALYSIS BY ION-EXCHANGE CHROMATOGRAPHY [J].
FESTE, AS .
JOURNAL OF CHROMATOGRAPHY-BIOMEDICAL APPLICATIONS, 1992, 574 (01) :23-34
[7]   ORTHO-PHTHALDIALDEHYDE PRECOLUMN DERIVATIZATION AND REVERSED-PHASE HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY OF POLYPEPTIDE HYDROLYSATES AND PHYSIOLOGICAL FLUIDS [J].
JONES, BN ;
GILLIGAN, JP .
JOURNAL OF CHROMATOGRAPHY, 1983, 266 (AUG) :471-482
[8]  
QURESHI GA, 1984, J CHROMATOGR, V297, P91, DOI 10.1016/S0021-9673(01)89032-4
[9]   HPLC QUALITATIVE AMINO-ACID-ANALYSIS IN THE CLINICAL LABORATORIES [J].
RUTLEDGE, JC ;
RUDY, J .
AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 1987, 87 (05) :614-618
[10]   MEASUREMENT OF PLASMA AND URINE AMINO-ACIDS BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY WITH ELECTROCHEMICAL DETECTION USING PHENYLISOTHIOCYANATE DERIVATIZATION [J].
SHERWOOD, RA ;
TITHERADGE, AC ;
RICHARDS, DA .
JOURNAL OF CHROMATOGRAPHY-BIOMEDICAL APPLICATIONS, 1990, 528 (02) :293-303