A comprehensive model for power line interference in biopotential measurements

被引:49
作者
Chimeno, MF [1 ]
Pallàs-Areny, R [1 ]
机构
[1] Univ Politecn Catalunya, Dept Elect Engn, Barcelona 08034, Spain
关键词
amplifier; biomedical measurements; electrodes; interference; shielding;
D O I
10.1109/19.850390
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Power line interference is a major problem in high-resolution biopotential measurements. Because interference coupling is mostly capacitive, shielding electrode leads and a high common-mode rejection ratio (CMRR) are quite effective in reducing power-line interference but do not completely eliminate it. We propose a model that includes both interference external to the measuring system and interference coming from its internal power supply. Moreover, the model considers interference directly coupled to the measuring electrodes, because, as opposed to connecting leads, electrodes are not usually shielded. Experimental results confirm that reducing interference coupled through electrodes yields a negligible interference. The proposed model can be applied to other differential measurement systems, particulary those involving electrodes or sensors placed far apart.
引用
收藏
页码:535 / 540
页数:6
相关论文
共 9 条
[1]  
[Anonymous], 1970, 123 HEWL PACK
[2]   60-HZ INTERFERENCE IN ELECTROCARDIOGRAPHY [J].
HUHTA, JC ;
WEBSTER, JG .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1973, BM20 (02) :91-101
[4]   GROUND-FREE ECG RECORDING WITH 2 ELECTRODES [J].
THAKOR, NV ;
WEBSTER, JG .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1980, 27 (12) :699-704
[5]   HIGH-QUALITY RECORDING OF BIOELECTRIC EVENTS .1. INTERFERENCE REDUCTION, THEORY AND PRACTICE [J].
VANRIJN, ACM ;
PEPER, A ;
GRIMBERGEN, CA .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1990, 28 (05) :389-397
[6]   THE ISOLATION MODE REJECTION RATIO IN BIOELECTRIC AMPLIFIERS [J].
VANRIJN, ACM ;
PEPER, A ;
GRIMBERGEN, CA .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1991, 38 (11) :1154-1157
[7]  
WIECHERT L, 1985, INSTRUMENT SCI TECHN, V3, P1021
[8]   REDUCTION OF INTERFERENCE DUE TO COMMON-MODE VOLTAGE IN BIOPOTENTIAL AMPLIFIERS [J].
WINTER, BB ;
WEBSTER, JG .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1983, 30 (01) :58-62
[9]   COMPARATIVE-ANALYSIS OF POWER-LINE INTERFERENCE BETWEEN 2-ELECTRODE OR 3-ELECTRODE BIOPOTENTIAL AMPLIFIERS [J].
WOOD, DE ;
EWINS, DJ ;
BALACHANDRAN, W .