Dosimetric study of a new palladium seed

被引:23
作者
Reniers, B [1 ]
Vynckier, S [1 ]
Scalliet, P [1 ]
机构
[1] Catholic Univ Louvain, Clin Univ St Luc, Serv Radiotherapie Oncol, B-1200 Brussels, Belgium
关键词
brachytherapy; dosimetry; palladium(103); Monte Carlo simulation; MCNP4B;
D O I
10.1016/S0969-8043(02)00193-8
中图分类号
O61 [无机化学];
学科分类号
070301 [无机化学]; 081704 [应用化学];
摘要
In this paper, the dosimetric parameters for a new palladium seed design: the InterSource(103,1) are presented as recommended by the AAPM in the TG-43 formalism. Measurements are made with LiF thermoluminescent dosimeters (Size of 1 mm(3)) in solid water phantoms WT1 to obtain the dose constant, the radial dose function and the anisotropy function. The TLD were calibrated at 6 MV and an energy correction factor of 1.40 has been applied. The same dose parameters are also obtained by Monte Carlo calculations (MCNP4B) in solid water and in liquid water. The calculated and the measured TG-43 functions for solid water are in excellent agreement. In WT1, the calculated dose rate constant is 0.657 +/- 1% and the measured value is 0.672 +/- 7%. The calculated value for water is 0.692 +/- 1%. The comparison with the previous study (Med. Phys. 27(5) (2000)) shows a very good agreement for the dose rate constant. The agreement for the radial function is poorer. For the measurements, it can be due to the difference of TLD settings. For the calculations the discrepancy could come from the different cross-section data utilized in the different Monte Carlo codes. In conclusion, the dosimetric functions for the new iodine seed InterSource(103) have been determined using the MCNP4B Monte Carlo code and TLD measurement in solid water WT1. (C) 2002 Elsevier Science Ltd. All rights reserved.
引用
收藏
页码:805 / 811
页数:7
相关论文
共 17 条
[1]
[Anonymous], 1989, TISS SUBST RAD DOS M
[2]
Berger M.J., 1987, United States, Technical Report., DOI DOI 10.2172/6016002
[3]
BRIESMEISTER JF, MCNP GEN MONTE CARLO
[4]
EVERETT CJ, 1973, LA5240MS LOS AL SCI
[5]
Hubbell J. H., 1975, Journal of Physical and Chemical Reference Data, V4, P471, DOI 10.1063/1.555523
[6]
COMPARISON OF RADIATION-DOSIMETRY IN WATER AND IN SOLID PHANTOM MATERIALS FOR I-125 AND PD-103 BRACHYTHERAPY SOURCES - EGS4 MONTE-CARLO STUDY [J].
LUXTON, G .
MEDICAL PHYSICS, 1994, 21 (05) :631-641
[7]
INSTRUMENTATION AND DOSIMETER-SIZE ARTIFACTS IN QUANTITATIVE THERMOLUMINESCENCE DOSIMETRY OF LOW-DOSE FIELDS [J].
MEIGOONI, AS ;
MISHRA, V ;
PANTH, H ;
WILLIAMSON, J .
MEDICAL PHYSICS, 1995, 22 (05) :555-561
[8]
Dosimetric characteristics of the InterSource103 palladium brachytherapy source [J].
Meigooni, AS ;
Sowards, K ;
Soldano, M .
MEDICAL PHYSICS, 2000, 27 (05) :1093-1100
[9]
A COMPARISON OF SOLID PHANTOMS WITH WATER FOR DOSIMETRY OF I125 BRACHYTHERAPY SOURCES [J].
MEIGOONI, AS ;
MELI, JA ;
NATH, R .
MEDICAL PHYSICS, 1988, 15 (05) :695-701
[10]
MEIGOONI AS, 1990, ENDOCURIETHER HYPERT, V6, P107