Analysis of forces developed during mechanical preparation of extracted teeth using RaCe rotary instruments and ProFiles

被引:18
作者
da Silva, FM [1 ]
Kobayashi, C [1 ]
Suda, H [1 ]
机构
[1] Tokyo Med & Dent Univ, Grad Sch, Dept Restorat Sci, Bunkyo Ku, Tokyo 1138548, Japan
关键词
nickel-titanium; rotary instruments; torque; vertical load;
D O I
10.1111/j.1365-2591.2004.00890.x
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Aim To compare the torque and load during instrumentation with ProFile and RaCe nickel-titanium rotary instruments. Methodology Thirty human incisor roots 12 mm in length were embedded in epoxy resin, and divided into three groups of 10 specimens each. Instruments employed in each group were as follows: RaCe .02 Step Back Files in group 1; RaCe .04 Step Back Files in group 2; and .04 ProFiles in group 3. Two load cells, whose outputs were connected to a digital oscilloscope, recorded torque and vertical load. Data were analysed by one-way ANOVA. Results Torque values were statistically different between groups 1 and 3, and between groups 2 and 3 (torque was higher in group 3 than in groups 1 and 2). Vertical loads were statistically higher in group 3 than in groups 1 and 2. Conclusions Torsional and vertical forces can be evaluated during instrumentation of straight root canals using the device and methods described in this study. When the step-back technique was employed, torque and vertical load obtained with RaCe instruments were lower than that obtained with ProFiles.
引用
收藏
页码:17 / 21
页数:5
相关论文
共 16 条
[1]   The 'Lightspeed' preparation technique evaluated by Swiss clinicians after attending continuing education courses [J].
Barbakow, F ;
Lutz, F .
INTERNATIONAL ENDODONTIC JOURNAL, 1997, 30 (01) :46-50
[2]   Analysis of forces developed during mechanical preparation of extracted teeth using Profile NiTi rotary instruments [J].
Blum, JY ;
Cohen, A ;
Machtou, P ;
Micallef, JP .
INTERNATIONAL ENDODONTIC JOURNAL, 1999, 32 (01) :24-31
[3]  
Bryant ST, 1998, INT ENDOD J, V31, P275
[4]  
Gambarini G, 2001, Aust Endod J, V27, P99
[5]   Cyclic fatigue of Profile rotary instruments after prolonged clinical use [J].
Gambarini, G .
INTERNATIONAL ENDODONTIC JOURNAL, 2001, 34 (05) :386-389
[6]   A COMPARISON OF ROOT-CANAL PREPARATIONS USING NI-TI HAND, NI-TI ENGINE-DRIVEN, AND K-FLEX ENDODONTIC INSTRUMENTS [J].
GLOSSON, CR ;
HALLER, RH ;
DOVE, SB ;
DELRIO, CE .
JOURNAL OF ENDODONTICS, 1995, 21 (03) :146-151
[7]   A new engine-driven canal preparation system with electronic canal measuring capability [J].
Kobayashi, C ;
Yoshioka, T ;
Suda, H .
JOURNAL OF ENDODONTICS, 1997, 23 (12) :751-754
[8]  
KOBYAASHI C, 2002, JAPANESE J CONSERVAT, V45, P1044
[9]   Rotary Ni-Ti profile systems for preparing curved canals in resin blocks: influence of operator on instrument breakage [J].
Mandel, E ;
Adib-Yazdi, M ;
Benhamou, LM ;
Lachkar, T ;
Mesgouez, C ;
Sobel, M .
INTERNATIONAL ENDODONTIC JOURNAL, 1999, 32 (06) :436-443
[10]   ProTaper rotary root canal preparation:: assessment of torque and force in relation to canal anatomy [J].
Peters, OA ;
Peters, CI ;
Schönenberger, K ;
Barbakow, F .
INTERNATIONAL ENDODONTIC JOURNAL, 2003, 36 (02) :93-99