Comparison of the Micronaut Merlin automated broth microtiter system with the standard agar dilution method for antimicrobial susceptibility testing of mucoid and nonmucoid Pseudomonas aeruginosa isolates from cystic fibrosis patients

被引:14
作者
Balke, B
Hoy, L
Weissbrodt, H
Häussler, S
机构
[1] Gesell Biotechnol Forsch mbH, Abt Zellbiol, D-38124 Braunschweig, Germany
[2] Hannover Med Sch, Inst Biometrie, D-30625 Hannover, Germany
[3] Hannover Med Sch, Inst Med Mikrobiol & Krankenhaushyg, D-30625 Hannover, Germany
关键词
D O I
10.1007/s10096-004-1212-7
中图分类号
R51 [传染病];
学科分类号
100401 ;
摘要
The aim of this study was to compare a commercially available automated broth microdilution system (Merlin; Micronaut, Germany) with the standard agar dilution method for susceptibility testing of pulmonary isolates from cystic fibrosis patients. Accurate susceptibility testing of bacterial isolates from cystic fibrosis patients is known to pose problems. Although commercially available automated test systems could facilitate susceptibility testing of such isolates in routine diagnostics, these systems have not been recommended thus far. However, a pilot study recently indicated that the Merlin system, which is based on an endpoint measurement rather than on growth curves, might be applicable in the susceptibility testing of isolates from cystic fibrosis patients. In the present study, the Merlin system was further evaluated using an extended panel of nonmucoid and mucoid Pseudomonas aeruginosa isolates. The results showed that the MICs obtained with the Merlin system tended to be lower than those obtained with the agar dilution method, a finding that became increasingly apparent when mucoid Pseudomonas aeruginosa strains were tested. The correlation coefficients (r values) of the MIC results for all strains tested were between 0.6 and 0.8 for five of the seven antimicrobial agents, with r values exceeding 0.8 for only meropenem and ciprofloxacin. However, since the overall rate of serious discrepancies was within an acceptable range, the Merlin system appears to be applicable for routine clinical use in susceptibility testing of P. aeruginosa isolates from cystic fibrosis patients.
引用
收藏
页码:765 / 771
页数:7
相关论文
共 16 条
[1]   Comparison of two commercial systems (Vitek and MicroScan-WalkAway) for antimicrobial susceptibility testing of Pseudomonas aeruginosa isolates from cystic fibrosis patients [J].
Burns, JL ;
Saiman, L ;
Whittier, S ;
Krzewinski, J ;
Liu, ZL ;
Larone, D ;
Marshall, SA ;
Jones, RN .
DIAGNOSTIC MICROBIOLOGY AND INFECTIOUS DISEASE, 2001, 39 (04) :257-260
[2]   The Calgary Biofilm Device: New technology for rapid determination of antibiotic susceptibilities of bacterial biofilms [J].
Ceri, H ;
Olson, ME ;
Stremick, C ;
Read, RR ;
Morck, D ;
Buret, A .
JOURNAL OF CLINICAL MICROBIOLOGY, 1999, 37 (06) :1771-1776
[3]   Approach to resistant gram-negative bacterial pulmonary infections in patients with cystic fibrosis [J].
Chernish, RN ;
Aaron, SD .
CURRENT OPINION IN PULMONARY MEDICINE, 2003, 9 (06) :509-515
[4]   GENDER DIFFERENCES IN CYSTIC-FIBROSIS - PSEUDOMONAS-AERUGINOSA INFECTION [J].
DEMKO, CA ;
BYARD, PJ ;
DAVIS, PB .
JOURNAL OF CLINICAL EPIDEMIOLOGY, 1995, 48 (08) :1041-1049
[5]   Antimicrobial resistance of Pseudomonas aeruginosa biofilms [J].
Drenkard, E .
MICROBES AND INFECTION, 2003, 5 (13) :1213-1219
[6]   Evaluation of the Merlin, Micronaut system for automated antimicrobial susceptibility testing of Pseudomonas aeruginosa and Burkholderia species isolated from cystic fibrosis patients [J].
Häussler, S ;
Ziesing, S ;
Rademacher, G ;
Hoy, L ;
Weissbrodt, H .
EUROPEAN JOURNAL OF CLINICAL MICROBIOLOGY & INFECTIOUS DISEASES, 2003, 22 (08) :496-500
[7]   Alginate overproduction affects Pseudomonas aeruginosa biofilm structure and function [J].
Hentzer, M ;
Teitzel, GM ;
Balzer, GJ ;
Heydorn, A ;
Molin, S ;
Givskov, M ;
Parsek, MR .
JOURNAL OF BACTERIOLOGY, 2001, 183 (18) :5395-5401
[8]   Diagnosis and treatment of cystic fibrosis [J].
Koch, C ;
Holby, N .
RESPIRATION, 2000, 67 (03) :239-247
[9]   Lung infections associated with cystic fibrosis [J].
Lyczak, JB ;
Cannon, CL ;
Pier, GB .
CLINICAL MICROBIOLOGY REVIEWS, 2002, 15 (02) :194-+
[10]  
NCCLS (National Committee for Clinical Laboratory Standards), 2002, PERF STAND ANT SUSC