Fracture of a carotid stent: An unexpected complication

被引:51
作者
Valibhoy, Arif Rahim [1 ]
Mwipatayi, Bibombe Patrice [1 ]
Surg, Cert Vase [1 ]
Sieunarine, Kishore [1 ]
机构
[1] Royal Perth Hosp, Dept Vasc Surg, Perth, WA 6000, Australia
关键词
D O I
10.1016/j.jvs.2006.08.086
中图分类号
R61 [外科手术学];
学科分类号
摘要
We present the first reported case of a carotid artery stent fracture where a Nitinol stent was used for the management of carotid artery stenosis. An 83-year-old man underwent right carotid artery stenting in July 2005. On follow-up ultrasound, stenosis of the right internal carotid artery was noted. In January 2006, he underwent left carotid artery stenting and at this procedure, precontrast assessment of the right neck revealed a complete fracture through the waist of the stent in the right internal carotid artery. This fractured segment correlated with the area of stenosis reported at the previous ultrasound. In July 2006, this gentleman underwent successful removal of the carotid stent with uneventful carotid endarterectomy and synthetic patch repair. Carotid artery stenting is a newer and alternative modality to carotid endarterectomy for the treatment of carotid artery stenting. More data are emerging to support its safety and efficacy. An unexpected complication can happen with any procedure and complications help define the follow-up management of subsequent patients undergoing this procedure.
引用
收藏
页码:603 / 606
页数:4
相关论文
共 12 条
[1]
Renal artery thrombosis caused by stent fracture in a single kidney patient [J].
Bessias, N ;
Sfyroeras, G ;
Moulakakis, KG ;
Karalkasis, F ;
Ferentinou, E ;
Andrikopoulos, V .
JOURNAL OF ENDOVASCULAR THERAPY, 2005, 12 (04) :516-520
[2]
Stent fracture after endoluminal repair of a carotid artery pseudoaneurysm [J].
de Vries, JPPM ;
Meijer, RW ;
van den Berg, JC ;
Meijer, JMR ;
van de Pavoordt, EDWM .
JOURNAL OF ENDOVASCULAR THERAPY, 2005, 12 (05) :612-615
[3]
High-risk carotid endarterectomy: Challenges for carotid stent protocols [J].
Jordan, WD ;
Alcocer, F ;
Wirthlin, DJ ;
Fisher, WS ;
Warren, JA ;
McDowell, HA ;
Whitley, ND .
JOURNAL OF VASCULAR SURGERY, 2002, 35 (01) :16-21
[4]
Longitudinal Stent fracture 11 months after implantation in the left pulmonary artery and successful management by a stent-in-stent maneuver [J].
Knirsch, W ;
Haas, NA ;
Lewin, MAG ;
Uhlemann, F .
CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2003, 58 (01) :116-118
[5]
Axial stent strut angle influences wall shear stress after stent implantation: analysis using 3D computational fluid dynamics models of stent foreshortening [J].
LaDisa, John F., Jr. ;
Olson, Lars E. ;
Hettrick, Douglas A. ;
Warltier, David C. ;
Kersten, Judy R. ;
Pagel, Paul S. .
BIOMEDICAL ENGINEERING ONLINE, 2005, 4 (1)
[6]
McKelvey AL, 1999, J BIOMED MATER RES, V47, P301, DOI 10.1002/(SICI)1097-4636(19991205)47:3<301::AID-JBM3>3.0.CO
[7]
2-H
[8]
Serial stent implantation to relieve in-stent stenosis in obstructed total anomalous pulmonary venous return [J].
Michel-Behnke, I ;
Luedemann, M ;
Hagel, KJ ;
Schranz, D .
PEDIATRIC CARDIOLOGY, 2002, 23 (02) :221-223
[9]
The nitinol SMART stent vs Wallstent for suboptimal iliac artery angioplasty: CRISP-US trial results [J].
Ponec, D ;
Jaff, MR ;
Swischuk, J ;
Feiring, A ;
Laird, J ;
Mehra, M ;
Popma, JJ .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2004, 15 (09) :911-918
[10]
Prevalence and clinical impact of stent fractures after femoropopliteal stenting [J].
Scheinert, D ;
Scheinert, S ;
Sax, J ;
Piorkowski, C ;
Bräunlich, S ;
Ulrich, M ;
Biamino, G ;
Schmidt, A .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2005, 45 (02) :312-315