Induction of atrial fibrillation with low-energy defibrillator shocks in patients with implantable cardioverter defibrillators

被引:13
作者
Florin, TJ
Weiss, DN
Peters, RW
Shorofsky, SR
Gold, MR
机构
[1] Department of Medicine, Div. Cardiol., Univ. Maryland S., Baltimore, MD
关键词
D O I
10.1016/S0002-9149(97)00556-0
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
In a population of 151 consecutive patients who received an implantable cardioverter defibrillator, we found that atrial fibrillation was induced by low-energy shocks in 19% and was most common in patients with lead systems that included a right atrial electrode. Our finding that there was a fixed relation between the energy required to fibrillate (less than or equal to 3 J) and defibrillate (>3 J) suggests the presence of an upper limit of vulnerability in the human atrium.
引用
收藏
页码:960 / &
页数:4
相关论文
共 19 条
[11]   VENTRICULAR-FIBRILLATION INDUCED BY LOW-ENERGY SHOCKS FROM PROGRAMMABLE IMPLANTABLE CARDIOVERTER-DEFIBRILLATORS IN PATIENTS WITH CORONARY-ARTERY DISEASE [J].
LAUER, MR ;
YOUNG, C ;
LIEM, LB ;
OTTOBONI, L ;
PETERSON, J ;
GOOLD, P ;
SUNG, RJ .
AMERICAN JOURNAL OF CARDIOLOGY, 1994, 73 (08) :559-563
[12]   THE AUTOMATIC IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR - EFFICACY, COMPLICATIONS, AND DEVICE FAILURES [J].
MARCHLINSKI, FE ;
FLORES, BT ;
BUXTON, AE ;
HARGROVE, WC ;
ADDONIZIO, VP ;
STEPHENSON, LW ;
HARKEN, AH ;
DOHERTY, JU ;
GROGAN, EW ;
JOSEPHSON, ME .
ANNALS OF INTERNAL MEDICINE, 1986, 104 (04) :481-488
[13]   THE PROARRHYTHMIC POTENTIAL OF IMPLANTABLE CARDIOVERTER-DEFIBRILLATORS [J].
PINSKI, SL ;
FAHY, GJ .
CIRCULATION, 1995, 92 (06) :1651-1664
[14]  
SERDLOW CD, 1996, J AM COLL CARDIOL, V27, P1112
[15]   Effects of waveform and polarity on defibrillation thresholds in humans using a transvenous lead system [J].
Shorofsky, SR ;
Gold, MR .
AMERICAN JOURNAL OF CARDIOLOGY, 1996, 78 (03) :313-316
[16]   EFFECT OF SHOCK POLARITY ON VENTRICULAR DEFIBRILLATION THRESHOLD USING A TRANSVENOUS LEAD SYSTEM [J].
STRICKBERGER, SA ;
HUMMEL, JD ;
HORWOOD, LE ;
JENTZER, J ;
DAOUD, E ;
NIEBAUER, M ;
BAKR, O ;
MAN, KC ;
WILLIAMSON, BD ;
KOU, W ;
MORADY, F .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1994, 24 (04) :1069-1072
[17]   Chronic rise in defibrillation threshold with a hybrid lead system [J].
Tummala, RV ;
Riggio, DR ;
Peters, RW ;
Shorofsky, SR ;
Gold, MR .
AMERICAN JOURNAL OF CARDIOLOGY, 1996, 78 (03) :309-312
[18]   DYSRHYTHMIAS AFTER DIRECT-CURRENT CARDIOVERSION [J].
WALDECKER, B ;
BRUGADA, P ;
ZEHENDER, M ;
STEVENSON, W ;
WELLENS, HJJ .
AMERICAN JOURNAL OF CARDIOLOGY, 1986, 57 (01) :120-123
[19]   CATHETER-BASED ATRIAL DEFIBRILLATION [J].
WHARTON, JM ;
JOHNSON, EE .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1994, 17 (05) :1058-1066